Description
XP Power PCM50US3V3 is a medical-grade external desktop AC-DC power supply from the PCM50 series, where the nomenclature denotes a 50W-class platform (PCM50), universal AC input (U), single-output architecture (S), and a regulated 3.3 VDC output rail (3V3). Engineered with a high-reliability switch-mode topology operating at 64 kHz (PWM) and 70 kHz (PFC), this single-output unit delivers up to 40 W continuous output power and 8.00 A rated current across a universal 90 to 264 VAC (47–63 Hz) input range. It operates with 80% typical efficiency at 230 VAC full load and requires zero minimum load (0 A) for stable single-output regulation. Output quality is tightly controlled within a ±5% total regulation band (combining initial set tolerance, ±1% max. line regulation, and ±5% max. load regulation), while high-frequency ripple and noise is limited to 2% pk-pk max. (66 mVp-p) across a 20 MHz bandwidth.
Housed in an impact-resistant desktop enclosure measuring 76.0 mm (2.99 in) W × 43.0 mm (1.69 in) H × 146.0 mm (5.75 in) D with a top-mounted Power LED status indicator, the adapter connects to mains power via a standard IEC 320-C14 3-prong AC inlet and delivers DC power through a 1.22 m (48.0 in) output cable terminated with a 5-pin 180° DIN plug (Pins 1, 2, 4: Common; Pins 3, 5: +3.3 V Output; Shell: Earth Ground). Built for sensitive clinical environments, it integrates dual fuses on both Line and Neutral conductors, 112%–132% overvoltage protection (OVP), and 110%–200% auto-recovering overcurrent protection (OCP). The barrier structure provides 4000 VAC input-to-output reinforced medical isolation, 1500 VAC input-to-ground isolation, and low earth leakage current of 180 µA max. at 132 VAC / 60 Hz (300 µA max. at 264 VAC / 50 Hz). Certified to EN 60601-1 and EN 55032 Level B EMI standards, the unit delivers full 40 W power from 0 °C to +50 °C (+32 °F to +122 °F) and operates up to +70 °C (+158 °F) with linear derating to 50% load.
Key Features
- High-Current 3.3 VDC Logic Output: Supplies up to 8.00 A continuous current and 40 W rated output power with zero minimum load requirement (0 A) and a tight ±5% total regulation window.
- 4000 VAC Medical-Grade Galvanic Isolation: Provides 4000 VAC input-to-output, 1500 VAC input-to-ground, and 500 VAC output-to-ground withstand barriers alongside dual Line and Neutral fusing for EN 60601-1 patient and operator protection.
- Low Earth Leakage Current Architecture: Restricts earth leakage current to 180 µA max. at 132 VAC / 60 Hz and 300 µA max. at 264 VAC / 50 Hz for medical diagnostic and laboratory instruments.
- Universal 90–264 VAC Input & Hold-Up Resilience: Operates across global 90–264 VAC (47–63 Hz) mains with 80% typical efficiency at 230 VAC and a minimum 10 ms hold-up time at low line.
- Fast Transient Recovery & Low Ripple: Limits output voltage deviation to 5% max. with a 500 µs recovery time under a 50% step load change while holding ripple and noise within 2% pk-pk (20 MHz BW).
- Comprehensive OVP & Auto-Recovery OCP: Safeguards low-voltage 3.3 V digital circuits with 112%–132% overvoltage protection and 110%–200% overcurrent protection featuring automatic recovery upon fault removal.
- 5-Pin 180° DIN Plug & IEC 320 Inlet: Features a standard IEC 320-C14 AC receptacle and a 48.0 in (1.22 m) DC cord wired with dual parallel output pins (3 & 5) and triple return pins (1, 2 & 4).
- Wide 0 °C to +70 °C Thermal Envelope & EN 55032 Class B: Sustains 100% load up to +50 °C ambient (derating linearly to 50% at +70 °C) and meets EN 55032 Level B emissions and IEC 61000-4 Level 3 immunity.
Technical Specifications
| Manufacturer | XP Power |
|---|---|
| Manufacturer Part Number (MPN) | PCM50US3V3 |
| Product Series | PCM50 Series |
| Product Category | Medical Power Adapters |
| Component Type | External Desktop Medical AC-DC Power Adapter |
| Circuit Topology | Switched-Mode Power Supply (SMPS), Regulated Single Output with Dual Line & Neutral Fusing |
| Nominal Output Voltage | 3.3 VDC |
| Rated Output Current | 8.00 A |
| Output Current Range | 0 A to 8.00 A (No minimum load required for single-output models) |
| Rated Output Power | 40 W |
| Number of Outputs | 1 Output (Single Output) |
| Output Ripple & Noise (Max.) | 2% pk-pk max. (66 mVp-p max. at 3.3 VDC), measured at 20 MHz bandwidth |
| Output Voltage Tolerance | ±5% max. total regulation (includes initial set tolerance, line regulation, and load regulation) |
| Line Regulation | ±1% max. at full load |
| Load Regulation | ±5% max. |
| Transient Recovery Time & Deviation | 5% max. deviation, 500 µs recovery time for a 50% load step change |
| Setup Time | 2 s max. (Start-Up Delay) |
| Hold-Up Time (at 115 VAC) | 10 ms min. at low line |
| AC Input Voltage Range | 90 VAC to 264 VAC (Universal Input) |
| Input Frequency Range | 47 Hz to 63 Hz |
| Efficiency (at 230 VAC) | 80% typical (at 230 VAC, 100% full load) |
| AC Input Current (at 115 VAC) | 1.35 A max. |
| Inrush Current (Cold Start at 115 VAC) | 30 A max. at 25 °C cold start |
| Inrush Current (Cold Start at 230 VAC) | 60 A max. at 25 °C cold start |
| Power Factor (PF at 230 VAC) | 0.55 typical |
| Switching Frequency | 64 kHz typical (PWM), 70 kHz typical (PFC) |
| Mounting Style | Desktop / External Inline Enclosure |
| Cooling Method | Free Air Natural Convection |
| Overall Width (W) | 76.0 mm (2.99 in) |
| Overall Height (H) | 43.0 mm (1.69 in) |
| Overall Depth (D) | 146.0 mm (5.75 in) |
| Dimensional Tolerance | ±1.0 mm (±0.04 in) max. |
| Net Weight | 400 g to 550 g (0.88 lb to 1.21 lb) approx. |
| Termination / Connector Style | Input: IEC 320-C14 3-Pin AC Inlet; Output: 5-Pin DIN 180° Male Plug on captive DC cord (mates with standard 5-pin DIN 180° socket) |
| Input Terminal Pinout | IEC 320 Inlet: Line (L), Neutral (N), Earth Ground (E) — Both Line and Neutral internally fused |
| Output & Signal Terminal Pinout | 5-Pin DIN Plug: Pin 1 = Common, Pin 2 = Common, Pin 3 = +3.3V Output, Pin 4 = Common, Pin 5 = +3.3V Output, Connector Shell = Earth Ground (connected to AC input Earth pin) |
| Status Indicator | Top-panel AC Power-On LED Indicator |
| Lead Wire Length | 1220 mm (48.0 in / 1.22 m) approx. output DC cable |
| Overload Protection (OLP) Threshold | 110% to 200% of rated output current |
| Overload Protection (OLP) Mode | Auto-recovery upon fault removal |
| Overvoltage Protection (OVP) Threshold | 112% to 132% of nominal output voltage |
| Short Circuit Protection (SCP) | Continuous overcurrent protection with automatic recovery |
| Withstand Isolation Voltage (I/P to O/P) | 4000 VAC |
| Withstand Isolation Voltage (I/P to FG / Case) | 1500 VAC |
| Withstand Isolation Voltage (O/P to FG / Case) | 500 VAC |
| Earth Leakage Current | 180 µA max. at 132 VAC / 60 Hz; 300 µA max. at 264 VAC / 50 Hz |
| Operating Ambient Temperature Range | 0 °C to +70 °C (+32 °F to +158 °F) |
| Full-Load Operating Temperature Limit | +50 °C (+122 °F) at 100% load (40 W / 8.00 A) |
| High-Temperature Derating Rate | Derate linearly at 2.5% per °C (1.0 W / °C) from 100% load at +50 °C (+122 °F) to 50% load (20 W / 4.00 A) at +70 °C (+158 °F) |
| Storage Temperature Range | -40 °C to +80 °C (-40 °F to +176 °F) |
| Temperature Coefficient | ±0.05% / °C |
| Safety Approvals (IEC / EN / TÜV) | EN 60601-1 Medical Electrical Equipment Safety |
| Safety Approvals (Regional / Global) | CE (Meets all applicable directives), UKCA (Meets all applicable legislation) |
| EMC Conducted & Radiated Emissions | EN 55032 Level B (Conducted & Radiated Emissions) |
| EMC Harmonic & Flicker Compliance | EN 61000-3-2 Class A (Harmonic Currents), EN 61000-3-3 (Voltage Flicker) |
| EMC Immunity Standards | IEC 61000-4-2 Level 3 (ESD, Perf Criteria A), IEC 61000-4-3 Level 3 (Radiated 10 V/m, Perf Criteria A), IEC 61000-4-4 Level 3 (EFT/Burst, Perf Criteria A), IEC 61000-4-5 Level 3 (Surge, Perf Criteria A), IEC 61000-4-6 Level 3 (Conducted 10 V/m, Perf Criteria A), IEC 61000-4-8 (Magnetic Fields 10 A/m, Perf Criteria A), IEC 61000-4-11 (Voltage Dips & Interruptions: 30% 500 ms, 60% 100 ms, 100% 5000 ms, Perf Criteria A, B, B) |
| MTBF (MIL-HDBK-217F) | 100,000 Hours (100 kHrs) min. at full load and +25 °C (+77 °F) ambient |
Industrial Applications & Compatible Systems
The XP Power PCM50US3V3 is engineered for medical, clinical laboratory, and low-voltage digital instrumentation requiring a 4000 VAC isolated, high-current 3.3 VDC external power source:
- Point-of-Care Medical & Diagnostic Equipment: Powers bedside diagnostic analyzers, patient monitoring consoles, portable ultrasound docks, and ophthalmic instruments requiring EN 60601-1 compliance, dual Line/Neutral fusing, and low earth leakage current (180 µA max. at 132 VAC).
- Clinical Laboratory & In-Vitro Diagnostic (IVD) Systems: Provides regulated 3.3 VDC up to 8.00 A for benchtop hematology analyzers, PCR thermal cyclers, spectrophotometer logic boards, and automated fluid handling workstations.
- Dental & Aesthetic Treatment Consoles: Supplies isolated low-voltage DC power to tabletop dental imaging processors, optical scanners, and laser control interfaces subject to strict EN 55032 Class B EMI limits.
- FPGA, DSP & Embedded Single-Board Computer Enclosures: Feeds high-current 3.3 V digital logic rails directly via parallel-pinned 5-pin DIN connections without requiring internal AC-DC conversion stages or minimum load resistors.
Installation, Thermal Management & MRO Guidelines
To ensure reliable long-term operation and maintain medical safety compliance during OEM integration or MRO replacement, observe the following engineering guidelines:
- Parallel 5-Pin DIN Current Sharing & Mating Socket Wiring: Because the PCM50US3V3 delivers up to 8.00 A at 3.3 VDC, both output pins (Pins 3 and 5) and all three return pins (Pins 1, 2, and 4) on the mating 180° 5-pin DIN socket must be wired in parallel on the host PCB to minimize contact resistance and prevent voltage drop or pin overheating.
- Chassis Grounding & Output Shell Bond: Note that the metallic shell of the 5-pin DIN output plug is internally bonded to the IEC 320 AC input Earth pin, while the DC output faces a 500 VAC isolation barrier to ground; account for this earth bond when designing grounding schemes for sensitive analog front-ends.
- Thermal Convection & +50 °C Derating Curve: Provide unobstructed airflow around the desktop housing so heat can dissipate naturally; the unit sustains full 40 W (8.00 A) output up to +50 °C (+122 °F) ambient and must be derated linearly at 1.0 W/°C (0.20 A/°C) down to 20 W (4.00 A) at +70 °C (+158 °F).
- Hospital-Grade Mains Cord & Inrush Sizing: Pair the IEC 320-C14 inlet with an appropriately rated hospital-grade detachable mains cord for clinical environments, and size upstream branch circuit breakers to accommodate the 30 A (115 VAC) / 60 A (230 VAC) cold-start inrush current.