CUS250M-12/U TDK-Lambda 250W 12VDC Medical Power Supply

$140.88

TDK-Lambda CUS250M-12/U (CUS250M12U) 250W 12VDC 20.83A U-channel medical AC-DC power supply, 80-264VAC input, featuring 2xMOPP isolation and Class B EMI.

SKU: CUS250M-12/U Category: Brand:

Description

TDK-Lambda CUS250M-12/U (also known as CUS250M12U) is a high-density 250W AC-DC power supply from the medical and industrial CUS250M series. Its part number nomenclature designates a 12VDC nominal output housed in a rugged U-channel aluminum chassis (/U suffix) with standard JST input connectors, dual AC line fuses, and an auxiliary 11.4VDC / 0.5A fan supply. Engineered with an active PFC stage and variable-frequency multi-stage conversion topology (25 to 300 kHz), it accepts a wide 80 to 264VAC input range and delivers up to 20.83A (adjustable from 12.0 to 13.2VDC) via conduction or forced-air cooling. The supply achieves up to 94% peak efficiency (>91% average efficiency), tight <0.5% line regulation, and <1.0% load regulation across 0 to 100% load.

Packaged in a low-profile U-channel footprint measuring 64.0 x 119.2 x 39.5 mm (2.52 x 4.69 x 1.56 in) and weighing 305 g (10.76 oz), the unit utilizes a JST B2P3-VH input header and heavy-duty M3 screw output terminals. Comprehensive onboard protection includes hiccup-mode overcurrent protection (110–170%) with automatic recovery alongside latching overvoltage (115–140%) and overtemperature shutdown circuits. Certified for both Class I and Class II (ungrounded) installations, it provides 4,000VAC (2xMOPP) input-to-output and 1,500VAC (1xMOPP) input/output-to-ground isolation for B and BF medical equipment. It operates from -20°C to +85°C (-40°C startup) and holds IEC/ES/EN 60601-1, IEC/UL/EN 62368-1, IEC/EN 61010-1, IEC/EN 60335-1, and EN 55011/55032 Class B approvals.

Key Features

  • Dual-Rated Cooling Performance: Delivers a full 250W (12VDC at 20.83A) utilizing either 1 m/s forced air or conduction cooling via its integral U-channel baseplate on a 300x300x1mm aluminum chassis.
  • Medical 2xMOPP & Class I/II Isolation: Features 4,000VAC (2xMOPP) input-to-output and 1,500VAC (1xMOPP) input/output-to-ground withstand isolation, <150µA leakage current, and <10µA Class I / <70µA Class II touch current.
  • High-Efficiency Active PFC Architecture: Achieves up to 94% peak conversion efficiency and >91% four-point average efficiency across 80–264VAC inputs, with >0.90 power factor at 115VAC and >14ms hold-up time.
  • Integrated Auxiliary Fan Output: Supplies a dedicated 11.4VDC at 0.5A auxiliary fan output via a 2-pin Molex 50-37-5023 (J500) header alongside user-adjustable 12.0 to 13.2VDC main output regulation.
  • Comprehensive Fault Protection Suite: Incorporates automatic-recovery hiccup overcurrent protection (110–170%), latching overvoltage protection (115–140%), latching overtemperature protection, and dual internal 5A fast-acting AC input fuses.
  • Wide Thermal & Altitude Envelope: Operates across -20°C to +85°C ambient temperatures (with -40°C cold start-up capability) and up to 5,000m (16,404 ft) elevation for demanding medical and industrial environments.
  • Multi-Sector Global Safety & Class B EMI: Certified to IEC/ES/EN 60601-1, IEC/UL/EN 62368-1, IEC/EN 61010-1, and IEC/EN 60335-1 standards with EN 55011 / EN 55032 Class B conducted and radiated EMI compliance.

Technical Specifications

Manufacturer TDK-Lambda
Manufacturer Part Number (MPN) CUS250M-12/U
Alternate Part Number CUS250M12U
Product Series CUS250M Series
Product Category Medical Enclosed PSUs
Component Type U-Channel Conduction & Forced-Air Cooled Medical & ITE AC-DC Power Supply
Circuit Topology Variable-Frequency Active PFC + Resonant DC-DC & Flyback Conversion
Nominal Output Voltage 12 VDC
Output Voltage Adjustment Range 12.0 to 13.2 VDC
Rated Output Current 20.83 A (with forced air or conduction cooling)
Rated Output Power 250 W (with forced air or conduction cooling on metal baseplate)
Peak Output Power 250 W peak for 1 to 10 min pulse time (10% duty cycle in convection cooling; see derating curves)
Number of Outputs 1 Main Output + 1 Standard Auxiliary Fan Supply
Ch1 Output Voltage & Current 12 VDC at 20.83 A (Main Output)
Ch2 Output Voltage & Current 11.4 VDC at 0.5 A (Standard Auxiliary Fan Supply)
Minimum Load Current 0 A (No minimum load required)
Output Ripple & Noise (Max.) <1% (<120 mVp-p above 0 °C); <2% (<240 mVp-p at -20 °C or <10% load burst mode)
Line Regulation <0.5% (across 85 to 264 VAC)
Load Regulation <1.0% (across 0 to 100% load)
Hold-Up Time >14 ms (at 115 / 230 VAC)
Max. Capacitive Load 20,830 µF
AC Input Voltage Range 80 to 264 VAC (Operating)
Nominal Input Voltage 100 to 240 VAC (Safety certified for 80 to 264 VAC; Class I or Class II installation)
Input Frequency Range 47 to 63 Hz
Efficiency Up to 94% peak; >91% average efficiency (measured at 25%, 50%, 75%, and 100% load)
AC Input Current (at 115 VAC) 3.1 A max. (measured at 100 VAC)
Inrush Current (Cold Start at 230 VAC) <75 A (Inrush I²t is significantly below internal 5 A fast-acting dual fuses)
Power Factor (PF at 115 VAC) >0.90 (at >20% load)
Power Factor (PF at 230 VAC) >0.70 (at >20% load)
Power Factor Correction (PFC) Active PFC (Meets IEC/EN 61000-3-2 Class A)
No-Load Power Consumption <0.5 W (at 230 VAC when output is inhibited)
Switching Frequency 25 to 300 kHz variable (for PFC, DC-DC, and flyback stages, excluding burst mode)
Mounting Style Chassis Mount (U-Channel with 4 x M3 underside fixing holes)
Cooling Method Conduction, Natural Convection, or Forced Air (1 m/s)
Package / Footprint Standard U-Channel Chassis (2.52 x 4.69 in footprint; based on 2 x 4 in internal core)
Overall Width (W) 64.0 mm (2.52 in)
Overall Height (H) 39.5 mm (1.56 in)
Overall Depth (D) 119.2 mm (4.69 in)
Dimensional Tolerance ±0.4 mm (±0.016 in) unstated; ±0.2 mm (±0.008 in) mounting pitch
Mounting Hole Pitch 95.24 x 44.45 mm (3.75 x 1.75 in) or 112.0 x 35.0 mm (4.41 x 1.38 in)
Net Weight 305 g (0.67 lbs / 10.76 oz)
Termination / Connector Style AC Input: JST 3-Pin Header; DC Output: M3 Screw Terminals; Fan: Molex 2-Pin Header
Input Terminal Pinout J1 (JST B2P3-VH, mates with Housing VAR-2 / Crimp SVA-41T-P1.1): Pin 1 = Neutral, Pin 2 = Not Connected, Pin 3 = Live; Dual internal 5 A fast-acting fuses
Output & Signal Terminal Pinout J100 = +V CH1 (M3 Screw Tag); J101 = 0V CH1 (M3 Screw Tag); J500 Fan Header (Molex 22-05-7025, mates with 50-37-5023 / 08-70-1039): Pin 1 = +V Fan (11.4 V), Pin 2 = 0V Fan
Terminal Tightening Torque 0.5 to 0.6 Nm (4.4 to 5.3 lb-in) for M3 fixing screws (max penetration 3.75 mm)
Enclosure / Potting Material Aluminum U-Channel Chassis (Pollution Degree 2, Material Group IIIb)
Overload Protection (OLP) Threshold 110% to 170% of rated output current
Overload Protection (OLP) Mode Hiccup mode with automatic recovery
Overvoltage Protection (OVP) Threshold 115% to 140% of nominal output voltage (13.8 to 16.8 VDC)
Overvoltage Protection (OVP) Mode Latching shutdown (cycle AC input power to reset)
Overtemperature Protection (OTP) Mode Latching shutdown (cycle AC input power to reset)
Short Circuit Protection (SCP) Continuous hiccup mode with automatic recovery
Withstand Isolation Voltage (I/P to O/P) 4,000 VAC for 1 minute (2xMOPP)
Withstand Isolation Voltage (I/P to FG / Case) 1,500 VAC for 1 minute (1xMOPP)
Withstand Isolation Voltage (O/P to FG / Case) 1,500 VAC for 1 minute (1xMOPP)
Isolation Resistance (at 500 VDC) >100 MΩ (at 25 °C, 70% RH)
Earth Leakage Current <150 µA (at 264 VAC, 63 Hz); Touch Current: <10 µA (Class I), <70 µA (Class II)
Medical Isolation Rating 2xMOPP (Input to Output), 1xMOPP (Input/Output to Ground); Suitable for Type B and BF Medical Equipment
Operating Ambient Temperature Range -20 °C to +85 °C (-4 °F to +185 °F); -40 °C (-40 °F) cold start-up supported
Full-Load Operating Temperature Limit +45 °C (+113 °F) at 250 W with 1 m/s forced air or conduction cooling on 300x300x1 mm aluminum plate
High-Temperature Derating Rate Derates linearly above +45 °C up to +85 °C (see orientation and cooling method derating curves)
Low-Line Input Derating Derate 1%/VAC from 100 to 90 VAC (225 W at 90 VAC); 2%/VAC from 90 to 80 VAC (180 W at 80 VAC)
Altitude Derating Up to 5,000 m (16,404 ft) max. operating altitude
Storage Temperature Range -40 °C to +85 °C (-40 °F to +185 °F)
Operating Relative Humidity 5% to 95% RH (non-condensing)
Temperature Coefficient ±0.02%/°C
Vibration Resistance 2G, 10 to 500 Hz for 1 hour (non-operating); Shock: 30G, 11 ms half-sine (non-operating)
Safety Approvals (UL / North America) ES 60601-1, UL 62368-1
Safety Approvals (IEC / EN / TÜV) IEC/EN 60601-1, IEC/EN 62368-1, IEC/EN 61010-1, IEC/EN 60335-1
Safety Approvals (Regional / Global) CE Mark, UKCA Mark (Manufacturer Warranty: 5 Years)
EMC Conducted & Radiated Emissions EN 55032 Class B, EN 55011 Class B
EMC Harmonic & Flicker Compliance IEC/EN 61000-3-2 Class A, EN 61000-4-14 Class 3 Criteria A
EMC Immunity Standards IEC 60601-1-2 Ed.4.1, CISPR 35, IEC 61204-3, IEC 61000-6-2, SEMI F47, IEC 61000-4-2 (ESD Level 4 Enclosure/AC, Level 3 O/P), IEC 61000-4-3 (10 V/m 80MHz–6GHz), IEC 61000-4-4 (EFT 4 kV AC / 2 kV DC), IEC 61000-4-5 (Surge 2 kV CM / 1 kV NM), IEC 61000-4-6 (10 V), IEC 61000-4-8 (Level 4), IEC 61000-4-11 (Dips), IEC 61000-4-12 (Ringwave Level 3)
Environmental Compliance RoHS, REACH, CE, UKCA

Industrial Applications & Compatible Systems

The TDK-Lambda CUS250M-12/U is purpose-built for space-constrained medical, laboratory, and industrial subsystems requiring silent conduction cooling or high-power forced-air operation:

  • Type B & BF Medical Diagnostic Equipment: Powers patient monitors, ultrasound consoles, clinical blood analyzers, and medical lasers requiring 2xMOPP isolation, <10µA Class I touch current, and IEC 60601-1-2 Ed.4.1 immunity.
  • Laboratory & Analytical Instrumentation: Delivers low-ripple 12VDC power for mass spectrometers, chromatography systems, and automated liquid handlers certified under IEC/EN 61010-1.
  • Silent Broadcast & Pro-AV Systems: Eliminates acoustic fan noise in studio consoles and enclosed industrial PCs by transferring heat directly through the U-channel baseplate into the equipment enclosure.
  • Semiconductor & Commercial Appliances: Supports wafer handling controls and commercial equipment requiring SEMI F47 voltage-sag immunity and IEC/EN 60335-1 compliance.

Installation, Thermal Management & MRO Guidelines

Follow these critical thermal, mechanical, and electrical practices when integrating the CUS250M-12/U into host equipment:

  • Conduction vs. Forced-Air Thermal Mounting: To achieve the full 250W continuous rating without fans up to +45°C, bolt the U-channel flat onto a 300 x 300 x 1 mm aluminum chassis plate or provide 1 m/s forced airflow with at least 5 mm side clearance.
  • M3 Chassis Fastener Penetration & Torque: Secure the four underside “B” M3 mounting inserts using a tightening torque of 0.5 to 0.6 Nm, ensuring screw penetration into the U-channel baseplate never exceeds 3.75 mm.
  • Low-Line Derating & Latching Fault Reset: Account for linear power derating below 100VAC (225W at 90VAC; 180W at 80VAC), and remember that OVP (115–140%) and OTP triggers latch the unit off, requiring an AC input power cycle to reset.