Description
XP Power EMH350PS48-01 (also known as EMH350PS4801) is a high-density 350W single-output chassis-mount AC-DC power supply from the EMH350 series engineered for BF-rated medical and industrial ITE equipment. Within the series nomenclature, PS48 denotes a regulated 48 VDC main output at 7.3 A alongside dedicated 5.0 VDC / 2.0 A standby and 12 VDC / 0.6 A fan auxiliary rails for 354 W total output power, while the -01 suffix equips the unit with an expanded 12-pin analog control and monitoring interface. Operating across an ultra-wide 80 to 275 VAC input range with active power factor correction (>0.9) and 87% typical efficiency, it delivers ±10% output voltage trim, ±0.5% line regulation, ±1% load regulation, and 1% pk-pk maximum ripple and noise.
Packaged in a compact 76.2 × 127.0 × 36.3 mm (3.00 × 5.00 × 1.43 in) open-chassis footprint weighing 410 g (0.90 lb), the unit utilizes a 3-pin Tyco SL-156 AC input header, dual M4 screw output terminals, and a 12-pin JST S12B-PHDSS signal connector supporting remote sense, remote inhibit, AC OK/Power Fail warning, and single-wire active current sharing for up to three parallel modules. Comprehensive protection circuits include latching overvoltage protection (115–140%), power-limiting overload protection (110–150%), auto-recovery overtemperature shutdown, and continuous constant-current short-circuit protection. Certified to IEC/EN/ANSI/AAMI ES60601-1 (2×MOPP input-to-output, 4000 VAC isolation, ≤100 µA patient leakage) and IEC/UL/EN 62368-1, it operates reliably from 0 °C to +70 °C (32 °F to 158 °F) with 12 CFM forced-air cooling.
Key Features
- Medical 2×MOPP & ITE Safety Approvals: Certified to IEC/EN/ES60601-1 for BF-rated patient-connected equipment and IEC/UL/EN62368-1, providing 4000 VAC input-to-output isolation and 100 µA maximum patient leakage current.
- High-Density 48 VDC Triple-Rail Architecture: Delivers 48 VDC at 7.3 A (adjustable ±10%) plus always-on 5.0 VDC / 2.0 A standby and 12 VDC / 0.6 A fan outputs totaling 354 W at 16.3 W/in³.
- Expanded -01 Analog Control & Parallel Sharing: Features a 12-pin JST S12B-PHDSS interface supporting 0.5 V remote sense compensation, open-collector Power Fail warning (≥3 ms), remote inhibit, and 10% active current sharing across up to 3 units.
- Ultra-Wide 80–275 VAC Input with Active PFC: Operates globally from 80 to 275 VAC (47–63 Hz) with >0.9 power factor, 87% typical efficiency, dual line/neutral internal fuses, and linear low-line derating below 90 VAC.
- Comprehensive Fault Protection & Fast Transient Recovery: Incorporates latching OVP (115–140%), 110–150% overload power limiting, auto-reset OTP, continuous short-circuit protection, and 500 µs recovery within 1% for 50–75–50% load steps.
- Rugged Compact Chassis & EMI Class B Compliance: Measures 76.2 × 127.0 × 36.3 mm (3.00 × 5.00 × 1.43 in) with M4 output terminals, 400,000-hour MIL-HDBK-217F MTBF, 30g shock resistance, and EN55011/32 Class B conducted emissions.
Technical Specifications
| Manufacturer | XP Power |
|---|---|
| Manufacturer Part Number (MPN) | EMH350PS48-01 |
| Alternate Part Number | EMH350PS4801 |
| Product Series | EMH350 Series |
| Product Category | Medical Enclosed PSUs |
| Component Type | Medical & Industrial Chassis-Mount AC-DC Power Supply (with -01 Analog Signals) |
| Circuit Topology | Active PFC + Main Forward/Resonant Converter + Isolated Standby Converter (62–560 kHz PFC / 71 kHz Main / 50–130 kHz Standby) |
| Nominal Output Voltage | 48 VDC (Main V1); 5.0 VDC (Always-On Standby); 12 VDC (Fan Supply V3) |
| Output Voltage Adjustment Range | 43.2 to 52.8 VDC (±10% via onboard potentiometer) |
| Rated Output Current | 7.3 A (Main V1); 2.0 A (5V Standby); 0.6 A (12V Fan Supply V3) |
| Output Current Range | 0.1 to 7.3 A (100 mA minimum load required on V1) |
| Rated Output Power | 354 W (Total combined output power; 16.3 W/in³ power density) |
| Number of Outputs | 3 (Main 48 VDC + 5.0 VDC Standby + 12 VDC Fan Supply) |
| Output Ripple & Noise (Max.) | 480 mVp-p (1% pk-pk max., measured at 20 MHz bandwidth) |
| Output Voltage Tolerance | ±1% initial set accuracy (V1) |
| Line Regulation | ±0.5% max. |
| Load Regulation | ±1% (Main V1); ±5% (5V Standby) |
| Transient Recovery Time & Deviation | 500 µs recovery to within 1% (4% max. deviation for 50%–75%–50% step load change) |
| Setup Time | 1000 ms (1 s typical start-up delay) |
| Rise Time | 50 ms (start-up rise time) |
| Hold-Up Time | 16 ms typical (at 115 / 230 VAC) |
| AC Input Voltage Range | 80 to 275 VAC (single-phase; linear derating required below 90 VAC) |
| Input Frequency Range | 47 to 63 Hz |
| Efficiency | 87% typical |
| AC Input Current (at 115 VAC) | 3.6 A typical (at 350 W full load) |
| AC Input Current (at 230 VAC) | 1.8 A typical (at 350 W full load) |
| Inrush Current (Cold Start at 230 VAC) | 40 A max. (cold start at 25 °C) |
| Power Factor Correction (PFC) | Active PFC (>0.9 power factor) |
| Input Protection | Dual internal 10 AF / 250 V fuses (Line and Neutral) |
| Mounting Style | Chassis Mount (Open-frame U-channel with 4× M3 bottom mounting holes; compatible with -TF top fan cover) |
| Cooling Method | Forced Air Cooling (12 CFM required across chassis; 12 VDC / 0.6 A fan header provided) |
| Overall Width (W) | 76.2 mm (3.00 in) |
| Overall Height (H) | 36.3 mm (1.43 in) |
| Overall Depth (D) | 127.0 mm (5.00 in) |
| Dimensional Tolerance | ±0.5 mm (±0.02 in) |
| Mounting Hole Pitch | 115.6 × 64.8 mm (4.55 × 2.55 in) via 4× M3 threaded holes |
| Net Weight | 410 g (0.90 lb) |
| Termination / Connector Style | AC Input: 5-pin (3 loaded) Tyco SL-156 (P/N 640445-5); DC Output: Dual M4 Pan-Head Screws; Signals: 12-pin 2.0 mm JST S12B-PHDSS |
| Input Terminal Pinout | CON1 (Tyco 640445-5, pins 2 & 4 removed): Pin 1 = Line (L), Pin 3 = Neutral (N), Pin 5 = Safety Earth (E) |
| Output & Signal Terminal Pinout | Main Output: CON2 = +V1 (48 VDC), CON3 = -V1 (M4 screws); 12-Pin Signals Connector (JST S12B-PHDSS): Pins 1–2 = -VE Standby, Pins 3–4 = +VE Standby (5V/2A), Pin 5 = +VE Fan (12V/0.6A), Pin 6 = -VE Fan, Pin 7 = +VE Sense, Pin 8 = -VE Sense, Pin 9 = Current Share, Pin 10 = Inhibit, Pin 11 = Power Fail (AC OK), Pin 12 = N/C |
| Control & Monitoring Signals | Remote Sense (0.5 V drop compensation), Remote Inhibit (pull <0.4 V to disable V1 & V3), Open-Collector AC OK / Power Fail (≥3 ms warning), Single-Wire Current Share (up to 3 parallel units within 10%) |
| Terminal Wire Gauge Range | 26 to 22 AWG (for mating JST PHDR-12VS signal contacts JST SPHD-001T-P0.5) |
| Terminal Tightening Torque | 0.90 N·m (8 lb-in / 90 cN·m) max. on M4 output terminal screws |
| Overload Protection (OLP) Threshold | 110% to 150% of rated power |
| Overload Protection (OLP) Mode | Power limit mode |
| Overvoltage Protection (OVP) Threshold | 115% to 140% of nominal V1 (55.2 to 67.2 VDC) |
| Overvoltage Protection (OVP) Mode | Latching shutdown (recycle AC input to reset) |
| Overtemperature Protection (OTP) Mode | Thermal shutdown with automatic recovery |
| Short Circuit Protection (SCP) | Continuous protection, approximately constant current mode |
| Withstand Isolation Voltage (I/P to O/P) | 4000 VAC (2×MOPP reinforced medical isolation) |
| Withstand Isolation Voltage (I/P to FG / Case) | 1500 VAC (1×MOPP) |
| Withstand Isolation Voltage (O/P to FG / Case) | 1500 VAC (1×MOPP) |
| Earth Leakage Current | 250 µA max. (at 264 VAC, 60 Hz); Patient Leakage Current: 100 µA max. |
| Medical Isolation Rating | 2×MOPP (Input-to-Output) / 1×MOPP (Input/Output-to-Ground), BF Rated |
| Operating Ambient Temperature Range | 0 °C to +70 °C (32 °F to 158 °F) |
| Full-Load Operating Temperature Limit | +50 °C (122 °F) at 100% load with 12 CFM forced airflow |
| High-Temperature Derating Rate | Linear derating at 2.5% / °C from +50 °C to +70 °C (50% load capacity at +70 °C / 158 °F) |
| Low-Line Input Derating | Linear derating from 100% load at 90 VAC down to 80% load at 80 VAC |
| Altitude Derating | Rated for operation up to 3048 m (10,000 ft) altitude |
| Storage Temperature Range | -40 °C to +85 °C (-40 °F to +185 °F) |
| Operating Relative Humidity | Up to 95% RH, non-condensing |
| Temperature Coefficient | ±0.05% / °C |
| Vibration Resistance | 2g sweep (10 to 500 Hz) and resonance endurance across all 3 axes (conforms to EN 60068-2-6); Shock: ±3× 30g half-sine 11 ms pulses per axis (18 total, EN 60068-2-27) |
| Safety Approvals (UL / North America) | ANSI/AAMI ES60601-1, CSA C22.2 No. 60601-1 (cUL BF Rated), UL 62368-1:2014, CSA C22.2 No. 62368-1-14, UL 60950-1 |
| Safety Approvals (IEC / EN / TÜV) | IEC/EN 60601-1 (including Risk Management), IEC/EN 62368-1:2014+A11:2017, EN 60950-1 |
| Safety Approvals (Regional / Global) | CE, UKCA |
| EMC Conducted & Radiated Emissions | EN 55011 / EN 55032 Class B (Conducted), EN 55011 / EN 55032 Class A (Radiated) |
| EMC Harmonic & Flicker Compliance | EN 61000-3-2 Class A (Harmonic Currents), EN 61000-3-3 (Voltage Flicker) |
| EMC Immunity Standards | EN 61000-4-3 Level 3 Criteria A (Radiated RF), EN 61000-4-4 Level 3 Criteria A (EFT/Burst), EN 61000-4-5 Installation Class 3 Criteria A (Surge), EN 61000-4-6 Level 3 Criteria A (Conducted RF), EN 61000-4-11 & EN 60601-1-2 Criteria A/B (Voltage Dips & Interruptions) |
| MTBF (MIL-HDBK-217F) | 400,000 hours (at 25 °C ambient, Ground Benign) |
| Environmental Compliance | RoHS & REACH Compliant |
Industrial Applications & Compatible Systems
XP Power EMH350PS48-01 is engineered for mission-critical medical devices, laboratory diagnostic analyzers, and high-reliability industrial 48 VDC control systems requiring BF-rated isolation and active signal telemetry:
- Diagnostic Imaging & Patient-Connected Medical Equipment: Supplies clean 48 VDC bus power with 2×MOPP (4000 VAC) isolation and ≤100 µA patient leakage for ultrasound consoles, surgical lighting, endoscopy towers, and patient monitoring carts.
- Automated Clinical Chemistry & Laboratory Analyzers: Provides 354 W continuous power with an always-on 5.0 VDC / 2.0 A logic standby rail and Power Fail early-warning signal for graceful sample-handling shutdown.
- Parallel Redundant & Scalable 48 VDC Power Architectures: Utilizes pin-9 single-wire current sharing to parallel up to three EMH350PS48-01 units within 10% load balance for N+1 redundant medical or industrial racks.
- Industrial Robotics, Telecom & Precision Instrumentation: Powers 48 VDC servo actuators, optical inspection rigs, and test-and-measurement chassis requiring remote inhibit control and Class B conducted EMI compliance.
Installation, Thermal Management & MRO Guidelines
Follow these OEM mechanical, thermal, and electrical integration guidelines to ensure full datasheet performance and regulatory compliance:
- Forced-Air Cooling & Temperature Derating: Maintain at least 12 CFM longitudinal airflow across the chassis (or pair with an XP Power top-fan cover) to sustain 354 W full load up to +50 °C; above +50 °C, derate linearly at 2.5% / °C down to 50% load at +70 °C.
- Safety Earth Bonding & EMC Grounding: Connect all four M3 chassis mounting holes (marked with earth symbols on the mechanical drawing) directly to protective safety earth to achieve EN 55011/32 Class B conducted and Class A radiated EMI performance.
- Minimum Load, Remote Sense & Mating Connectors: Maintain at least 100 mA minimum load on the 48 VDC output for tight regulation, route twisted-pair remote sense leads (pins 7 & 8 of the JST S12B-PHDSS header) to compensate up to 0.5 V line drop, and torque M4 output screws to ≤0.90 N·m (8 lb-in).