CUS200M-18 TDK-Lambda 252W 18VDC Medical Open Frame PSU

$138.15

TDK-Lambda CUS200M-18 (CUS200M18) 201.6W/252W 18VDC medical open frame power supply, 85-265VAC input, featuring 2xMOPP isolation, 5V standby, and 94% efficiency.

SKU: CUS200M-18 Category: Brand:

Description

TDK-Lambda CUS200M-18 (also known as CUS200M18) is a high-efficiency AC-DC medical and ITE open-frame power supply from the CUS200M series, where the -18 suffix designates its 18 VDC nominal main output and the blank suffix denotes the standard industry 3 x 5 inch open-frame PCB configuration. Engineered with an active power factor corrected (PFC) topology achieving up to 94% efficiency at 230 VAC (92% at 115 VAC), it delivers 201.6 W (11.2 A) under silent natural convection cooling and scales up to 252.0 W (14.0 A) with 1.5 m/s forced air. The unit operates across a universal 85 to 265 VAC (47–63 Hz) or 120 to 370 VDC input range while maintaining tight 0.5% line regulation, 1% load regulation, and low output ripple and noise capped at 180 mVpp (1%). In addition to its adjustable 17.6 to 18.9 VDC main rail, the supply integrates an auxiliary 5 V standby output rated at 0.6 A in convection and 1.0 A under forced air, alongside isolated Remote On/Off control and a DC_OK power-good signal.

Constructed on a compact 76.2 x 34.0 x 127.0 mm (3.00 x 1.34 x 5.00 in) footprint weighing 350 g (0.77 lbs), the CUS200M-18 utilizes genuine JST headers including a B3P5-VH (CN1) AC input connector, a B6P-VH (CN51) main DC output connector, and an 8-pin CN61 control header plus an onboard VR51 trim potentiometer and green LED indicator. Comprehensive protection circuits safeguard downstream loads with overcurrent protection (>14.7 A) and latch-style overvoltage protection (19.8–24.3 VDC) requiring AC input cycling to reset. Designed for Class I installations and Type B and BF patient-applied medical equipment, it provides reinforced 4,000 VAC (2xMOPP) input-to-output isolation, 2,000 VAC (1xMOPP) input-to-ground isolation, and 1,500 VAC (1xMOPP) output-to-ground isolation with earth leakage current kept below 300 µA at 265 VAC. The supply operates from -20 °C to +70 °C (-4 °F to +158 °F) with full convection load up to +50 °C (+122 °F) and holds dual medical and ITE certifications to IEC/UL/EN 60601-1, IEC/UL/EN 62368-1, EN 55011/55022 Class B EMI, and IEC 60601-1-2 Ed4 immunity.

Key Features

  • Dual Convection and Forced-Air Power Ratings: Delivers 201.6 W (11.2 A) continuously with silent natural convection cooling for noise-sensitive hospital rooms, or up to 252.0 W (14.0 A) with 1.5 m/s forced airflow.
  • Reinforced 2xMOPP Medical Isolation Architecture: Certified for Type B and BF medical equipment with 4,000 VAC (2xMOPP) input-to-output, 2,000 VAC (1xMOPP) input-to-ground, and 1,500 VAC (1xMOPP) output-to-ground dielectric withstand.
  • Up to 94% Conversion Efficiency & Active PFC: Achieves 92% efficiency at 115 VAC and 94% at 230 VAC with active power factor correction (0.99 / 0.95) and <0.5 W standby no-load power draw.
  • Integrated 5V Standby, DC_OK & Remote Inhibit: Built-in CN61 signal interface provides a 5 VDC auxiliary output (0.6 A convection / 1.0 A forced air), DC Good status monitoring, and isolated Remote On/Off shutdown.
  • Class B EMI & 4th Edition Medical Immunity: Meets EN 55011-B, EN 55022-B, and FCC Class B conducted and radiated emissions alongside IEC 60601-1-2 Ed4 and SEMI F47 line dip immunity standards.
  • Compact 3 x 5 Inch Open-Frame Footprint: Measures 76.2 x 34.0 x 127.0 mm (3.00 x 1.34 x 5.00 in) with low <300 µA earth leakage current and wide -20 °C to +70 °C ambient operation.
  • Precision Output Trim & Hardware Protections: Onboard VR51 potentiometer adjusts the 18 VDC rail from 17.6 to 18.9 VDC, backed by >14.7 A overcurrent protection and 19.8–24.3 V latching overvoltage protection.

Technical Specifications

Manufacturer TDK-Lambda
Manufacturer Part Number (MPN) CUS200M-18
Alternate Part Number CUS200M18
Product Series CUS200M Series
Product Category Medical Open Frame
Component Type AC-DC Medical & ITE Open Frame Power Supply
Circuit Topology Switched-Mode Power Supply (SMPS) with Active PFC
Nominal Output Voltage 18 VDC (Main Output) / 5 VDC (Standby Output)
Output Voltage Adjustment Range 17.6 to 18.9 VDC (via onboard potentiometer VR51)
Rated Output Current 11.2 A (Convection Cooling) / 14.0 A (Forced Air Cooling at 1.5 m/s)
Rated Output Power 201.6 W (Convection Cooling) / 252.0 W (Forced Air Cooling at 1.5 m/s)
Number of Outputs 1 Main Output + 1 Auxiliary 5 V Standby Output
Ch1 Output Voltage & Current 18 VDC at 11.2 A (Convection) / 14.0 A (Forced Air)
Ch2 Output Voltage & Current 5 VDC Standby (5Vsb) at 0.6 A (Convection) / 1.0 A (Forced Air)
Output Ripple & Noise (Max.) 180 mVpp (180 mV or 1% of nominal output, whichever is greater)
Line Regulation 0.5% of nominal output
Load Regulation 1.0% of nominal output (No minimum load required)
Hold-Up Time 16 ms typ. (at 200 W output) / 12 ms typ. (at 250 W output, 115 / 230 VAC)
AC Input Voltage Range 85 to 265 VAC operating (100 to 240 VAC nominal; safety certified for 90 to 264 VAC)
DC Input Voltage Range 120 to 370 VDC (Safety certified for AC input only)
Nominal Input Voltage 100 to 240 VAC
Input Frequency Range 47 to 63 Hz (Safety certified for 50 to 60 Hz)
Efficiency (at 115 VAC) 92% (at 200 W output)
Efficiency (at 230 VAC) 94% (at 200 W output)
AC Input Current (at 115 VAC) 2.2 A (at 200 W output) / 3.0 A (at 250 W output)
AC Input Current (at 230 VAC) 1.1 A (at 200 W output) / 1.5 A (at 250 W output)
Inrush Current (Cold Start at 115 VAC) 35 A typ.
Inrush Current (Cold Start at 230 VAC) 70 A typ.
Power Factor (PF at 115 VAC) 0.99
Power Factor (PF at 230 VAC) 0.95
Power Factor Correction (PFC) Active PFC (Meets EN 61000-3-2 Class A)
No-Load Power Consumption < 0.5 W at 230 VAC (with Remote On/Off activated and no load on 5Vsb)
Mounting Style Chassis / PCB Standoff Mount (Open Frame, Class I Installation)
Cooling Method Natural Convection (201.6 W) or Forced Air Cooling at 1.5 m/s across terminals (252.0 W)
Package / Footprint Standard 3 x 5 Inch Industry Standard Open Frame
Overall Width (W) 76.2 mm (3.00 in)
Overall Height (H) 34.0 mm (1.34 in)
Overall Depth (D) 127.0 mm (5.00 in)
Dimensional Tolerance ±1.0 mm on outer L x W dimensions; ±0.5 mm on mounting hole pitch
Mounting Hole Pitch 115.6 x 64.8 mm (4.55 x 2.55 in)
Net Weight 350 g (0.77 lbs / 12.35 oz)
Termination / Connector Style JST VH Series PCB Headers (CN1 Input, CN51 Output) & Signal Header (CN61)
Input Terminal Pinout CN1 (JST B3P5-VH(LF)(SN), mates with VHR-5N housing): Live (L), Neutral (N), Earth Ground
Output & Signal Terminal Pinout CN51 (JST B6P-VH(LF)(SN), mates with VHR-6N): Main DC Output (+V / -V); CN61 (8-Pin Signal Header): Pin 1 STBY+ (5V+), Pin 2 STBY- (5V-), Pin 3-5 NC, Pin 6 DC_OK, Pin 7 R+ (Remote On/Off +), Pin 8 R- (Remote On/Off -)
Control & Monitoring Signals 5V Standby Output, DC_OK Signal, Remote On/Off (Apply voltage to isolated R+/R- terminals to shut unit down); Remote Sense & Parallel Operation: Not supported
Status Indicator Green Output Indicator LED on PCB (marked none on external chassis specification)
Overload Protection (OLP) Threshold > 14.7 A
Overvoltage Protection (OVP) Threshold 19.8 to 24.3 VDC (110% to 135% of nominal output)
Overvoltage Protection (OVP) Mode Latching shutdown (Cycle AC input power to reset)
Withstand Isolation Voltage (I/P to O/P) 4,000 VAC for 1 minute (2xMOPP)
Withstand Isolation Voltage (I/P to FG / Case) 2,000 VAC for 1 minute (1xMOPP)
Withstand Isolation Voltage (O/P to FG / Case) 1,500 VAC for 1 minute (1xMOPP)
Isolation Resistance (at 500 VDC) > 100 MΩ at 25 °C (77 °F) and 70% RH
Earth Leakage Current < 300 µA at 265 VAC, 60 Hz
Medical Isolation Rating 2xMOPP (Input to Output), 1xMOPP (Input to Ground & Output to Ground); Suitable for Type B and BF rated medical equipment
Operating Ambient Temperature Range -20 °C to +70 °C (-4 °F to +158 °F)
Full-Load Operating Temperature Limit +50 °C (+122 °F) for Convection Mounting A, B, E; +40 °C (+104 °F) for Convection Mounting C, D and Forced Air Cooling
High-Temperature Derating Rate Convection (Mounting A, B): Derate linearly 2.5%/°C from 100% load at +50 °C to 50% load at +70 °C; Convection (Mounting C, D): Derate from 100% at +40 °C to 80% at +50 °C and 40% at +70 °C; Forced Air: Derate from 100% at +40 °C to 50% at +60 °C
Low-Line Input Derating Derate linearly from 100% load at 115 VAC to 80% load at 85 VAC input
Altitude Derating Up to 5,000 m (16,404 ft) for IEC 62368-1; Up to 4,000 m (13,123 ft) for IEC 60601-1
Storage Temperature Range -40 °C to +85 °C (-40 °F to +185 °F)
Operating Relative Humidity 10% to 95% RH (non-condensing)
Storage Relative Humidity 10% to 95% RH (non-condensing)
Temperature Coefficient < 0.02% / °C
Vibration Resistance 10 to 55 Hz at 19.6 m/s² (2G) constant sweep 1 min along X, Y, Z axes for 1 hour (non-operating); Shock: < 196.1 m/s² (20G)
Safety Approvals (UL / North America) UL/CSA/ES 60601-1, UL/CSA 62368-1, UL 60950-1
Safety Approvals (IEC / EN / TÜV) IEC/EN 60601-1, IEC/EN 62368-1, EN 60950-1
Safety Approvals (Regional / Global) CE Mark, UKCA Mark (3-Year Manufacturer Warranty)
EMC Conducted & Radiated Emissions EN 55011 Class B, EN 55022 Class B, FCC Class B
EMC Harmonic & Flicker Compliance EN 61000-3-2 Class A
EMC Immunity Standards IEC 60601-1-2 Ed4, EN 61000-4-2 (ESD Level 2 & 3, Criteria A), EN 61000-4-3 (Radiated Level 3, Criteria A), EN 61000-4-4 (EFT Level 3, Criteria A), EN 61000-4-5 (Surge Level 3 & 4, Criteria A), EN 61000-4-6 (Conducted Level 3, Criteria A), EN 61000-4-8 (Magnetic Field Level 4, Criteria A), EN 61000-4-11 (Voltage Dips & Short Interruptions Class 3, Criteria B), SEMI F47 Line Dip (at >200 VAC)
MTBF (Telcordia SR-332 / Bellcore) 126,383 Hours (per RCR-9102B; Telcordia calculations yield significantly higher values)
Environmental Compliance RoHS Compliant, Lead-Free (LF) JST Connectors

Industrial Applications & Compatible Systems

The TDK-Lambda CUS200M-18 is engineered for medical, laboratory, and low-noise industrial systems requiring a compact 18 VDC power source with reinforced patient isolation:

  • Hospital & Patient-Care Medical Devices: Certified to IEC/EN/UL 60601-1 with 2xMOPP input-to-output isolation, 1xMOPP to ground, and <300 µA leakage current for Type B and BF patient-contact diagnostic monitors, infusion pumps, and surgical support carts.
  • Dental Chairs & Clinical Imaging Equipment: Operates silently at 201.6 W (11.2 A) via natural convection cooling without fan noise, making it ideal for chairside dental treatment units, optical diagnostics, and point-of-care clinical analyzers.
  • Broadcast & Professional Audio Consoles: Fanless convection capability coupled with Class B conducted and radiated EMI (EN 55011-B / EN 55022-B) eliminates acoustic hum and RF interference in studio mixing consoles and broadcast transmitters.
  • Automated Test & Semiconductor Equipment: Complies with IEC/UL 62368-1 and SEMI F47 line dip immunity (above 200 VAC), providing a stable 18 VDC bus with 5 V standby and DC_OK telemetry for industrial instrumentation.

Installation, Thermal Management & MRO Guidelines

Follow these engineering guidelines to ensure reliable thermal, electrical, and safety performance during integration:

  • Mounting Orientation & Thermal Derating: Mount in standard horizontal (Mounting A) or vertical (Mounting B) orientation to sustain 100% convection load (201.6 W) up to +50 °C (+122 °F), derating linearly to 50% load at +70 °C; to extract the full 252.0 W (14.0 A) rating, supply at least 1.5 m/s forced airflow across the terminals.
  • Low-Line AC Input & Breaker Sizing: Account for linear input voltage derating down to 80% load between 115 VAC and 85 VAC, and size upstream AC overcurrent protection to withstand the cold-start inrush surge of 35 A at 115 VAC and 70 A at 230 VAC.
  • JST Mating Harness & Class I Earthing: Mate input header CN1 (B3P5-VH) with a JST VHR-5N housing and output header CN51 (B6P-VH) with a VHR-6N housing, ensuring PCB lead protrusions beneath the board remain under 3 mm and the unit is bonded to protective earth for Class I EMC and safety compliance.