Description
TDK-Lambda CUS30M-18 (also known as CUS30M18) is a compact, 2″ × 3″ open-frame single-output AC-DC power supply from the CUS30M & CUS60M medical and ITE series, where the base part number designates the 30W convection-cooled platform, the -18 suffix specifies an 18 VDC nominal output, and the blank option suffix indicates an open-frame chassis fitted with industry-standard JST VH series connectors. Engineered for universal 85 to 264 VAC (47–63 Hz) operation without requiring external EMI filtering, this regulated switching converter delivers 1.7 A continuous output current (30.6 W rated power) with 87% typical efficiency at 115 VAC and 88% at 230 VAC, while maintaining <0.3 W no-load power draw at 265 VAC. Tight regulation is maintained at 72 mV line regulation, 144 mV load regulation, and 150 mV peak-to-peak ripple and noise across zero-to-full load conditions without any minimum load requirement.
Built for demanding clinical and industrial environments, the CUS30M-18 features reinforced 4 kVAC input-to-output isolation (2×MOPP), 2 kVAC input-to-ground (1×MOPP), and 1.5 kVAC output-to-ground (1×MOPP) suitable for Type B and BF patient-connected equipment in Class I or Class II (double-insulated) installations. Integrated protection circuits safeguard downstream electronics via >105% overcurrent limiting (UL/IEC Class 2 Limited Power Source) and >115% latching overvoltage shutdown requiring an AC input cycle to reset. Housed in a lightweight 62 g (2.19 oz) open-frame assembly measuring 76.2 × 50.8 × 24.2 mm (3.0 × 2.0 × 0.95 in), the module supports -30 °C cold start-up, full-load convection operation up to +45 °C (with linear derating to 50% load at +70 °C), and carries global IEC/EN/UL/ES/CSA 60601-1 3rd Edition, IEC 62368-1, EN 60335-1, and EN 60601-1-2:2015 Edition 4 approvals.
Key Features
- Medical 2×MOPP Reinforced Isolation: Certified to IEC/EN/UL/ES/CSA 60601-1 3rd Edition with 4 kVAC I/P-O/P and 1.5 kVAC O/P-FG isolation for Type B and BF medical equipment.
- Ultra-Compact 2″ × 3″ Convection Footprint: Measures 76.2 × 50.8 × 24.2 mm (3.0 × 2.0 × 0.95 in) and weighs just 62 g (2.19 oz), operating fanlessly for silent clinical environments.
- High Efficiency & Ultra-Low Standby Draw: Achieves 87% efficiency at 115 VAC and 88% at 230 VAC (87% average active efficiency) with <0.3 W no-load consumption at 265 VAC.
- Class I & Class II Built-In Class B EMI: Meets EN 55011-B, EN 55032-B, and FCC Class B conducted and radiated limits in grounded or double-insulated setups without external filters.
- Comprehensive Immunity & SEMI F47 Compliance: Tested to EN 60601-1-2:2015 Edition 4, Level 4 contact ESD (±8 kV), Level 4 common-mode surge, and SEMI F47 voltage dips above 200 VAC.
- Class 2 LPS & Latching Overvoltage Protection: Provides >105% overcurrent protection compliant with Class 2 Limited Power Source standards alongside >115% latching OVP shutdown.
- Wide Thermal & Altitude Operating Envelope: Supports -30 °C cold start-up, -20 °C to +70 °C ambient operation (full load to +45 °C), and deployment up to 5,000 m altitude.
Technical Specifications
| Manufacturer | TDK-Lambda |
|---|---|
| Manufacturer Part Number (MPN) | CUS30M-18 |
| Alternate Part Number | CUS30M18 |
| Product Series | CUS30M & CUS60M Series |
| Product Category | Medical Open Frame |
| Component Type | Single-Output Medical & ITE AC-DC Open-Frame Power Supply |
| Installation Class | Class I or Class II (Double Insulated) Construction |
| Nominal Output Voltage | 18 VDC |
| Rated Output Current | 1.7 A |
| Current Range | 0 to 1.7 A (No Minimum Load Required) |
| Rated Output Power | 30.6 W |
| Output Ripple & Noise (Max) | 150 mVp-p (Increases at light load to meet efficiency standards) |
| Line Regulation (Max) | 72 mV |
| Load Regulation (Max) | 144 mV |
| Minimum Load Requirement | 0 A (No minimum load required) |
| Temperature Coefficient | < 0.02 %/°C |
| Hold-Up Time (at 115 VAC) | 20 ms (Typical) |
| Hold-Up Time (at 230 VAC) | 100 ms (Typical) |
| Parallel Operation | Not Supported |
| AC Input Voltage Range | 85 to 264 VAC (Single-Phase) |
| Input Frequency Range | 47 to 63 Hz |
| Efficiency (at 115 VAC) | 87% (Typical, Full Load) |
| Efficiency (at 230 VAC) | 88% (Typical, Full Load) |
| Average Active Efficiency (at 115 VAC) | 87% (Typical, measured at 25%, 50%, 75%, and 100% load) |
| Average Active Efficiency (at 230 VAC) | 87% (Typical, measured at 25%, 50%, 75%, and 100% load) |
| AC Input Current (at 110 VAC) | 0.6 A |
| AC Input Current (at 230 VAC) | 0.4 A |
| Inrush Current (Cold Start at 115 VAC) | 30 A (Typical) |
| Inrush Current (Cold Start at 230 VAC) | 60 A (Typical) |
| No-Load Power Consumption | < 0.3 W (at 265 VAC) |
| Overload / Overcurrent Protection (OLP) Threshold | > 105% of Rated Current |
| Overload / Overcurrent Protection (OLP) Mode | Class 2 Limited Power Source (LPS) Compliant Current Limiting |
| Overvoltage Protection (OVP) Threshold | > 115% of Nominal Output Voltage (> 20.7 VDC) |
| Overvoltage Protection (OVP) Mode | Latching Shutdown (Cycle AC input to reset) |
| Withstand Isolation Voltage (I/P to O/P) | 4,000 VAC (4 kVAC) for 1 Minute (2×MOPP) |
| Withstand Isolation Voltage (I/P to FG / Ground) | 2,000 VAC (2 kVAC) for 1 Minute (1×MOPP) |
| Withstand Isolation Voltage (O/P to FG / Ground) | 1,500 VAC (1.5 kVAC) for 1 Minute (1×MOPP, Type B & BF Rated) |
| Isolation Resistance (at 500 VDC) | > 100 MΩ (Output to Ground at 25 °C, 70% RH) |
| Earth Leakage Current | < 0.25 mA (250 µA) at 265 VAC, 60 Hz |
| Patient / Enclosure Touch Current | 100 µA at 265 VAC, 60 Hz |
| Operating Ambient Temperature | -20 °C to +70 °C (-4 °F to +158 °F) with -30 °C (-22 °F) Cold Start-Up |
| Full-Load Operating Temperature Limit | +45 °C (+113 °F) for CUS30M-18 |
| High-Temperature Derating Rate | Linear derating to 50% load (15.3 W / 0.85 A) from +45 °C to +70 °C (2.0 %/°C) |
| Low-Line Input Voltage Derating | Linear derating to 80% load (24.48 W / 1.36 A) from 115 VAC down to 85 VAC |
| Storage Temperature Range | -40 °C to +85 °C (-40 °F to +185 °F) |
| Operating & Storage Humidity | 10% to 90% RH (Non-Condensing) |
| Maximum Operating Altitude | 5,000 m (16,404 ft); derate ambient temperature by 5 °C/1,000 m above 3,000 m |
| Vibration Resistance (Non-Operating) | 10 to 500 Hz, 19.6 m/s² (2G) constant sweep 1 min/cycle along X, Y, Z axes for 1 hour |
| Mechanical Shock Resistance | < 196.1 m/s² (< 20G) |
| MTBF (Telcordia SR-332 Issue 3) | 9,960,323 Hours (25 °C ambient, full load, 230 VAC input) |
| Medical Safety Certifications | IEC 60601-1, EN 60601-1, UL/ES 60601-1, CSA C22.2 No. 60601-1 (3rd Edition, 2×MOPP) |
| ITE & Household Safety Certifications | IEC/EN/UL/CSA 62368-1, 60950-1, EN 61558-1, EN 60335-1, Designed to meet IEC 61010-1 |
| Regional Compliance Marks | CE Mark, UKCA Mark |
| Conducted & Radiated EMI Emissions | EN 55011 Class B, EN 55032 Class B, FCC Part 15 Class B (Class I & Class II) |
| Harmonic Current Emissions | EN 61000-3-2 Compliant |
| Medical EMC Immunity Standard | EN 60601-1-2:2015 Edition 4 (Class 2 Voltage Dips Criteria B at 100–120 VAC, Criteria A at 200–240 VAC) |
| ESD Immunity (EN 61000-4-2) | Level 3 & Level 4 (Level 4 Contact Discharge Only), Performance Criteria A |
| Radiated RF Immunity (EN 61000-4-3) | Level 3, Performance Criteria A |
| EFT / Burst Immunity (EN 61000-4-4) | Level 3, Performance Criteria A |
| Surge Immunity (EN 61000-4-5) | Level 3 & Level 4 (Level 4 Common Mode Only), Performance Criteria A |
| Conducted RF Immunity (EN 61000-4-6) | Level 3, Performance Criteria A |
| Power Frequency Magnetic Immunity (EN 61000-4-8) | Level 4, Performance Criteria A |
| Voltage Dips & Short Interruptions (EN 61000-4-11) | Class 3 (Criteria B at 100–120 VAC; Criteria A at 200–240 VAC) |
| Semiconductor Sag Immunity (SEMI F47) | Compliant at Input Voltages > 200 VAC |
| Input Connector (CN1) | JST B2P3-VH(LF)(SN) 3-Position Header (Center Pin Removed; Mates with JST VHR-3N Housing) |
| Output Connector (CN-51) | JST B4P-VH(LF)(SN) 4-Position Header (Mates with JST VHR-4N Housing) |
| Cooling Method | Natural Convection Cooled (Fanless) |
| Form Factor & Mounting Style | Open-Frame PCB Assembly (2″ × 3″ Industry-Standard Footprint) |
| Overall Length (L) | 76.2 mm (3.00 in) |
| Overall Width (W) | 50.8 mm (2.00 in) |
| Overall Height (H) | 24.2 mm (0.95 in) |
| Overall Dimensions (L × W × H) | 76.2 × 50.8 × 24.2 mm (3.00 × 2.00 × 0.95 in) |
| Net Weight | 62 g (2.19 oz / 0.137 lbs) |
| Manufacturer Warranty | 3 Years |
Industrial Applications & Compatible Systems
TDK-Lambda CUS30M-18 is engineered for space-constrained medical, laboratory, and industrial equipment requiring silent convection cooling and 2×MOPP patient-grade isolation:
- Diagnostic & Patient-Connected Medical Devices: Infusion pumps, patient monitors, portable ultrasound systems, blood analyzers, and Type B / BF clinical instrumentation requiring 4 kVAC isolation and <100 µA touch current.
- Home Healthcare & Dental Equipment: CPAP respiratory therapy units, nebulizers, dental chairs, and surgical lighting where EN 60601-1-2 Edition 4 immunity and Class II double-insulated operation are mandatory.
- Test, Measurement & Laboratory Instruments: Benchtop analytical meters, spectrometers, and IEC 61010-1 compliant diagnostic analyzers sensitive to fan vibration and acoustic noise.
- Industrial Automation & Semiconductor Tooling: Embedded 18 VDC sensor controllers, human-machine interfaces (HMIs), and SEMI F47-compliant auxiliary control boards.
Installation, Thermal Management & MRO Guidelines
Installation & Engineering Guidelines: When integrating the open-frame CUS30M-18 into an end-product enclosure, maintain adequate primary-to-secondary creepage and clearance distances to preserve 2×MOPP reinforced isolation, particularly in Class II installations without a protective earth connection. Unlike 12 V, 15 V, and 24 V CUS30M variants rated to +50 °C, the CUS30M-18 maintains 100% output (30.6 W) up to +45 °C ambient in free-air convection and must be linearly derated at 2.0 %/°C down to 50% load (15.3 W) at +70 °C, as well as linearly derated to 80% load below 115 VAC down to 85 VAC. Use genuine JST VHR-3N (input CN1) and VHR-4N (output CN-51) crimp housings with appropriate SVH-21T-P1.1 contacts, and size upstream branch protection to withstand the 60 A cold-start inrush peak at 230 VAC.