PBL500PS24C XP Power 500W 24VDC Medical Power Supply

$288.11

XP Power PBL500PS24C 500W 24VDC 20.84A enclosed medical AC-DC power supply, 80-264VAC input, featuring integral fan cooling, 5V/2A standby, and 2xMOPP BF safety.

SKU: PBL500PS24C Category: Brand:

Description

XP Power PBL500PS24C is a high-density 500 W enclosed AC-DC switching power supply from the medical-grade PBL500 series, where the “PS24” nomenclature designates a single 24 VDC main output rail and the “C” suffix specifies the factory-enclosed chassis configuration with an integral temperature-controlled cooling fan. Engineered for both Class I and Class II medical equipment requiring Body Floating (BF) applied-part insulation, the unit accepts a universal 80–264 VAC (47–63 Hz) input and delivers up to 20.84 A continuous output current (500 W) alongside an independent 5.0 VDC / 2.0 A standby rail (V2). Its active power factor corrected (PFC) front end operates at a fixed 65 kHz switching frequency paired with a variable 90–300 kHz main converter stage to achieve 90% typical efficiency at 230 VAC, ±2% combined line and load regulation, fast <500 µs transient recovery, and low output ripple of 240 mV pk-pk maximum.

Constructed in a compact covered metal housing measuring 180.3 × 101.6 × 67.0 mm (7.10 × 4.00 × 2.64 in) and weighing 1.14 kg (2.52 lbs), the supply incorporates an intelligent internal cooling fan regulated by transformer T1 thermal sensing (remaining off below ~30 °C and scaling to full speed at ~60 °C). Field wiring is streamlined via a Dinkle DT-35C-B01W-03 M3 screw terminal block (P1) for AC input, heavy-duty M4×0.7 screw terminals (P2) for the 24 VDC main output, and an 8-pin Molex 87833-08 header (P3) providing Remote Sense (0.5 V drop compensation), TTL Inhibit, Power OK (POK), and 5 VDC standby power. Comprehensive protection encompasses latching overvoltage (112–140%), trip-and-restart overcurrent (105–140%), latching overtemperature shutdown, and dual internal line/neutral fusing. Certified to IEC/ES/EN 60601-1 and CSA C22.2 No. 60601-1 up to 5000 m altitude, it provides 4000 VAC (2×MOPP) input-to-output and input-to-ground isolation, 1500 VAC (1×MOPP) output-to-ground isolation, ultra-low 50 µA typical patient leakage current, and EN 55011 / EN 55032 Class B EMI compliance.

Key Features

  • 500W Regulated 24VDC Output & 5V Standby: Delivers 24 VDC at up to 20.84 A continuous load (500 W) with ±2% total regulation, 240 mV pk-pk maximum ripple, and an auxiliary 5.0 VDC / 2.0 A standby rail.
  • 2xMOPP & BF-Rated Medical Isolation: Certified to IEC/EN/ES 60601-1 with 4000 VAC input-to-output (2xMOPP), 4000 VAC input-to-ground (2xMOPP), and 1500 VAC output-to-ground (1xMOPP) isolation for Body Floating (BF) medical systems.
  • Ultra-Low Patient Leakage Current: Restricts patient leakage current to 50 µA typical and 80 µA maximum at 264 VAC / 63 Hz, supporting Class II medical installations and sensitive patient-connected diagnostic equipment.
  • Acoustic-Optimized Thermal Fan Control: Features an integral top-mounted fan governed by internal transformer T1 temperature monitoring, remaining stationary below 30 °C and ramping smoothly to full airflow at 60 °C.
  • Remote Sense, Inhibit & Power OK Telemetry: Dedicated 8-pin Molex signal header (P3) provides 0.5 V remote sense line-drop compensation, TTL-high remote inhibit (reducing no-load power to 1 W), and Power OK monitoring.
  • Class B EMI & High Surge Immunity: Meets EN 55011 and EN 55032 Class B conducted and radiated emissions without external filtering, alongside EN 61000-3-2 Class A harmonics and Level 4 ESD immunity.
  • Wide-Altitude & Extended Thermal Operation: Operates across 0 °C to +70 °C ambient (full 500 W load up to +40 °C, linear derating to 50% at +60 °C) at altitudes up to 5000 meters.

Technical Specifications

Manufacturer XP Power
Manufacturer Part Number (MPN) PBL500PS24C
Compatible Equipment / Series BF-Rated Medical Devices, Class I & Class II Medical Systems, Surgical & Diagnostic Instrumentation
Product Series PBL500 Series
Product Category Converter Heatsinks
Component Type Enclosed Fan-Cooled Medical AC-DC Switching Power Supply
Circuit Topology Switched-Mode Power Supply (SMPS) with Active PFC, Variable-Frequency Main Converter & Standby Converter
Nominal Output Voltage 24 VDC (Main Output V1)
Rated Output Current 20.84 A (at 24 VDC with Integral Fan Cooling)
Rated Output Power 500 W Continuous
Number of Outputs 2 Outputs (Main Output V1 + Standby Output V2; V3 Fan Output Omitted on Covered C Version)
Ch1 Output Voltage & Current V1 Main Output: 24 VDC at 20.84 A (500 W)
Ch2 Output Voltage & Current V2 Standby Supply: 5.0 VDC at 2.0 A (10 W)
Output Ripple & Noise (Max.) 240 mV pk-pk (1% pk-pk max, 20 MHz bandwidth with 10 µF tantalum and 0.1 µF ceramic parallel capacitors)
Output Voltage Tolerance ±2% (Combined Line and Load Regulation; 0.1% minimum load required)
Transient Recovery Time & Deviation 4% max deviation, recovery within 1% in <500 µs (for a 25% step load change)
Hold-Up Time (at 115 VAC) 20 ms typical (at 110 VAC and 500 W full load)
AC Input Voltage Range 80 VAC to 264 VAC
Input Frequency Range 47 Hz to 63 Hz
Efficiency (at 230 VAC) 90% typical (at 230 VAC, 100% full load)
AC Input Current (at 115 VAC) 5.2 A rms typical (at 115 VAC, 60 Hz, full load)
AC Input Current (at 230 VAC) 2.6 A rms typical (at 230 VAC, 50 Hz, full load)
Inrush Current (Cold Start at 115 VAC) 30 A maximum (at 115 VAC, 25 °C cold start)
Inrush Current (Cold Start at 230 VAC) 60 A maximum (at 230 VAC, 25 °C cold start)
Power Factor Correction (PFC) Active PFC Stage (EN 61000-3-2 Class A Compliant)
No-Load Power Consumption 1 W typical (when Remote Inhibit is active)
Switching Frequency PFC Converter: 55–75 kHz (65 kHz typ. fixed); Main Converter: 90–300 kHz (variable); Standby Converter: 80–120 kHz (variable)
Mounting Style Chassis Mount (8× M3×0.5 or #6-32 threaded mounting inserts, 4 mm max screw penetration)
Cooling Method Forced Air via Integral Temperature-Controlled Top Fan (Monitored via Transformer T1: Off <30 °C, Full Speed at 60 °C)
Package / Footprint Standard Covered U-Channel Enclosure with Top-Mounted Fan (Power Density: 0.684 W/cm³ / 11.2 W/in³)
Overall Width (W) 101.6 mm (4.00 in) body width; 107.0 mm (4.21 in) overall across mounting tabs
Overall Height (H) 67.0 mm (2.64 in) including integral top fan
Overall Depth (D) 180.3 mm (7.10 in) nominal; 180.0 mm (7.09 in) drawing length
Dimensional Tolerance ±0.5 mm (±0.02 in) maximum
Mounting Hole Pitch Bottom Pattern: 150.0 × 60.0 mm (5.91 × 2.36 in); Side Pattern: 166.0 × 20.8 mm (6.54 × 0.82 in)
Net Weight 1.14 kg (2.52 lbs) approximate
Termination / Connector Style Input P1: Dinkle DT-35C-B01W-03 M3 Screw Terminal; Output P2: M4×0.7 Screw Terminals; Signals P3: Molex 87833-08 8-Pin Header (mates with Molex 51110-0850)
Input Terminal Pinout Input Connector P1: Pin 1 = Line (L), Pin 2 = Neutral (N) (Dual internal fuses fitted in Line and Neutral)
Output & Signal Terminal Pinout Output P2: Pin 1 = +V1 Main Output (24 VDC), Pin 2 = -V1 Common Return; Signal Header P3: Pin 1 & 8 = Common Return, Pin 2 = +V1 Sense, Pin 3 = -V1 Sense, Pin 4 = Power OK / AC OK, Pin 5 = Inhibit, Pin 6 & 7 = +V2 (5 VDC Standby)
Control & Monitoring Signals Remote Sense (compensates up to 0.5 V total drop), Remote Inhibit (apply TTL High to disable output), Power OK / AC OK (TTL High normal, 100–1000 ms turn-on delay, 1 ms pre-warning on AC loss)
Overload Protection (OLP) Threshold 105% to 140% of rated output current
Overload Protection (OLP) Mode Trip and Restart (Hiccup / Auto-Recovery Mode)
Overvoltage Protection (OVP) Threshold 112% to 140% of nominal V1 output voltage
Overvoltage Protection (OVP) Mode Latching Shutdown (Requires AC Input Recycle to Reset)
Overtemperature Protection (OTP) Mode Latching Thermal Shutdown
Short Circuit Protection (SCP) Continuous Trip and Restart (Auto-Recovery) with Dual Internal Input Fuses (Line & Neutral)
Withstand Isolation Voltage (I/P to O/P) 4000 VAC (2 × MOPP)
Withstand Isolation Voltage (I/P to FG / Case) 4000 VAC (2 × MOPP)
Withstand Isolation Voltage (O/P to FG / Case) 1500 VAC (1 × MOPP)
Earth Leakage Current 50 µA typical, 80 µA maximum Patient Leakage Current (at 264 VAC, 63 Hz)
Medical Isolation Rating 2 × MOPP (Input to Output & Input to Ground), 1 × MOPP (Output to Ground), Medical Body Floating (BF) Rated, Class II Applications
Operating Ambient Temperature Range 0 °C to +70 °C (+32 °F to +158 °F); Safety Approved up to +40 °C (+104 °F)
Full-Load Operating Temperature Limit +40 °C (+104 °F) at 100% Load (500 W)
High-Temperature Derating Rate Derate linearly at 2.5% per °C (12.5 W/°C) from 100% load (500 W) at +40 °C to 50% load (250 W) at +60 °C
Low-Line Input Derating Derate output power to 90% (450 W) at 85 VAC and to 80% (400 W) at 80 VAC
Altitude Derating Up to 5000 m (16,404 ft) Maximum Operating Altitude
Storage Temperature Range -10 °C to +85 °C (+14 °F to +185 °F)
Operating Relative Humidity 5% to 95% RH (Non-Condensing)
Temperature Coefficient ±0.04% / °C
Safety Approvals (UL / North America) ANSI/AAMI ES60601-1, CSA C22.2 No. 60601-1 (Medical)
Safety Approvals (IEC / EN / TÜV) IEC 60601-1, EN 60601-1, CB Scheme Certified (Medical)
Safety Approvals (Regional / Global) CE Marked, UKCA Marked
EMC Conducted & Radiated Emissions EN 55032 Class B, EN 55011 Class B (Conducted and Radiated)
EMC Harmonic & Flicker Compliance EN 61000-3-2 Class A (Harmonic Currents), EN 61000-3-3 (Voltage Flicker)
EMC Immunity Standards EN 61000-4-2 (ESD Level 4: ±8 kV contact, ±15 kV air, Criteria A), EN 61000-4-3 (Radiated: 10 V/m, Criteria A), EN 61000-4-4 (EFT/Burst: ±2 kV, Criteria A), EN 61000-4-5 (Surge Level 3: ±2 kV L-N, Criteria A), EN 61000-4-6 (Conducted RF: 10 Vrms, Criteria A), EN 61000-4-8 (Magnetic Field: 30 A/m, Criteria A), Voltage Dips & Interruptions (100 VAC & 240 VAC / 50 Hz per datasheet table)
MTBF (MIL-HDBK-217F) 100,000 Hours minimum (Full load at +25 °C, Ground Benign)
Environmental Compliance RoHS & CE / UKCA Directive Compliant

Industrial Applications & Compatible Systems

The XP Power PBL500PS24C is purpose-built for medical electronics and high-reliability industrial systems requiring Body Floating (BF) patient protection and Class II insulation:

  • Patient-Connected Medical & Surgical Equipment: Powers electrosurgical generators, patient monitors, ultrasound carts, and infusion pumps requiring 2×MOPP isolation, BF applied-part compliance, and <80 µA patient leakage current.
  • Clinical & Molecular Diagnostic Analyzers: Provides low-noise 24 VDC main power (240 mV pk-pk max) and a continuous 5 VDC / 2 A standby rail for automated hematology analyzers, PCR thermal cyclers, and clinical laboratory robotics.
  • Home Healthcare & Point-of-Care Devices: Supports Class II (unearthed) home respiratory therapy machines, dialysis systems, and mobile imaging carts operating on residential mains (EN 55011 Class B emissions) at altitudes up to 5000 meters.
  • Precision Analytical & Semiconductor Instrumentation: Drives optical inspection platforms, laser etching controllers, and precision laboratory automation utilizing remote sense compensation and TTL inhibit control.

Installation, Thermal Management & MRO Guidelines

Follow these engineering and MRO guidelines when integrating the PBL500PS24C into host enclosures:

  • Chassis Mounting & Screw Penetration Limits: Secure the enclosure using either the M3×0.5 or #6-32 threaded mounting inserts (8 places), ensuring fixing screws penetrate no more than 4.0 mm (0.16 in) from the outer surface of the chassis to maintain internal creepage and clearance barriers.
  • Thermal Derating & Low-Line Management: Provide unobstructed intake clearance above the top-mounted cooling fan; derate continuous load linearly from 500 W at +40 °C to 250 W at +60 °C, and apply low-line derating below 90 VAC (450 W at 85 VAC; 400 W at 80 VAC).
  • High-Current Wiring & Remote Sense Connection: Terminate the 20.84 A main 24 VDC output onto the M4×0.7 screw terminals (P2) using appropriately sized lugs, and wire Pin 2 (+V1 Sense) and Pin 3 (-V1 Sense) on the Molex 87833-08 header (P3) directly to the load to compensate for up to 0.5 V of cable drop.