Description
TDK-Lambda XMS500DP is a configurable 500W Class II double-insulated AC-DC enclosed power supply from the XMS500 series, decoded under OEM configuration string XMS500D-24-5H-TFSBDInA. Operating across a universal 90–264 VAC (47–63 Hz) input with active power factor correction (0.95 minimum at 110 VAC) and up to 92% conversion efficiency at 230 VAC, it delivers a regulated 24 VDC main output at 20.8 A alongside an isolated 5 V / 2 A (5H) auxiliary standby rail. Integrated factory-fitted top fan cooling (TF) on a U-chassis (S) enables full 500 W continuous output without requiring external system airflow, while maintaining tight ±1% initial voltage accuracy, less than 2% total regulation, 1% pk-pk ripple and noise, and 16 ms minimum hold-up time.
Engineered for patient-connected and body-floating (BF) medical equipment, the TDK-Lambda XMS500DP provides reinforced 2xMOPP input-to-output galvanic isolation rated at 4 kVAC (5.7 kVDC) and ultra-low touch current below 100 µA without requiring a protective earth connection. Comprehensive circuit safeguards include auto-recovering hiccup-mode short-circuit protection, latching overvoltage and overtemperature protection, remote on/off inhibit control (In), and an active AC Fail status signal (A). Housed in a compact 107.0 × 67.4 × 180.0 mm (4.21 × 2.65 × 7.09 in) metal chassis, it operates at full load from 0 °C to +50 °C (derating linearly at 2.5%/°C up to +70 °C) and carries full IEC/EN/UL 60601-1, IEC/EN/UL 62368-1, CE, and UKCA certifications with native CISPR 11/32 Class B emissions compliance.
Key Features
- IEC 60601-1 2xMOPP Reinforced Isolation: Delivers 4 kVAC (5.7 kVDC) input-to-output galvanic isolation and less than 100 µA touch current for Class II double-insulated BF medical devices.
- 500W Continuous 24VDC Output with Standby: Supplies 24 VDC at 20.8 A with 1% pk-pk ripple, ±1% setting accuracy, and an always-on 5 V / 2 A auxiliary standby output.
- Integrated Top-Fan U-Chassis Cooling: Factory-mounted top cooling fan (TF) provides self-contained forced-air thermal management for full 500 W continuous load from 0 °C to +50 °C.
- Universal AC Input & High Efficiency: Accepts 90–264 VAC (300 VAC surge withstand for 5 s) at 47–63 Hz, achieving up to 92% efficiency at 230 VAC and 0.95 minimum power factor.
- Comprehensive Protection & Supervisory Signals: Features hiccup short-circuit recovery, latching OVP and OTP, remote on/off inhibit logic (In), and dedicated AC Fail monitoring (A).
- Native Class B Conducted & Radiated EMC: Meets EN 55011, EN 55032, and FCC Part 15 Subpart B Class B limits in Class II configuration without external EMI filters.
- Rugged Shock, Vibration & High-Altitude Rating: Certified to MIL-STD-810G 30 g shock and 2 g vibration, operating up to 3,000 m for medical and 5,000 m for industrial installations.
Technical Specifications
| Manufacturer | TDK-Lambda |
|---|---|
| Manufacturer Part Number (MPN) | XMS500DP |
| OEM Product Configuration Code | XMS500D-24-5H-TFSBDInA |
| Product Series | XMS500 Series |
| Product Category | Open Frame PSUs |
| Protection Class | Class II (Double Insulated, No Protective Earth) |
| Nominal Main Output Voltage | 24 VDC |
| Rated Main Output Current | 20.8 A |
| Rated Continuous Output Power | 500 W (with factory-fitted top fan) |
| Auxiliary Standby Output (Option 5H) | 5 VDC / 2 A (10 W) |
| Internal Fan Supply (Option TF) | Dedicated to factory-fitted top fan (not for external customer load) |
| Minimum Load Requirement | 0 A (No minimum load required) |
| Voltage Setting Accuracy | ±1% (measured at 50% load) |
| Total Output Regulation | < 2% (combined line, load, and thermal regulation) |
| Line Regulation | 0.1% (across 90–264 VAC input change) |
| Load Regulation | 1.0% (across 0–100% load change) |
| Thermal Regulation Coefficient | 0.02%/°C (across 0 °C to +50 °C) |
| Output Ripple & Noise (20 MHz BW) | 1% pk-pk (240 mVp-p max, EIAJ test method) |
| Transient Response Voltage Deviation | < 5% of set voltage (25% to 100% step load change) |
| Transient Response Recovery Time | < 2 ms (to within 2% of set voltage) |
| Turn-On / Setup Time (at 90 VAC) | 1.0 s max (at 90 VAC input, 100% rated output power) |
| Hold-Up Time (at 100% Load) | 16 ms minimum (at nominal input, 100% rated load) |
| AC Input Voltage Range | 90–264 VAC (100–240 VAC nominal) |
| Input Surge Voltage Withstand | 300 VAC for up to 5 seconds |
| Input Line Frequency | 47–63 Hz |
| Efficiency (at 230 VAC) | Up to 92% (at 230 VAC, 100% rated load) |
| Power Factor (PF at 110 VAC) | 0.95 minimum (at 110 VAC input) |
| Inrush Current (Cold Start at 264 VAC) | < 40 A (at +25 °C ambient, 264 VAC cold start) |
| Input Fusing Configuration | Dual fast-acting internal fuses (non-user replaceable) |
| Short-Circuit Protection (SCP) Mode | Hiccup mode with automatic recovery upon fault removal |
| Overvoltage Protection (OVP) Mode | Latching shutdown (requires AC power cycle to reset) |
| Overtemperature Protection (OTP) Mode | Latching shutdown (requires AC power cycle to reset) |
| Withstand Isolation Voltage (I/P to O/P) | 4,000 VAC / 5,700 VDC (Reinforced, 2 x MOPP) |
| Medical Means of Patient Protection | 2 x MOPP (Input to Output, BF-ready) |
| Touch Current (Option B) | < 100 µA (measured at 264 VAC, 63 Hz) |
| Earth Leakage Current (Option D) | Class II construction (no protective earth connection) |
| Operating Ambient Temperature Range | 0 °C to +70 °C (+32 °F to +158 °F) with derating |
| Full-Load Operating Temperature Limit | 0 °C to +50 °C (+32 °F to +122 °F) at 100% load (500 W) with top fan |
| High-Temperature Derating Rate | 2.5%/°C linear derating from +50 °C to +70 °C (50% load at +70 °C) |
| Low-Temperature Cold Startup | -20 °C (-4 °F) |
| Storage Temperature Range | -40 °C to +70 °C (-40 °F to +158 °F) for Top Fan (TF) models |
| Operating Relative Humidity | 5% to 95% RH (non-condensing) |
| Operating & Storage Altitude | -200 m to 3,000 m (IEC 60601-1 medical) / -200 m to 5,000 m (IEC 62368-1 industrial) |
| Pollution Degree & Material Group | Pollution Degree 2, Material Group III |
| Mechanical Shock Resistance | ±3 x 30 g half-sine shocks (11 ms ± 0.5 ms) in each plane (18 total); IEC 60068-2-27 & MIL-STD-810G Method 516.6 Pro IV, VI |
| Mechanical Vibration Resistance | 2 g single-axis sweep & endurance at resonance, 10–200 Hz in all 3 planes; IEC 60068-2-6 & MIL-STD-810G Method 514.4 Pro I, Cat 1, 9 |
| Medical Safety Approvals | IEC/EN 60601-1, UL/CSA 60601-1, ANSI/AAMI ES60601-1, CAN/CSA-C22.2 No. 60601-1-08 (UL File E349607) |
| ITE / Audio-Video Safety Approvals | IEC/EN 62368-1, UL 62368-1, CSA C22.2 No. 62368-1, IEC/EN/UL/CSA 60950-1 (UL File E135494) |
| Additional Designed-to-Meet Safety Standards | IEC/EN 61010-1 (Measurement/Lab), IEC/EN/UL/CSA 60065 |
| Regional Compliance Marks & Quality System | CE (LVD/EMC/RoHS), UKCA, CB Scheme; manufactured under ISO 9001 & ISO 13485 |
| Conducted & Radiated Emissions | EN 55011, EN 55032 (CISPR 11/22) Class B, FCC 47 Part 15 Subpart B Class B, EN 61000-6-3 |
| Harmonic Current Emissions | EN 61000-3-2 Class A |
| Voltage Fluctuations & Flicker | EN 61000-3-3 Compliant (dmax only) |
| ESD Immunity (EN 61000-4-2) | Level 4 (in chassis) / Level 3 (I/O & signal pins), Performance Criteria A |
| Radiated RF Field Immunity (EN 61000-4-3) | Level 3, Performance Criteria A |
| EFT / Burst Transient Immunity (EN 61000-4-4) | Level 4, Performance Criteria A |
| Surge Immunity (EN 61000-4-5) | Level 3, Performance Criteria A |
| Conducted RF Immunity (EN 61000-4-6) | Level 3, Performance Criteria A |
| Power Frequency Magnetic Field (EN 61000-4-8) | Level 3, Performance Criteria A |
| Voltage Dips & Interruptions (EN 61000-4-11) | Class 3, Criteria A (Criteria B for 1-cycle and 5-second interruptions) |
| Ring Wave Immunity (EN 61000-4-12) | Level 3, Performance Criteria B |
| Voltage Fluctuations Immunity (EN 61000-4-14) | Class 3, Performance Criteria A |
| Remote On/Off Control (Option In) | Inhibit logic (J2 Pin 5): Connect to Standby + (2 mA max) to inhibit Channel 1; open circuit or 0 V to enable |
| AC Fail Supervisory Signal (Option A) | J2 Pin 6: Logic High = AC Fail; 0 V = AC Good |
| AC Input Connector (J1) & Pinout | 3-Pin Class II Header (Mates with Molex 09-50-1031 housing / 08-70-1030 pins): Pin 1 = Live, Pin 2 = NC, Pin 3 = Neutral |
| Control & Standby Signal Connector (J2) & Pinout | 8-Pin Header (Mates with Molex 90142-0008 housing / 90119-0109 pins): Pin 1 = Standby +, Pin 2 = 0V, Pin 3 = Standby +, Pin 4 = 0V, Pin 5 = Remote On/Off, Pin 6 = AC Fail, Pin 7 = Temp Sense, Pin 8 = Temp Sense |
| Internal Fan Connector (J3) & Pinout | 2-Pin Header (Mates with Molex 51191-0200 housing / 50802-9001 pins): Pin 1 = 0V, Pin 2 = Fan + (factory-wired to top fan) |
| Cooling Method | Forced-air cooling via integrated factory-mounted top fan (Option TF) |
| Chassis & Enclosure Style | U-Chassis with Top Fan Cover (Option S + TF) |
| Overall Width (W) | 107.00 mm (4.21 in) |
| Overall Height (H) | 67.40 mm (2.65 in) |
| Overall Length / Depth (L/D) | 180.00 mm (7.09 in) |
| Overall Dimensions (W x H x L) | 107.00 x 67.40 x 180.00 mm (4.21 x 2.65 x 7.09 in), ±0.5 mm general tolerance |
| Chassis Mounting Provisions & Torque | Bottom & side M3 threaded mounting holes (max penetration 3.0 mm bottom / 2.4 mm side); max tightening torque 0.5–0.6 Nm |
| Warranty | 5 Years OEM Manufacturer Warranty |
Industrial Applications & Compatible Systems
TDK-Lambda XMS500DP is engineered for mission-critical medical, clinical, and precision instrumentation systems requiring Class II double insulation, 2xMOPP isolation, and self-contained forced-air cooling:
- BF-Rated Patient-Connected Medical Devices: Surgical consoles, patient monitors, anesthesia workstations, and defibrillator-proof clinical equipment demanding 4 kVAC (2xMOPP) isolation and under 100 µA touch current.
- Home Healthcare & Class II Diagnostic Systems: Portable ultrasound scanners, infusion systems, respiratory therapy units, and dental chairs operating from ungrounded 2-prong wall outlets.
- In-Vitro Diagnostic (IVD) & Laboratory Analyzers: Automated blood chemistry analyzers, PCR thermal cyclers, and IEC 61010-1 analytical instruments utilizing the 5 V / 2 A standby rail for system wake-up and supervisory control.
- Broadcast, Test & Measurement Equipment: Low-noise studio audio/video processors, RF test benches, and industrial automation controllers benefiting from native CISPR 11/32 Class B emissions compliance without external line filters.
Installation, Thermal Management & MRO Guidelines
Installation & Engineering Guidelines: Mount the TDK-Lambda XMS500DP using M3 fasteners into the base (3.0 mm maximum screw penetration) or side chassis holes (2.4 mm maximum screw penetration) tightened to 0.5–0.6 Nm, ensuring unobstructed intake and exhaust clearance above the top-mounted cooling fan. For ambient temperatures exceeding +50 °C, apply a linear output power derating of 2.5%/°C down to 250 W at the +70 °C upper operating limit. Because this is a Class II double-insulated unit, use the 3-position Molex 09-50-1031 mating connector on J1 with Live and Neutral only (do not connect a protective earth conductor to the chassis), and size upstream branch-circuit protection to accommodate the < 40 A cold-start inrush current at 264 VAC.