CUS800M-48 TDK-Lambda 800W 48VDC Medical Power Supply

$267.81

TDK-Lambda CUS800M-48 (CUS800M48) 801.6W 48VDC 16.7A medical enclosed AC-DC power supply, 85–265VAC input, 94% efficiency, 2xMOPP IEC 60601-1 &

SKU: CUS800M-48 Category: Brand:

Description

TDK-Lambda CUS800M-48 (also known as CUS800M48) is an 801.6W enclosed medical and industrial AC-DC power supply from the ultra-compact, low-profile CUS800M series, where the 800M prefix designates the 800W medical-grade chassis platform and the -48 suffix specifies a regulated 48VDC single output. Built around a high-efficiency LLC resonant primary-to-secondary converter operating at 50 kHz to 300 kHz with active power factor correction (0.99 at 115 VAC, 0.95 at 230 VAC), it accepts a wide universal input from 85 to 265 VAC (47–63 Hz) and delivers 16.7A continuous current across an adjustable 46.8 to 51.8 VDC trim range. Conversion efficiency reaches 92.2% at 115 VAC and 94.0% at 230 VAC, while maintaining tight line regulation within 240 mV, load regulation within 480 mV, ripple and noise below 480 mVp-p, and less than 0.6 W no-load power consumption when inhibited.

Engineered for Type B and BF patient-connected clinical systems as well as space-constrained instrumentation, the CUS800M-48 provides full 2xMOPP galvanic isolation rated at 4,000 VAC (input-to-output), 2,000 VAC (input-to-ground, 1xMOPP), and 1,500 VAC (output-to-ground, 1xMOPP), alongside ultra-low earth leakage current under 250 µA at 265 VAC. Comprehensive circuit protections include hiccup-mode overcurrent protection above 17.6A, overvoltage protection tripping between 55.2 V and 64.8 V, and latching overtemperature shutdown, complemented by an always-on 5 V / 2 A standby rail, 0.5 V total remote sense compensation, opto-isolated remote on/off, and an opto-isolated Power Good (AC Fail / DC OK) signal. Housed in a 1.67-inch (42.5 mm) high chassis measuring 171 × 85 × 42.5 mm (810 g) with an acoustic-optimized variable-speed cooling fan (30 dBA at 20% load, 42 dBA at full load), it operates from -20 °C to +70 °C (-25 °C cold start-up, linear derating to 60% load above +40 °C) up to 5,000 m altitude and carries IEC/EN/ES/CSA 60601-1, IEC/EN/UL/CSA 62368-1, EN 55011/55032 Class B, EN 60601-1-2:2015 (Ed. 4), CE, and UKCA approvals.

Key Features

  • Medical 2xMOPP Safety & Ultra-Low Leakage: Certified to IEC/EN/ES/CSA 60601-1 and 62368-1 with 4,000 VAC input-to-output isolation and under 250 µA leakage current at 265 VAC for Type B and BF medical equipment.
  • High-Density 801.6W Output & 94% Efficiency: Delivers 48 VDC (adjustable 46.8–51.8 V) at 16.7 A continuous using a 50–300 kHz LLC resonant topology achieving 92.2% efficiency at 115 VAC and 94.0% at 230 VAC.
  • Ultra-Low 1.67-Inch Profile & Quiet Variable Fan: Compact 171 × 85 × 42.5 mm (6.73 × 3.35 × 1.67 in) chassis features an intelligent variable-speed fan generating just 30 dBA at 20% load and 42 dBA at full load.
  • Comprehensive Auxiliary & Supervisory Controls: Includes an isolated 5 V / 2 A standby supply, 0.5 V total remote sense, opto-isolated remote on/off inhibit, and combined AC Fail / DC OK Power Good telemetry.
  • Class B EMI & 4th-Edition Medical EMC Immunity: Complies natively with EN 55011-B, EN 55032-B, and FCC Class B emissions plus EN 60601-1-2:2015 (Ed. 4), IEC 61000-4-2 through -11, and SEMI F47 line-dip standards.
  • High-Altitude & Wide-Temperature Reliability: Supports -25 °C cold start-up, full-load operation to +40 °C (linear derating to 60% at +70 °C), 5,000 m operating altitude, and a 452,227-hour Telcordia SR-332 MTBF backed by a 5-year warranty.

Technical Specifications

Manufacturer TDK-Lambda
Manufacturer Part Number (MPN) CUS800M-48
Alternate Part Number CUS800M48
Product Series CUS800M / CUS1000M / CUS1200M Series
Product Category Medical Enclosed PSUs
Component Type Enclosed Single-Output Medical & Industrial AC-DC Power Supply
Circuit Topology Active PFC + LLC Resonant Primary-to-Secondary Converter (50 kHz to 300 kHz)
Nominal Output Voltage 48 VDC
Output Voltage Adjustment Range 46.8 VDC to 51.8 VDC (via onboard potentiometer)
Rated Output Current 16.7 A
Output Current Range 0 A to 16.7 A (No minimum load required)
Rated Output Power 801.6 W
Standby Output Rating 5 VDC / 2 A (10 W auxiliary standby rail)
Output Ripple & Noise (Max) 480 mVp-p
Line Regulation (Max) 240 mV
Load Regulation (Max) 480 mV
Maximum External Load Capacitance 7,000 µF
Temperature Coefficient ±0.02 %/°C
Hold-Up Time (Typical at 115 VAC) 11 ms
Parallel Operation Capability Not supported (Parallel operation not possible)
Operating AC Input Voltage Range 85 VAC to 265 VAC (Safety certified for 90 VAC to 264 VAC)
Nominal AC Input Voltage Range 100 VAC to 240 VAC
Input Line Frequency 47 Hz to 63 Hz
Efficiency (at 115 VAC) 92.2 %
Efficiency (at 230 VAC) 94.0 %
AC Input Current (at 115 VAC) 8.0 A
AC Input Current (at 230 VAC) 4.0 A
Inrush Current (Cold Start at 115 VAC) 25.0 A
Inrush Current (Cold Start at 230 VAC) 50.0 A
Power Factor (PF at 115 VAC) 0.99
Power Factor (PF at 230 VAC) 0.95
No-Load Power Consumption (at 230 VAC) < 0.6 W typical (with Remote Off active and no load on 5 Vsb)
Overcurrent / Overload Protection (OCP/OLP) Threshold > 17.6 A
Overcurrent / Overload Protection (OCP/OLP) Mode Hiccup mode with automatic recovery (AC power reset may be required if output is dead-shorted)
Overvoltage Protection (OVP) Threshold 55.2 VDC to 64.8 VDC
Overvoltage Protection (OVP) Mode Output shutdown (latched; cycle AC input power to reset)
Overtemperature Protection (OTP) Mode Latching unit shutdown; cycle AC input power to reset
Withstand Isolation Voltage (Input to Output) 4,000 VAC for 1 minute (2 x MOPP medical isolation)
Withstand Isolation Voltage (Input to Ground / FG) 2,000 VAC for 1 minute (1 x MOPP medical isolation)
Withstand Isolation Voltage (Output to Ground / FG) 1,500 VAC for 1 minute (1 x MOPP medical isolation)
Insulation Protection Class Class I (Protective earth connection required)
Isolation Resistance (at 500 VDC) > 100 MΩ (at 25 °C, 70% RH, 500 VDC)
Earth Leakage Current < 250 µA (at 265 VAC, 60 Hz)
Operating Ambient Temperature Range -20 °C to +70 °C (-4 °F to +158 °F); capable of -25 °C (-13 °F) cold start-up
Full-Load Operating Temperature Limit +40 °C (+104 °F) at 100% rated load
High-Temperature Derating Rate Linear derating from 100% load at +40 °C (+104 °F) to 60% load at +70 °C (+158 °F) (1.33 %/°C)
Low-Line Input Derating Linear derating from 100% load at 90 VAC to 90% load at 85 VAC
Storage Temperature Range -40 °C to +75 °C (-40 °F to +167 °F)
Operating & Storage Relative Humidity 10% to 95% RH (non-condensing)
Operating & Storage Altitude Up to 5,000 m (16,404 ft)
Vibration Resistance (Non-Operating) 10 Hz to 55 Hz (1 min sweep), max. 19.6 m/s² (2 G), 1 hour per axis
Mechanical Shock Resistance < 196 m/s² (20 G)
MTBF (Telcordia SR-332) 452,227 hours (24V model baseline at full load, 115 VAC input, +40 °C ambient)
Manufacturer Warranty 5 Years
Medical Safety Certifications IEC 60601-1, EN 60601-1, ANSI/AAMI ES 60601-1, CSA C22.2 No. 60601-1 (2 x MOPP)
ITE / Industrial Safety Certifications IEC 62368-1, EN 62368-1, UL 62368-1, CSA C22.2 No. 62368-1
Regional Compliance Markings CE Mark, UKCA Mark
Conducted & Radiated EMI Emissions EN 55032 Class B, EN 55011 Class B, FCC Part 15 Class B
Harmonic Current Emissions (EN 61000-3-2) IEC / EN 61000-3-2 Class A
Medical EMC Immunity Standard EN 60601-1-2:2015 (4th Edition)
ESD Immunity (EN 61000-4-2) Levels 1, 2, 3, 4 — Performance Criteria A
Radiated RF Immunity (EN 61000-4-3) Level 3 — Performance Criteria A
EFT / Burst Immunity (EN 61000-4-4) Level 4 — Performance Criteria A
Surge Immunity (EN 61000-4-5) Levels 1, 2, 3, 4 (AC input power port) — Performance Criteria A
Conducted RF Immunity (EN 61000-4-6) Level 3 — Performance Criteria A
Power Frequency Magnetic Field Immunity (EN 61000-4-8) Level 4 — Performance Criteria A
Voltage Dips & Interruptions (EN 61000-4-11 & SEMI F47) EN 61000-4-11 Class 3 & EN 60601-1-2:2015 (Criteria A < 530 W for 0% 1/2 cycle; Criteria A < 330 W at 100 VAC / 200 VAC
Remote Sense Compensation Up to 0.5 V total load-line voltage drop compensation (+S / -S)
Remote On / Off Control Opto-isolated inhibit input (Logic Low = Output ON, Logic High = Output OFF)
Power Good / DC OK Telemetry Combined AC Fail and DC OK opto-isolated open-collector signal (PG)
Input Terminal Block & Pinout (TB1) 3-pin M3 screw terminal block (TIANLI TC25-01-03P-KC): Pin 1 = Live (L), Pin 2 = Neutral (N), Pin 3 = Protective Earth (FG); Torque: 0.5 to 0.8 N·m (5.1 to 8.2 kgf·cm)
Output Terminal Busbars & Torque (CN51A / CN51B) Dual M4 screw busbar terminals (CN51A = +Vout, CN51B = -Vout); Recommended torque: 1.18 N·m (12 kgf·cm)
Control Signal Connector & Pinout (CN51) JST B8B-PHDSS(LF)(SN) 8-pin header (Mating housing: PHDR-08VS, Contacts: SPHD-001T-P0.5) — Pin 1: +5V Standby, Pin 2: 5V Standby GND, Pin 3: -S (Remote Sense -), Pin 4: +S (Remote Sense +), Pin 5: SG (Signal Ground), Pin 6: PG (Power Good), Pin 7: R+ (Remote On/Off +), Pin 8: R- (Remote On/Off -)
Cooling Method & Acoustic Noise Internal variable-speed DC fan (airflow exits across output terminals); 30 dBA typical at 20% load, 42 dBA typical at 100% load
Chassis Mounting Provisions 12 x M3 tapped and standoff mounting holes (bottom & sides); max screw penetration 5 mm, recommended torque 0.49 N·m (5 kgf·cm)
Overall Length (L) 171.0 mm (6.73 in)
Overall Width (W) 85.0 mm (3.35 in)
Overall Height (H) 42.5 mm (1.67 in)
Overall Dimensions (L × W × H) 171.0 × 85.0 × 42.5 mm (6.73 × 3.35 × 1.67 in)
Net Weight 810 g (1.79 lbs / 28.57 oz)

Industrial Applications & Compatible Systems

TDK-Lambda CUS800M-48 is purpose-built for compact, acoustically sensitive clinical equipment and precision 48 VDC industrial electronics requiring dual medical/ITE safety approvals:

  • Diagnostic Imaging & Clinical Treatment Systems: Ultrasound scanners, patient monitors, surgical lasers, dental consoles, and automated laboratory analyzers requiring 2xMOPP isolation and under 250 µA earth leakage.
  • Test, Measurement & Semiconductor Fabrication Tools: High-density ATE racks, optical inspection stages, and SEMI F47-compliant wafer handling subsystems operating in 1U/low-profile enclosures.
  • Professional Broadcast & Audio/Video Infrastructure: Studio routers, LED wall processors, and acoustic-sensitive control room racks benefiting from 30 dBA idle fan noise and Class B EMI suppression.
  • Industrial Automation & 48V Motion Subsystems: Robotics controllers, PoE++ midspan injectors, and 48 VDC servo/actuator bus architectures in high-altitude facilities up to 5,000 m.

Installation, Thermal Management & MRO Guidelines

Installation & Engineering Guidelines: Mount the chassis using all 12 M3 tapped/standoff holes (0.49 N·m torque, strictly limiting screw penetration to 5 mm maximum) to ensure full EMI shielding and 2 G vibration compliance, while maintaining unobstructed airflow exiting across the output terminals. For ambient temperatures above +40 °C, apply linear load derating down to 60% (481 W) at +70 °C, and apply an additional linear derating to 90% load when operating between 90 VAC and 85 VAC. Because the CUS800M-48 does not support parallel current sharing, external ORing diodes are required for redundant N+1 architectures; size upstream branch protection for 8.0 A continuous input current and a 50.0 A cold-start inrush at 230 VAC.