PBL500PS36C XP Power 500W 36VDC Medical Power Supply

$288.11

XP Power PBL500PS36C 500W 36VDC 13.89A enclosed medical AC-DC power supply, 80-264VAC input, featuring BF-rated 2xMOPP isolation, 5V/2A standby, and integral fan.

SKU: PBL500PS36C Category: Brand:

Description

XP Power PBL500PS36C is an enclosed, fan-cooled AC-DC switching power supply from the medical-grade PBL500 series, where the “PS36” nomenclature specifies a single 36 VDC main output and the “C” suffix designates a protective top cover with an integral temperature-controlled cooling fan. Engineered with an active power factor correction (PFC) front end switching at 65 kHz typical and a variable-frequency main converter stage (90 to 300 kHz), this unit delivers a full 500 W continuous output power (13.89 A at 36 VDC) across a universal 80 to 264 VAC (47 to 63 Hz) input range. It also provides an isolated 5.0 VDC / 2.0 A standby supply (V2) and consumes just 1 W of no-load input power when the TTL remote inhibit signal is active. Achieving 90% typical efficiency at 230 VAC full load, the supply maintains combined line and load regulation within ±2% and limits maximum output ripple and noise to 360 mV pk-pk across a 20 MHz bandwidth.

Constructed on a rugged U-channel chassis enclosed with a top fan cover measuring 101.6 mm (4.00 in) wide by 67.0 mm (2.64 in) high by 180.3 mm (7.10 in) deep and weighing 1.14 kg (2.52 lbs), the unit installs via #6-32 or M3 threaded mounting inserts. Electrical connections are made through a 2-pin Dinkle DT-35C-B01W-03 input terminal block (P1), heavy-duty M4 screw output terminals (P2), and an 8-pin Molex 87833-08 header (P3) providing Remote Sense (0.5 V compensation), TTL Inhibit, and Power OK (AC OK) supervisory signaling. Built for direct patient-contact medical systems requiring Body Floating (BF) insulation and Class II operation, it provides 4000 VAC (2xMOPP) input-to-output and input-to-ground isolation alongside a low 50 µA typical (80 µA max.) patient leakage current. Comprehensive fault protection includes latching overvoltage (112% to 140%), trip-and-restart overcurrent (105% to 140%), latching thermal shutdown, and certification to IEC/EN/ANSI/AAMI ES60601-1 and EN 55011/EN 55032 Class B up to 5000 m altitude.

Key Features

  • Continuous 500W Forced-Air Medical Output: Delivers 36 VDC at up to 13.89 A (500 W continuous) alongside an auxiliary 5.0 VDC at 2.0 A standby rail for supervisory logic and wake-up circuits.
  • BF-Rated 2xMOPP Galvanic Isolation: Engineered for Body Floating (BF) and Class II medical equipment with 4000 VAC (2xMOPP) input-to-output/ground isolation, 1500 VAC (1xMOPP) output-to-ground isolation, and 80 µA maximum patient leakage.
  • Smart Thermal-Controlled Integral Fan: The ‘C’ covered enclosure integrates an internal fan controlled by transformer T1 temperature, remaining off below 30 °C and ramping to full speed at 60 °C.
  • Comprehensive Supervisory & Sense Signals: Features an 8-pin Molex control header providing +V1/-V1 Remote Sense (compensating up to 0.5 V drop), TTL high Inhibit control, and Power OK (POK) status output.
  • Tight Regulation & Fast Transient Recovery: Maintains ±2% total output tolerance and 360 mV pk-pk maximum ripple while recovering within 1% in under 500 µs following a 25% step load change.
  • Multi-Layer Fault & Line Protection: Incorporates dual internal fuses on both Line and Neutral, 112–140% latching overvoltage protection, 105–140% trip-and-restart overcurrent protection, and latching overtemperature shutdown.
  • Global IEC 60601-1 & Class B EMC Compliance: Certified to UL/ANSI/AAMI ES60601-1, CSA C22.2 No. 60601-1, and EN/IEC 60601-1 with EN 55011/EN 55032 Class B conducted and radiated EMI performance up to 5000 m.

Technical Specifications

Manufacturer XP Power
Manufacturer Part Number (MPN) PBL500PS36C
Compatible Equipment / Series BF-Rated Medical Diagnostic Devices, Surgical Equipment, Class II Healthcare Systems & Analytical Instruments
Product Series PBL500 Series
Product Category Converter Heatsinks
Component Type Enclosed Fan-Cooled Medical AC-DC Switching Power Supply
Circuit Topology Switched-Mode Power Supply (SMPS) with Fixed-Frequency Active PFC & Variable-Frequency Main Converter
Nominal Output Voltage 36 VDC (Main Output V1); 5.0 VDC (Standby Output V2)
Rated Output Current 13.89 A (at 36 VDC Main Output V1 with integral fan); 2.0 A (at 5.0 VDC Standby Output V2)
Rated Output Power 500 W (Continuous with integral fan)
Number of Outputs 2 Outputs (Main V1 + 5 VDC Standby V2; V3 external fan tap omitted on covered ‘C’ version)
Ch1 Output Voltage & Current V1 Main Output: 36 VDC at 13.89 A (500 W Max.)
Ch2 Output Voltage & Current V2 Standby Supply: 5.0 VDC at 2.0 A (10 W)
Output Ripple & Noise (Max.) 360 mV pk-pk (1% pk-pk max., 20 MHz bandwidth with 10 µF tantalum and 0.1 µF ceramic parallel capacitors)
Output Voltage Tolerance ±2% (Combined line and load regulation, 0.1% minimum load required)
Transient Recovery Time & Deviation 4% max. deviation, recovery within 1% in <500 µs for a 25% step load change
Hold-Up Time (at 115 VAC) 20 ms typical (at 110 VAC and 500 W full load)
AC Input Voltage Range 80 to 264 VAC (Wide-range universal input)
Nominal Input Voltage 100 to 240 VAC
Input Frequency Range 47 to 63 Hz
Efficiency (at 230 VAC) 90% typical (at 230 VAC, 100% full load)
AC Input Current (at 115 VAC) 5.2 A rms (at 115 VAC, 60 Hz full load)
AC Input Current (at 230 VAC) 2.6 A rms (at 230 VAC, 50 Hz full load)
Inrush Current (Cold Start at 115 VAC) 30 A peak (at 115 VAC, 25 °C cold start)
Inrush Current (Cold Start at 230 VAC) 60 A peak (at 230 VAC, 25 °C cold start)
Power Factor Correction (PFC) Active PFC meeting EN 61000-3-2 Class A harmonic current limits
No-Load Power Consumption 1 W typical (When TTL Inhibit control is active)
Switching Frequency PFC Converter: 55 to 75 kHz (65 kHz typ., fixed); Main Converter: 90 to 300 kHz (variable); Standby Converter: 80 to 120 kHz (variable)
Mounting Style Chassis Mount (8x #6-32 UNC or 8x M3x0.5 threaded mounting inserts on bottom/sides; 4 mm max. screw penetration)
Cooling Method Forced Air via Integral Temperature-Controlled Top Fan (Starts at T1 ~30 °C, full speed at T1 ~60 °C)
Enclosure Case Number PBL500PSxxC Covered U-Channel Chassis with Top Fan Assembly
Overall Width (W) 101.6 mm (4.00 in) chassis body; 107.0 mm (4.21 in) max. across cover
Overall Height (H) 67.0 mm (2.64 in) including integral top fan assembly
Overall Depth (D) 180.3 mm (7.10 in)
Dimensional Tolerance ±0.5 mm (±0.02 in) maximum
Mounting Hole Pitch Bottom Holes: 150.0 x 60.0 mm (5.91 x 2.36 in); Side Holes: 166.0 x 20.8 mm (6.54 x 0.82 in)
Net Weight 1.14 kg (2.52 lbs) approx.
Termination / Connector Style Input P1: Dinkle DT-35C-B01W-03 M3 Screw Terminal Block; Output P2: M4x0.7 Screw Terminals; Signals P3: Molex 87833-08 8-Pin Header (mates with Molex 51110-0850)
Input Terminal Pinout Connector P1 (Dinkle M3 Screw Terminal): Pin 1 = Line (L), Pin 2 = Neutral (N); Dual internal fuses fitted in Line and Neutral
Output & Signal Terminal Pinout Output P2 (M4 Screws): Pin 1 = +V1 Main Output (36 VDC), Pin 2 = -V1 Common Return; Signals P3 (Molex 8-Pin): Pin 1 & 8 = Common Return, Pin 2 = +V1 Sense, Pin 3 = -V1 Sense, Pin 4 = POK / AC OK, Pin 5 = Inhibit, Pin 6 & 7 = +V2 5 VDC Standby
Control & Monitoring Signals Remote Sense (compensates up to 0.5 V total cable drop), Inhibit (TTL High to disable output), Power OK / AC OK (TTL High normal, 100–1000 ms turn-on delay, 1 ms loss-of-output warning)
Overload Protection (OLP) Threshold 105% to 140% of rated output current
Overload Protection (OLP) Mode Trip and restart (Hiccup / auto-recovery characteristic)
Overvoltage Protection (OVP) Threshold 112% to 140% of nominal V1 output voltage
Overvoltage Protection (OVP) Mode Latching shutdown (Recycle AC input power to reset)
Overtemperature Protection (OTP) Mode Latching thermal shutdown
Short Circuit Protection (SCP) Continuous trip and restart (Auto-recovery)
Withstand Isolation Voltage (I/P to O/P) 4000 VAC (2 x MOPP medical insulation)
Withstand Isolation Voltage (I/P to FG / Case) 4000 VAC (2 x MOPP medical insulation, suitable for Class II applications)
Withstand Isolation Voltage (O/P to FG / Case) 1500 VAC (1 x MOPP medical insulation, BF applied part rated)
Earth Leakage Current Patient Leakage Current: 50 µA typical, 80 µA maximum (at 264 VAC, 63 Hz)
Medical Isolation Rating 2xMOPP (Primary-to-Secondary & Primary-to-Ground), 1xMOPP (Secondary-to-Ground), Body Floating (BF) Applied Part Compatible
Operating Ambient Temperature Range 0 °C to +70 °C (+32 °F to +158 °F)
Full-Load Operating Temperature Limit +40 °C (+104 °F) at 100% load (500 W safety approved up to 40 °C)
High-Temperature Derating Rate Derate linearly at 2.5% per °C from 100% load (500 W) at +40 °C to 50% load (250 W) at +60 °C
Low-Line Input Derating Derate output power to 90% at 85 VAC and to 80% at 80 VAC
Altitude Derating Up to 5000 m (16,404 ft) operating altitude
Storage Temperature Range -10 °C to +85 °C (+14 °F to +185 °F)
Operating Relative Humidity 5% to 95% RH (Non-condensing)
Temperature Coefficient ±0.04% / °C
Safety Approvals (UL / North America) ANSI/AAMI ES60601-1, CSA C22.2 No. 60601-1 (Medical Electrical Equipment)
Safety Approvals (IEC / EN / TÜV) IEC 60601-1, EN 60601-1 CB Scheme (3rd Edition Medical with 2xMOPP & BF Rating)
Safety Approvals (Regional / Global) CE Marked, UKCA Compliant
EMC Conducted & Radiated Emissions EN 55011 Class B, EN 55032 Class B (Conducted and Radiated)
EMC Harmonic & Flicker Compliance EN 61000-3-2 Class A (Harmonic Currents), EN 61000-3-3 (Voltage Fluctuations and Flicker)
EMC Immunity Standards EN 61000-4-2 (ESD Level 4: ±8 kV contact, ±15 kV air, Criteria A), EN 61000-4-3 (Radiated RF 10 V/m, Criteria A), EN 61000-4-4 (EFT/Burst ±2 kV, Criteria A), EN 61000-4-5 (Surge Level 3: ±2 kV Line-to-Neutral, Criteria A), EN 61000-4-6 (Conducted RF 10 Vrms, Criteria A), EN 61000-4-8 (Magnetic Field 30 A/m, Criteria A), Voltage Dips & Interruptions (100/240 VAC per Medical Standard)
MTBF (MIL-HDBK-217F) 100,000 Hours minimum (MIL-HDBK-217F, Ground Benign, Full Load at +25 °C)
Environmental Compliance RoHS, REACH, CE & UKCA Directives

Industrial Applications & Compatible Systems

Engineered for critical patient-adjacent healthcare systems and precision instrumentation, the XP Power PBL500PS36C delivers BF-ready 36 VDC power across demanding OEM and MRO environments:

  • Body-Floating (BF) Medical Diagnostic & Surgical Systems: Powers patient-contact ultrasound scanners, surgical lasers, electrosurgical generators, and endoscopy carts requiring 4000 VAC (2xMOPP) primary isolation, 1500 VAC (1xMOPP) output-to-ground isolation, and ultra-low 80 µA max patient leakage current.
  • Clinical Laboratory & Analytical Instrumentation: Supplies low-noise 36 VDC main power and 5 VDC standby logic for hematology analyzers, automated immunoassay platforms, DNA sequencers, and mass spectrometers.
  • Home Healthcare & Mobile Point-of-Care Carts: Supports Class II (unearthed) double-insulated medical mobility equipment, respiratory therapy devices, and high-altitude medical installations operating up to 5000 meters.
  • Industrial Printing, Laser Etching & Precision Robotics: Drives 36 VDC stepper/servo motion axes, printhead arrays, and optical inspection stages requiring tight ±2% regulation and fast <500 µs transient recovery.

Installation, Thermal Management & MRO Guidelines

Ensure long-term clinical reliability and regulatory compliance by following these mechanical, electrical, and thermal integration practices:

  • Chassis Screw Penetration & Clearance: Secure the U-channel chassis using either the #6-32 UNC or M3x0.5 threaded inserts, strictly limiting mounting screw penetration to 4.0 mm maximum from the outer chassis surface to maintain internal 2xMOPP creepage and clearance distances.
  • Thermal Derating & Top Fan Airflow: Allow unobstructed intake and exhaust clearance above the integral top fan; operate at up to 500 W continuous load up to +40 °C ambient, and apply a 2.5% per °C linear derating down to 250 W at +60 °C (plus low-line derating to 90% at 85 VAC and 80% at 80 VAC).
  • Remote Sense & Minimum Load Configuration: Connect Pin 2 (+V1 Sense) and Pin 3 (-V1 Sense) on Molex header P3 directly to the load terminals using twisted-pair wiring to compensate for up to 0.5 V of cable drop, and maintain at least a 0.1% minimum load (~14 mA) to guarantee ±2% voltage regulation.