EMH350PS48-01 XP Power 350W 48VDC Medical Power Supply

$347.76

XP Power EMH350PS48-01 (EMH350PS4801) 354W 48VDC 7.3A medical and ITE chassis power supply, 80-275VAC input, featuring 2xMOPP isolation, 5V standby.

SKU: EMH350PS48-01 Category: Brand:

Description

XP Power EMH350PS48-01 (also known as EMH350PS4801) is a high-density 350W single-output chassis-mount AC-DC power supply from the EMH350 series engineered for BF-rated medical and industrial ITE equipment. Within the series nomenclature, PS48 denotes a regulated 48 VDC main output at 7.3 A alongside dedicated 5.0 VDC / 2.0 A standby and 12 VDC / 0.6 A fan auxiliary rails for 354 W total output power, while the -01 suffix equips the unit with an expanded 12-pin analog control and monitoring interface. Operating across an ultra-wide 80 to 275 VAC input range with active power factor correction (>0.9) and 87% typical efficiency, it delivers ±10% output voltage trim, ±0.5% line regulation, ±1% load regulation, and 1% pk-pk maximum ripple and noise.

Packaged in a compact 76.2 × 127.0 × 36.3 mm (3.00 × 5.00 × 1.43 in) open-chassis footprint weighing 410 g (0.90 lb), the unit utilizes a 3-pin Tyco SL-156 AC input header, dual M4 screw output terminals, and a 12-pin JST S12B-PHDSS signal connector supporting remote sense, remote inhibit, AC OK/Power Fail warning, and single-wire active current sharing for up to three parallel modules. Comprehensive protection circuits include latching overvoltage protection (115–140%), power-limiting overload protection (110–150%), auto-recovery overtemperature shutdown, and continuous constant-current short-circuit protection. Certified to IEC/EN/ANSI/AAMI ES60601-1 (2×MOPP input-to-output, 4000 VAC isolation, ≤100 µA patient leakage) and IEC/UL/EN 62368-1, it operates reliably from 0 °C to +70 °C (32 °F to 158 °F) with 12 CFM forced-air cooling.

Key Features

  • Medical 2×MOPP & ITE Safety Approvals: Certified to IEC/EN/ES60601-1 for BF-rated patient-connected equipment and IEC/UL/EN62368-1, providing 4000 VAC input-to-output isolation and 100 µA maximum patient leakage current.
  • High-Density 48 VDC Triple-Rail Architecture: Delivers 48 VDC at 7.3 A (adjustable ±10%) plus always-on 5.0 VDC / 2.0 A standby and 12 VDC / 0.6 A fan outputs totaling 354 W at 16.3 W/in³.
  • Expanded -01 Analog Control & Parallel Sharing: Features a 12-pin JST S12B-PHDSS interface supporting 0.5 V remote sense compensation, open-collector Power Fail warning (≥3 ms), remote inhibit, and 10% active current sharing across up to 3 units.
  • Ultra-Wide 80–275 VAC Input with Active PFC: Operates globally from 80 to 275 VAC (47–63 Hz) with >0.9 power factor, 87% typical efficiency, dual line/neutral internal fuses, and linear low-line derating below 90 VAC.
  • Comprehensive Fault Protection & Fast Transient Recovery: Incorporates latching OVP (115–140%), 110–150% overload power limiting, auto-reset OTP, continuous short-circuit protection, and 500 µs recovery within 1% for 50–75–50% load steps.
  • Rugged Compact Chassis & EMI Class B Compliance: Measures 76.2 × 127.0 × 36.3 mm (3.00 × 5.00 × 1.43 in) with M4 output terminals, 400,000-hour MIL-HDBK-217F MTBF, 30g shock resistance, and EN55011/32 Class B conducted emissions.

Technical Specifications

Manufacturer XP Power
Manufacturer Part Number (MPN) EMH350PS48-01
Alternate Part Number EMH350PS4801
Product Series EMH350 Series
Product Category Medical Enclosed PSUs
Component Type Medical & Industrial Chassis-Mount AC-DC Power Supply (with -01 Analog Signals)
Circuit Topology Active PFC + Main Forward/Resonant Converter + Isolated Standby Converter (62–560 kHz PFC / 71 kHz Main / 50–130 kHz Standby)
Nominal Output Voltage 48 VDC (Main V1); 5.0 VDC (Always-On Standby); 12 VDC (Fan Supply V3)
Output Voltage Adjustment Range 43.2 to 52.8 VDC (±10% via onboard potentiometer)
Rated Output Current 7.3 A (Main V1); 2.0 A (5V Standby); 0.6 A (12V Fan Supply V3)
Output Current Range 0.1 to 7.3 A (100 mA minimum load required on V1)
Rated Output Power 354 W (Total combined output power; 16.3 W/in³ power density)
Number of Outputs 3 (Main 48 VDC + 5.0 VDC Standby + 12 VDC Fan Supply)
Output Ripple & Noise (Max.) 480 mVp-p (1% pk-pk max., measured at 20 MHz bandwidth)
Output Voltage Tolerance ±1% initial set accuracy (V1)
Line Regulation ±0.5% max.
Load Regulation ±1% (Main V1); ±5% (5V Standby)
Transient Recovery Time & Deviation 500 µs recovery to within 1% (4% max. deviation for 50%–75%–50% step load change)
Setup Time 1000 ms (1 s typical start-up delay)
Rise Time 50 ms (start-up rise time)
Hold-Up Time 16 ms typical (at 115 / 230 VAC)
AC Input Voltage Range 80 to 275 VAC (single-phase; linear derating required below 90 VAC)
Input Frequency Range 47 to 63 Hz
Efficiency 87% typical
AC Input Current (at 115 VAC) 3.6 A typical (at 350 W full load)
AC Input Current (at 230 VAC) 1.8 A typical (at 350 W full load)
Inrush Current (Cold Start at 230 VAC) 40 A max. (cold start at 25 °C)
Power Factor Correction (PFC) Active PFC (>0.9 power factor)
Input Protection Dual internal 10 AF / 250 V fuses (Line and Neutral)
Mounting Style Chassis Mount (Open-frame U-channel with 4× M3 bottom mounting holes; compatible with -TF top fan cover)
Cooling Method Forced Air Cooling (12 CFM required across chassis; 12 VDC / 0.6 A fan header provided)
Overall Width (W) 76.2 mm (3.00 in)
Overall Height (H) 36.3 mm (1.43 in)
Overall Depth (D) 127.0 mm (5.00 in)
Dimensional Tolerance ±0.5 mm (±0.02 in)
Mounting Hole Pitch 115.6 × 64.8 mm (4.55 × 2.55 in) via 4× M3 threaded holes
Net Weight 410 g (0.90 lb)
Termination / Connector Style AC Input: 5-pin (3 loaded) Tyco SL-156 (P/N 640445-5); DC Output: Dual M4 Pan-Head Screws; Signals: 12-pin 2.0 mm JST S12B-PHDSS
Input Terminal Pinout CON1 (Tyco 640445-5, pins 2 & 4 removed): Pin 1 = Line (L), Pin 3 = Neutral (N), Pin 5 = Safety Earth (E)
Output & Signal Terminal Pinout Main Output: CON2 = +V1 (48 VDC), CON3 = -V1 (M4 screws); 12-Pin Signals Connector (JST S12B-PHDSS): Pins 1–2 = -VE Standby, Pins 3–4 = +VE Standby (5V/2A), Pin 5 = +VE Fan (12V/0.6A), Pin 6 = -VE Fan, Pin 7 = +VE Sense, Pin 8 = -VE Sense, Pin 9 = Current Share, Pin 10 = Inhibit, Pin 11 = Power Fail (AC OK), Pin 12 = N/C
Control & Monitoring Signals Remote Sense (0.5 V drop compensation), Remote Inhibit (pull <0.4 V to disable V1 & V3), Open-Collector AC OK / Power Fail (≥3 ms warning), Single-Wire Current Share (up to 3 parallel units within 10%)
Terminal Wire Gauge Range 26 to 22 AWG (for mating JST PHDR-12VS signal contacts JST SPHD-001T-P0.5)
Terminal Tightening Torque 0.90 N·m (8 lb-in / 90 cN·m) max. on M4 output terminal screws
Overload Protection (OLP) Threshold 110% to 150% of rated power
Overload Protection (OLP) Mode Power limit mode
Overvoltage Protection (OVP) Threshold 115% to 140% of nominal V1 (55.2 to 67.2 VDC)
Overvoltage Protection (OVP) Mode Latching shutdown (recycle AC input to reset)
Overtemperature Protection (OTP) Mode Thermal shutdown with automatic recovery
Short Circuit Protection (SCP) Continuous protection, approximately constant current mode
Withstand Isolation Voltage (I/P to O/P) 4000 VAC (2×MOPP reinforced medical isolation)
Withstand Isolation Voltage (I/P to FG / Case) 1500 VAC (1×MOPP)
Withstand Isolation Voltage (O/P to FG / Case) 1500 VAC (1×MOPP)
Earth Leakage Current 250 µA max. (at 264 VAC, 60 Hz); Patient Leakage Current: 100 µA max.
Medical Isolation Rating 2×MOPP (Input-to-Output) / 1×MOPP (Input/Output-to-Ground), BF Rated
Operating Ambient Temperature Range 0 °C to +70 °C (32 °F to 158 °F)
Full-Load Operating Temperature Limit +50 °C (122 °F) at 100% load with 12 CFM forced airflow
High-Temperature Derating Rate Linear derating at 2.5% / °C from +50 °C to +70 °C (50% load capacity at +70 °C / 158 °F)
Low-Line Input Derating Linear derating from 100% load at 90 VAC down to 80% load at 80 VAC
Altitude Derating Rated for operation up to 3048 m (10,000 ft) altitude
Storage Temperature Range -40 °C to +85 °C (-40 °F to +185 °F)
Operating Relative Humidity Up to 95% RH, non-condensing
Temperature Coefficient ±0.05% / °C
Vibration Resistance 2g sweep (10 to 500 Hz) and resonance endurance across all 3 axes (conforms to EN 60068-2-6); Shock: ±3× 30g half-sine 11 ms pulses per axis (18 total, EN 60068-2-27)
Safety Approvals (UL / North America) ANSI/AAMI ES60601-1, CSA C22.2 No. 60601-1 (cUL BF Rated), UL 62368-1:2014, CSA C22.2 No. 62368-1-14, UL 60950-1
Safety Approvals (IEC / EN / TÜV) IEC/EN 60601-1 (including Risk Management), IEC/EN 62368-1:2014+A11:2017, EN 60950-1
Safety Approvals (Regional / Global) CE, UKCA
EMC Conducted & Radiated Emissions EN 55011 / EN 55032 Class B (Conducted), EN 55011 / EN 55032 Class A (Radiated)
EMC Harmonic & Flicker Compliance EN 61000-3-2 Class A (Harmonic Currents), EN 61000-3-3 (Voltage Flicker)
EMC Immunity Standards EN 61000-4-3 Level 3 Criteria A (Radiated RF), EN 61000-4-4 Level 3 Criteria A (EFT/Burst), EN 61000-4-5 Installation Class 3 Criteria A (Surge), EN 61000-4-6 Level 3 Criteria A (Conducted RF), EN 61000-4-11 & EN 60601-1-2 Criteria A/B (Voltage Dips & Interruptions)
MTBF (MIL-HDBK-217F) 400,000 hours (at 25 °C ambient, Ground Benign)
Environmental Compliance RoHS & REACH Compliant

Industrial Applications & Compatible Systems

XP Power EMH350PS48-01 is engineered for mission-critical medical devices, laboratory diagnostic analyzers, and high-reliability industrial 48 VDC control systems requiring BF-rated isolation and active signal telemetry:

  • Diagnostic Imaging & Patient-Connected Medical Equipment: Supplies clean 48 VDC bus power with 2×MOPP (4000 VAC) isolation and ≤100 µA patient leakage for ultrasound consoles, surgical lighting, endoscopy towers, and patient monitoring carts.
  • Automated Clinical Chemistry & Laboratory Analyzers: Provides 354 W continuous power with an always-on 5.0 VDC / 2.0 A logic standby rail and Power Fail early-warning signal for graceful sample-handling shutdown.
  • Parallel Redundant & Scalable 48 VDC Power Architectures: Utilizes pin-9 single-wire current sharing to parallel up to three EMH350PS48-01 units within 10% load balance for N+1 redundant medical or industrial racks.
  • Industrial Robotics, Telecom & Precision Instrumentation: Powers 48 VDC servo actuators, optical inspection rigs, and test-and-measurement chassis requiring remote inhibit control and Class B conducted EMI compliance.

Installation, Thermal Management & MRO Guidelines

Follow these OEM mechanical, thermal, and electrical integration guidelines to ensure full datasheet performance and regulatory compliance:

  • Forced-Air Cooling & Temperature Derating: Maintain at least 12 CFM longitudinal airflow across the chassis (or pair with an XP Power top-fan cover) to sustain 354 W full load up to +50 °C; above +50 °C, derate linearly at 2.5% / °C down to 50% load at +70 °C.
  • Safety Earth Bonding & EMC Grounding: Connect all four M3 chassis mounting holes (marked with earth symbols on the mechanical drawing) directly to protective safety earth to achieve EN 55011/32 Class B conducted and Class A radiated EMI performance.
  • Minimum Load, Remote Sense & Mating Connectors: Maintain at least 100 mA minimum load on the 48 VDC output for tight regulation, route twisted-pair remote sense leads (pins 7 & 8 of the JST S12B-PHDSS header) to compensate up to 0.5 V line drop, and torque M4 output screws to ≤0.90 N·m (8 lb-in).