Description
Delta Electronics MDS-065APS18 BA (also known as MDS065APS18BA) is a compact 65W single-output medical open-frame AC-DC power supply from the Delta MDS Series. Per Delta MPN nomenclature, MDS denotes Medical Power Supply, 065 specifies the 65W maximum series wattage, APS identifies the open-frame single-output family code, 18 indicates the regulated 18VDC output rail, and BA signifies the standard Delta configuration. Engineered with a universal 90–264VAC (47–63Hz) input stage and optoisolated PWM flyback architecture, it delivers 3.61A continuous current with a 4.98A peak load capability for up to 8 seconds (10% duty cycle). The supply achieves 87.7% typical conversion efficiency, limits no-load standby power draw to 0.3W maximum, and maintains ±1.5% total voltage regulation with 20.8mV pk-pk typical ripple and noise.
Built for patient-connected and clinical instrumentation, the MDS-065APS18 BA integrates 2×MOPP (Means of Patient Protection) galvanic isolation rated at 4000VAC input-to-output, 1500VAC input-to-ground, and 500VAC output-to-ground, alongside ultra-low touch current below 0.1mA in normal conditions and 0.3mA in single-fault conditions. Comprehensive protection circuitry covers latching Overvoltage Protection (150% max) plus auto-recovering hiccup-mode Overload/Overcurrent (150%–210%), Overtemperature, and Short-Circuit protection. Housed on an industry-standard 2.0 × 4.0 × 1.2 in (50.8 × 101.6 × 30.0 mm) open-frame PCB weighing 130g (4.6 oz) with JST VH input/output headers, it operates across -10°C to +70°C (full load up to +50°C) and carries global IEC/EN/UL/CSA 60601-1, IEC 62368-1, and IEC 60601-1-2 4th Edition EMC certifications.
Key Features
- Medical & ITE Dual Safety Certifications: Approved to IEC/EN/ANSI/AAMI ES 60601-1, CAN/CSA-C22.2 No. 60601-1, IEC 62368-1, and IEC 60950-1 with CB, UL/cUL, CE, UKCA, and CQC marks.
- 2×MOPP Isolation & Ultra-Low Touch Current: Provides 4000VAC input-to-output galvanic isolation and limits touch current to 0.1mA maximum (normal) and 0.3mA maximum (single fault) at 264VAC.
- 18VDC Regulated Output with Peak Load Surge: Delivers 3.61A (65W) continuous output with ±1.5% total regulation, 20.8mV pk-pk ripple, and 4.98A peak output capability for 8 seconds.
- High Efficiency & Green Standby Power: Reaches 87.7% typical conversion efficiency across universal 90–264VAC input while keeping no-load standby power consumption under 0.3W.
- IEC 60601-1-2 4th Edition EMC Immunity: Complies with EN 55011 / EN 55032 / FCC Class B conducted and radiated emissions plus 15kV air ESD and 28V/m radiated RF immunity.
- Full Fault Protection Suite: Features latching Overvoltage Protection (150% max) and auto-recovery hiccup-mode Overload (150%–210%), Overtemperature, and Short-Circuit protection.
- Compact 2×4-Inch Footprint & High Reliability: Measures 50.8 × 101.6 × 30.0 mm (2.0 × 4.0 × 1.2 in) at 130g with JST VH connectors and achieves 2.5 million hours MTBF (Telcordia SR-332).
Technical Specifications
| Manufacturer | Delta |
|---|---|
| Manufacturer Part Number (MPN) | MDS-065APS18 BA |
| Alternate Part Number | MDS065APS18BA |
| Product Series | MDS Series |
| Product Category | Medical Open Frame |
| Component Type | Medical AC-DC Open Frame Power Supply |
| Circuit Topology | Isolated Switching Mode Power Supply (Optoisolated PWM Flyback) |
| Number of Outputs | 1 (Single Output) |
| Nominal Output Voltage | 18 VDC |
| Rated Output Current | 3.61 A |
| Rated Output Power | 65 W |
| Peak Output Current | 4.98 A (8 seconds maximum duration, 10% duty cycle, within ±5% regulation) |
| Total Output Voltage Regulation | ±1.5% |
| Line Regulation (Max) | ±0.5% |
| Load Regulation (Max) | ±1.0% |
| Output Ripple & Noise (Typ) | 20.8 mV pk-pk (at full load, 20 MHz bandwidth with 0.47 µF ceramic + 10 µF electrolytic capacitors) |
| Dynamic Load Response | ±5% output voltage overshoot/undershoot (50% to 100% load step) |
| Start-Up Time (at 115 VAC) | 3000 ms max (time to reach 90% of nominal Vo) |
| Hold-Up Time (at 115 VAC) | 15 ms typ at 65 W full load (20 ms typ at 100 VAC with 45 W load) |
| Nominal Input Voltage | 100 to 240 VAC |
| Input Voltage Range (AC) | 90 to 264 VAC (Single-Phase) |
| Nominal Input Frequency | 50 / 60 Hz |
| Input Frequency Range | 47 to 63 Hz |
| Efficiency (Typical) | 87.7% typ (at 25°C ambient) |
| AC Input Current (at 100 VAC) | 1.5 A max |
| AC Input Current (at 240 VAC) | 0.75 A max |
| Inrush Current (Cold Start at 115 VAC) | 40 A typ |
| Inrush Current (Cold Start at 230 VAC) | 80 A typ |
| No-Load / Standby Power Consumption | 0.3 W max |
| Overload / Overcurrent Protection (OLP/OCP) Threshold | 150% to 210% of rated output load current |
| Overload / Overcurrent Protection (OLP/OCP) Mode | Hiccup Mode (Non-Latching, Auto-Recovery after fault removal) |
| Overvoltage Protection (OVP) Threshold | 150% max of nominal output voltage |
| Overvoltage Protection (OVP) Mode | Latch Mode (Requires AC input power recycle to restart) |
| Overtemperature Protection (OTP) Mode | Hiccup Mode (Non-Latching, Auto-Recovery after cooling and AC recycle) |
| Short-Circuit Protection (SCP) Mode | Hiccup Mode (Non-Latching, Auto-Recovery) |
| Protection Against Electric Shock | Class I (with Protective Earth PE connection) / 2 × MOPP (Means of Patient Protection) |
| Withstand Isolation Voltage (Input to Output) | 4000 VAC (2 × MOPP) |
| Withstand Isolation Voltage (Input to Ground / PE) | 1500 VAC |
| Withstand Isolation Voltage (Output to Ground / PE) | 500 VAC |
| Touch Current (Normal Condition) | 0.1 mA max (at 264 VAC) |
| Touch Current (Single Fault Condition) | 0.3 mA max (at 264 VAC) |
| Operating Ambient Temperature Range | -10°C to +70°C (+14°F to +158°F) |
| Full-Load Operating Temperature Limit | -10°C to +50°C (+14°F to +122°F) at 100% rated load (65 W) |
| High-Temperature Derating Rate | 2.5% / °C linear derating from +50°C to +70°C (50% load / 32.5 W at +70°C) |
| Storage Temperature Range | -40°C to +85°C (-40°F to +185°F) |
| Operating & Storage Relative Humidity | 5% to 95% RH (Non-Condensing) |
| Operating Altitude | Up to 3,000 m (9,842 ft / 106 to 70 kPa) |
| Non-Operating Shock Resistance | 50 G, 11 ms half-sine, 3 shocks per axis/direction |
| Non-Operating Vibration Resistance | 5 to 500 Hz, 2.09 Grms, 20 minutes per each of 3 axes |
| MTBF (Telcordia SR-332) | 2,500,000 hours (2.5 Million Hours typical) |
| Medical Safety Standards | IEC 60601-1, EN 60601-1, ANSI/AAMI ES 60601-1, CAN/CSA-C22.2 No. 60601-1 |
| ITE / Audio-Video Safety Standards | IEC 62368-1, IEC 60950-1 |
| Regional Compliance & Marks | CB Certified, UL / cUL, CE (EMC 2014/30/EU & LVD 2014/35/EU), UKCA (UK MDR 2002), CQC (GB 17625.1, GB 4943.1, GB/T 9254.1), ISO 13485:2016 |
| EMC Compliance Standard | IEC 60601-1-2 4th Edition & EN 60601-1-2:2015 |
| Conducted & Radiated EMI Emissions | EN/BS EN 55011 Class B, EN/BS EN 55032 Class B, FCC Title 47 Part 15 Class B |
| Harmonic Current Emissions (IEC 61000-3-2) | Meets IEC 61000-3-2 Class A Limit |
| Voltage Fluctuations & Flicker (IEC 61000-3-3) | Complies with IEC 61000-3-3 |
| ESD Immunity (IEC 61000-4-2) | Level 4, Criteria A: ±15 kV Air Discharge, ±8 kV Contact Discharge |
| Radiated RF Field Immunity (IEC 61000-4-3) | Criteria A: 10 V/m AM (80 MHz–2700 MHz); 28 V/m Pulse/Modulated (385 MHz–5785 MHz) |
| Electrical Fast Transient / Burst Immunity (IEC 61000-4-4) | Level 3, Criteria A: ±2 kV |
| Surge Immunity (IEC 61000-4-5) | Level 3, Criteria A: ±2 kV Common Mode (Line-to-Earth), ±1 kV Differential Mode (Line-to-Line) |
| Conducted RF Immunity (IEC 61000-4-6) | Level 2, Criteria A: 3 Vrms (150 kHz–80 MHz), 6 Vrms at ISM & Amateur Radio Bands |
| Power Frequency Magnetic Field Immunity (IEC 61000-4-8) | Criteria A: 30 A/m |
| Voltage Dips & Interruptions Immunity (IEC 61000-4-11) | 0% UT 0.5 cycle (10ms) Crit A; 0% UT 1 cycle (20ms) Crit B; 70% UT 25 cycles (500ms) Crit A; 0% UT 250 cycles (5000ms) Crit B |
| AC Input Connector (CN1) | JST B2P3-VH(LF)(SN) 2-Pin Header (White, mates with JST VHR-3N housing) |
| DC Output Connector (CN2) | JST B4P-VH(LF)(SN) 4-Pin Header (White, mates with JST VHR-4N housing) |
| Cooling Method | Free Air Convection |
| Mounting Method | Chassis / Standoff PCB Mounting (4 mounting holes; 2 PE ground mounting holes) |
| Mounting Hole Pitch | 95.3 ± 0.5 mm (3.75 ± 0.02 in) length pitch |
| Overall Width (W) | 50.8 mm (2.00 in) |
| Overall Length (L) | 101.6 ± 0.5 mm (4.00 ± 0.02 in) |
| Overall Height (H) | 30.0 mm (1.20 in) |
| Overall Dimensions (W × L × H) | 50.8 × 101.6 × 30.0 mm (2.0 × 4.0 × 1.2 in) |
| Net Weight | 130 g (4.6 oz) |
| Environmental & Hazardous Substance Compliance | RoHS Directive Compliant |
| Warranty | 3 Years Manufacturer Warranty |
Industrial Applications & Compatible Systems
Delta Electronics MDS-065APS18 BA is engineered for space-constrained medical devices, diagnostic analyzers, and Class I clinical instrumentation requiring 2×MOPP patient isolation and low leakage current:
- Patient Monitoring & Bedside Diagnostics: Powers vital signs monitors, infusion pumps, pulse oximeters, and portable ECG/EEG bedside units needing 18VDC at up to 3.61A with 4000VAC 2×MOPP isolation.
- In-Vitro Diagnostic (IVD) & Laboratory Analyzers: Supplies low-noise 18VDC rails (20.8mV pk-pk ripple) for automated hematology analyzers, immunoassay readers, spectrophotometers, and PCR thermal cyclers.
- Dental, Surgical & Respiratory Therapy Equipment: Supports dental imaging sensors, CPAP/BiPAP respiratory humidifiers, endoscopic light sources, and surgical suction consoles compliant with IEC 60601-1-2 4th Edition EMC.
- Medical Carts & Point-of-Care Computing: Integrates into mobile nursing workstations, medical-grade panel PCs, and battery-charging cradles requiring compact 2 × 4-inch open-frame footprints.
Installation, Thermal Management & MRO Guidelines
Installation & Engineering Guidelines: For Class I safety compliance and optimal EMI suppression, bond both PCB mounting holes marked with the ground symbol to the host system protective earth (PE) chassis using conductive metal standoffs and screws. Maintain full 65W (3.61A) continuous output in ambient temperatures from -10°C to +50°C with free-air convection; above +50°C, derate output power linearly at 2.5%/°C down to 32.5W (50%) at +70°C. Use mating JST VHR-3N (AC input CN1) and JST VHR-4N (DC output CN2) crimp housings, and size upstream AC branch protection to withstand the 80A typical cold-start inrush current at 230VAC.