GCS150PS12 XP Power 150W 12VDC Open Frame Medical PSU

$91.95

XP Power GCS150PS12 150W 12VDC open-frame AC-DC power supply, 85-264VAC input, 2xMOPP IEC 60601-1 & 62368-1 approved, 12V/0.6A fan rail, -40°C to +70°C.

SKU: GCS150PS12 Category: Brand:

Description

XP Power GCS150PS12 is a high-efficiency single-output open-frame AC-DC power supply from the GCS Series, engineered for dual industrial ITE and medical compliance. The MPN nomenclature decodes to the 150 W frame size (GCS150), single power output (PS), and 12.0 VDC primary rail (12) without optional covers or remote inhibit suffixes. Utilizing an active power-factor-corrected topology (PF greater than 0.9), it delivers 12.0 VDC at 9.2 A (110.4 W) under natural convection cooling and 12.5 A (150 W) with 7 CFM forced airflow, plus an auxiliary 12.0 VDC / 0.6 A fan supply for a 157 W maximum combined output. Operating across a universal 85–264 VAC input with 93% typical efficiency and under 0.5 W no-load standby consumption, it maintains ±0.5% line and load regulation with 1.5% peak-to-peak maximum ripple and noise.

Packaged in an industry-standard 3.00 × 5.00 × 1.42 in (76.2 × 127.0 × 36.3 mm) open-frame footprint weighing 0.29 kg (0.65 lbs), the unit interfaces via Molex header connectors (J1 AC input, J2 DC output, J3 fan) and a JST signal header (J4) with 0.5 V remote sense compensation and a 0.25 in Faston ground tab. Comprehensive protection includes dual-line F3.15 A / 250 V input fusing, latching overvoltage protection (110–140%), trip-and-restart overload protection (110–150%), and continuous short-circuit protection. Certified to IEC/UL/EN 60601-1 Edition 3 (2×MOPP primary-to-secondary, 4000 VAC isolation, 200 μA max leakage) and IEC/UL/EN 62368-1 for Class I and Class II installations, it operates from -40 °C to +70 °C (-40 °F to +158 °F) with EN 55011/32 Class B emissions.

Key Features

  • Dual Convection & Forced-Air Ratings: Delivers 12.0 VDC at 9.2 A (110.4 W) convection-cooled or 12.5 A (150 W) with 7 CFM airflow, plus a dedicated 12.0 VDC / 0.6 A auxiliary fan output.
  • Medical 2×MOPP & ITE Safety Approvals: Certified to IEC/UL/EN 60601-1 Edition 3 and IEC/UL/EN 62368-1 for Class I and Class II systems with 4000 VAC primary-to-secondary isolation and dual-line fusing.
  • Low Earth Leakage Current: Limits leakage current to 60 μA typical at 115 VAC / 50 Hz, 120 μA typical at 230 VAC / 50 Hz, and 200 μA maximum at 264 VAC / 60 Hz.
  • High Efficiency & Eco-Design Standby: Achieves 93% typical efficiency (80 Plus Bronze) with active power factor correction greater than 0.9 and less than 0.5 W no-load input power consumption.
  • Wide Industrial Operating Temperature: Operates reliably from -40 °C to +70 °C (-40 °F to +158 °F) at altitudes up to 5000 m, delivering full rated power up to +50 °C (+122 °F).
  • Native Class B EMC & High Immunity: Complies with EN 55011/32 Class B conducted and radiated emissions without external filtering, alongside EN 61204-3 and EN 60601-1-2 Level 3 immunity.
  • Standard 3 × 5 Inch Open-Frame Footprint: Compact 76.2 × 127.0 × 36.3 mm (3.00 × 5.00 × 1.42 in) chassis fits 1U enclosures and supports remote voltage sense up to 0.5 V drop.

Technical Specifications

Manufacturer XP Power
Manufacturer Part Number (MPN) GCS150PS12
Product Series GCS Series (GCS150)
Product Category Medical Open Frame
Component Type Open-Frame AC-DC Medical & Industrial Power Supply
Circuit Topology Active PFC + High-Frequency Switching Converter
Nominal Output Voltage 12.0 VDC (Primary V1); 12.0 VDC (Auxiliary Vfan)
Output Voltage Adjustment Range ±2% (11.76 to 12.24 VDC via onboard potentiometer; Vfan tracks V1)
Rated Output Current 9.2 A (Convection-cooled); 12.5 A (Forced-cooled, 7 CFM); 0.6 A (Vfan)
Output Current Range 0 to 9.2 A (Convection); 0 to 12.5 A (Forced-cooled at 7 CFM)
Rated Output Power 110.4 W (Convection-cooled); 150 W (Forced-cooled V1); 157 W (Combined Max with Vfan)
Number of Outputs 1 Main Output (V1) + 1 Auxiliary Fan Output (Vfan)
Output Ripple & Noise (Max.) 1.5% pk-pk (180 mVp-p max., 20 MHz bandwidth)
Output Voltage Tolerance ±1% (V1 initial set accuracy at 50% load, 115/230 VAC); ±5% (Vfan)
Line Regulation ±0.5% (90 to 264 VAC)
Load Regulation ±0.5% (V1, 0 to 100% load); ±5% (Vfan)
Transient Recovery Time & Deviation 4% max. deviation, recovery within 1% in less than 500 μs (50–75–50% load step)
Setup Time 2 s max. (Start-up delay at 115 / 230 VAC full load)
Hold-Up Time 20 ms typical (at 115 / 230 VAC)
AC Input Voltage Range 85 to 264 VAC (1-Phase; linear derating below 90 VAC)
Nominal Input Voltage 115 / 230 VAC
Input Frequency Range 47 to 63 Hz (50 / 60 Hz nominal)
Efficiency 93% typical (230 VAC full load; 80 Plus Bronze rated)
AC Input Current (at 115 VAC) 1.5 A typical (full load)
AC Input Current (at 230 VAC) 0.7 A typical (full load)
Inrush Current (Cold Start at 230 VAC) 80 A max. (cold start at 25 °C)
Power Factor (PF at 230 VAC) Greater than 0.9 (at 100% load)
Power Factor Correction (PFC) Built-in Active PFC (EN 61000-3-2 Class A / Class C above 50 W)
No-Load Power Consumption 0.5 W max.
Switching Frequency 60 to 200 kHz (PFC stage); 90 to 150 kHz (Main converter)
Mounting Style Chassis / Open-Frame PCB Mount (4 × Ø3.96 mm / 0.156 in mounting holes)
Cooling Method Natural Convection (up to 110.4 W) or Forced Air Cooling (7 CFM required above 110 W)
Package / Footprint Standard Industry-Standard 3.0 × 5.0 in Open-Frame (7.4 W/in³ power density)
Overall Width (W) 76.2 mm (3.00 in)
Overall Height (H) 36.3 mm (1.42 in)
Overall Depth (D) 127.0 mm (5.00 in)
Dimensional Tolerance ±0.50 mm (±0.02 in) for .xx; ±0.25 mm (±0.01 in) for .xxx
Mounting Hole Pitch 115.57 × 64.77 mm (4.55 × 2.55 in)
Net Weight 0.29 kg (0.65 lbs)
Termination / Connector Style PCB Pin Headers (Molex & JST) with 0.25 in Faston Ground Tab
Input Terminal Pinout J1 (Molex 09-65-2038, mates with 09-50-1031): Pin 1 Line, Pin 2 Neutral; Chassis Ground: 0.25 in Faston Tab
Output & Signal Terminal Pinout J2 Output (Molex 09-65-2068): Pins 1–3 +V1, Pins 4–6 RTN; J3 Fan (Molex 22-04-1021): Pin 1 Fan +(12V), Pin 2 Fan -; J4 Signal (JST B10B-PHDSS): Pin 1 +Sense, Pin 2 -Sense, Pins 3–10 Reserved/NC
Control & Monitoring Signals Remote Sense (compensates up to 0.5 V total cable drop); Remote On/Off Inhibit available on -R suffix variants only
Overload Protection (OLP) Threshold 110% to 150% of nominal output current (Inom)
Overload Protection (OLP) Mode Trip and Restart (Auto-recovery)
Overvoltage Protection (OVP) Threshold 110% to 140% of nominal output voltage (Vnom)
Overvoltage Protection (OVP) Mode Latching Shutdown (Recycle AC input power to reset)
Short Circuit Protection (SCP) Continuous (Auto-recovery upon fault removal)
Withstand Isolation Voltage (I/P to O/P) 4000 VAC (2 × MOPP)
Withstand Isolation Voltage (I/P to FG / Case) 1500 VAC (1 × MOPP)
Withstand Isolation Voltage (O/P to FG / Case) 1500 VAC (1 × MOPP)
Earth Leakage Current 60 μA typ. (115 VAC / 50 Hz); 120 μA typ. (230 VAC / 50 Hz); 200 μA max. (264 VAC / 60 Hz)
Medical Isolation Rating 2 × MOPP (Primary to Secondary), 1 × MOPP (Primary/Secondary to Earth), Class I & Class II
Operating Ambient Temperature Range -40 °C to +70 °C (-40 °F to +158 °F)
Full-Load Operating Temperature Limit +50 °C (+122 °F) at 100% rated load
High-Temperature Derating Rate Linear derating from 100% load at +50 °C (+122 °F) to 50% load at +70 °C (+158 °F) (2.5%/°C)
Low-Line Input Derating Linear derating below 90 VAC down to 85 VAC (approx. 90% load at 85 VAC)
Altitude Derating Rated up to 5000 m (16,404 ft) maximum operating altitude
Storage Temperature Range -40 °C to +85 °C (-40 °F to +185 °F)
Operating Relative Humidity 5% to 95% RH (Non-condensing)
Temperature Coefficient ±0.05%/°C
Vibration Resistance 2g sweep and endurance at resonance, 10 to 500 Hz, 3 axes (EN 60068-2-6)
Safety Approvals (UL / North America) UL 62368-1, CAN/CSA C22.2 No. 62368-1-14 (ITE); ANSI/AAMI ES60601-1:2005, CSA C22.2 No. 60601-1:08 (Medical)
Safety Approvals (IEC / EN / TÜV) IEC 62368-1:2014, IEC 60950-1 Ed 2, EN 62368-1:2014/A11:2017 (ITE); IEC 60601-1 Ed 3, EN 60601-1/A12:2006 (Medical)
Safety Approvals (Regional / Global) CE (LVD & RoHS), CB Scheme Report (ITE & Medical)
EMC Conducted & Radiated Emissions EN 55011 / EN 55032 Class B (Conducted & Radiated)
EMC Harmonic & Flicker Compliance EN 61000-3-2 Class A (Class C above 50 W), EN 61000-3-3
EMC Immunity Standards EN 61204-3 High Severity, EN 55035, EN 60601-1-2; EN 61000-4-3 (Radiated Level 3), EN 61000-4-4 (EFT Level 3), EN 61000-4-5 (Surge Installation Class 3), EN 61000-4-6 (Conducted Level 3), Voltage Dips & Interruptions
MTBF (MIL-HDBK-217F) 569,000 hours (MIL-HDBK-217F Notice 2, +25 °C, Ground Benign)
Environmental Compliance RoHS Compliant

Industrial Applications & Compatible Systems

The XP Power GCS150PS12 is engineered for space-constrained medical, laboratory, and industrial automation systems requiring dual 2×MOPP patient safety isolation and low conducted/radiated EMI.

  • Medical Diagnostic & Patient Monitoring Systems: Powers clinical analyzers, infusion pumps, ultrasound peripherals, and bedside monitors demanding IEC 60601-1 Edition 3 compliance and under 200 μA earth leakage.
  • 1U Rackmount Test & Measurement Instrumentation: Fits compact 1U (44.45 mm) chassis enclosures in automated test equipment (ATE), data acquisition units, and RF benchtop instruments.
  • Industrial Process Control & Robotics: Supplies clean 12 VDC bus power for industrial HMIs, vision inspection cameras, solenoid valve manifolds, and embedded edge controllers.
  • Class II Double-Insulated Equipment: Suitable for portable diagnostic carts and ungrounded enclosures where Class II isolation eliminates protective earth reliance.

Installation, Thermal Management & MRO Guidelines

Proper mechanical grounding, airflow routing, and connector selection ensure full 150 W reliability and regulatory compliance in end-use assemblies.

  • Airflow & Thermal Management: Maintain at least 7 CFM of forced airflow across the PCB when continuous load exceeds 110.4 W (9.2 A), and ensure hot-spot temperatures on magnetic coils T1/L3 remain below 120 °C and capacitors C6/C23 remain below 105 °C.
  • Chassis Grounding for Class B EMI: Bond all four mounting hole pads (marked ‘A’ on the mechanical drawing) together to the metal chassis or ground plane and connect the 0.25 in Faston tab to achieve optimal EN 55011/32 Class B radiated emissions.
  • Mating Connector & Sense Wiring: Use Molex 09-50-1031 (J1 AC) and 09-50-1061 (J2 DC) housings with series 5194 crimp terminals, and route twisted-pair sense leads from JST PHDR-10VS (J4 Pins 1–2) to the point of load to compensate up to 0.5 V drop.
  • Upstream Overcurrent & Inrush Sizing: Although dual internal F3.15 A / 250 V fuses are built into both Line and Neutral, select slow-blow (Type C or D) upstream breakers capable of withstanding the 80 A cold-start inrush spike at 230 VAC.