Description
TDK-Lambda DTM250PW360D is a high-power 250W external AC-DC desktop adapter from the DTM250-D series, engineered for dual-certified medical and industrial equipment. Within the model nomenclature, DTM250PW designates the 250W medical/ITE desktop platform, while 360D specifies the fixed 36 VDC regulated output and series revision. Operating from a universal 90 to 264 VAC (47–63 Hz) input, the unit delivers up to 6.94 A continuous output current with ±5% total regulation, 300 mVp-p maximum ripple and noise, and a switching frequency of 75 to 90 kHz. Compliant with DoE Level VI and EU Tier 2 v5 energy efficiency mandates, it achieves an average active efficiency exceeding 89% while restricting off-load power consumption to under 150 mW.
Built for patient-connected and clinical environments, the adapter integrates 4,000 VAC (2xMoPP) input-to-output galvanic isolation, 1,500 VAC (1xMoPP) input/output-to-ground isolation, and ultra-low touch current below 100 µA. Comprehensive auto-recovery protection circuits guard against overcurrent (105–160%), overvoltage (37.8–43.2 VDC), and continuous short-circuit faults via hiccup mode. Housed in a fanless, convection-cooled IP30 black polycarbonate enclosure measuring 200 × 90 × 45 mm (7.874 × 3.543 × 1.770 in) with a 3-prong IEC 320-C14 inlet and an 800 mm #16 AWG output cable terminated in a 6-pin Mini-Fit connector, it operates across -20 °C to +60 °C up to 5,000 m altitude and carries full IEC/ES/UL/CSA/EN 60601-1 and 62368-1 certifications.
Key Features
- High-Power 36 VDC Medical & ITE Output: Delivers 250 W continuous power at 36 VDC (6.94 A max) with ±5% total regulation, 300 mVp-p ripple, and zero minimum load requirement.
- Dual 2xMoPP Medical & ITE Safety Certifications: Certified to IEC/ES/UL/CSA/EN 60601-1 (suitable for Type B and BF applied parts) and IEC/UL/CSA/EN 62368-1 for global medical and industrial deployment.
- Reinforced 4 kVAC Isolation & Low Touch Current: Provides 4,000 VAC (2xMoPP) input-to-output withstand voltage, 1,500 VAC (1xMoPP) ground isolation, and less than 100 µA touch current.
- DoE Level VI & EU Tier 2 Efficiency Compliance: Exceeds 89% average active efficiency across 25%–100% loads while consuming under 150 mW in no-load standby mode.
- Comprehensive Auto-Recovery Fault Protection: Incorporates hiccup-mode continuous short-circuit protection, 105–160% overcurrent limiting, and 37.8–43.2 V overvoltage protection with automatic recovery.
- High-Altitude & Wide-Temperature Operation: Operates reliably from -20 °C to +60 °C (full load up to +40 °C) and up to 5,000 m elevation using silent, fanless convection cooling.
- High-Current 6-Pin Mini-Fit Output Interface: Equipped with a 3-prong IEC 320-C14 AC inlet and an 800 mm UL 2464 #16 AWG cable terminated in a 6-pin Molex Mini-Fit P4.20 plug.
Technical Specifications
| Manufacturer | TDK-Lambda |
|---|---|
| Manufacturer Part Number (MPN) | DTM250PW360D |
| Product Series | DTM250-D Series |
| Product Category | Medical Power Adapters |
| Component Type | External Desktop AC-DC Power Supply (Class I) |
| Nominal Output Voltage | 36 VDC |
| Output Voltage Adjustment | Fixed (None) |
| Rated Output Current | 6.94 A |
| Minimum Load Requirement | 0 A (No minimum load required) |
| Rated Output Power | 250 W |
| Output Ripple & Noise (20 MHz BW) | 300 mVp-p max (measured with 0.1 µF ceramic & 47 µF electrolytic capacitors in parallel) |
| Total Output Regulation | ±5% (includes line regulation, load regulation, and initial voltage accuracy) |
| Temperature Coefficient | ±0.05%/°C |
| Hold-Up Time (at 115 VAC) | 10 ms typical |
| Parallel Operation | Not supported |
| AC Input Voltage Range | 90 to 264 VAC (Universal) |
| Input Frequency Range | 47 to 63 Hz |
| AC Input Current (at 115 VAC) | 2.5 A max |
| AC Input Current (at 230 VAC) | 1.3 A max |
| Inrush Current (Cold Start at 230 VAC) | < 80 A (at 25 °C / 77 °F ambient) |
| Average Active Efficiency | > 89% (measured at 25%, 50%, 75%, and 100% load) |
| Energy Efficiency Compliance | DoE Level VI, EU Tier 2 v5 |
| No-Load Power Consumption | < 150 mW (< 0.15 W) |
| Switching Frequency | 75 to 90 kHz (at full load) |
| Overcurrent Protection (OCP) Threshold | 105% to 160% of rated current |
| Overcurrent Protection (OCP) Mode | Auto-Recovery |
| Overvoltage Protection (OVP) Threshold | 37.8 to 43.2 VDC |
| Overvoltage Protection (OVP) Mode | Auto-Recovery |
| Short-Circuit Protection (SCP) | Continuous, Hiccup Mode (Auto-Recovery) |
| Withstand Isolation Voltage (I/P to O/P) | 4,000 VAC for 1 minute (2xMoPP) |
| Withstand Isolation Voltage (I/P to FG) | 1,500 VAC for 1 minute (1xMoPP) |
| Withstand Isolation Voltage (O/P to FG) | 1,500 VAC for 1 minute (1xMoPP) |
| Isolation Resistance | 20 MΩ min |
| Touch / Leakage Current | < 100 µA |
| Medical Applied Part Suitability | Type B and Type BF rated equipment (2xMoPP) |
| Operating Temperature Range | -20 °C to +60 °C (-4 °F to +140 °F) with derating |
| Full-Load Operating Temperature Range | 0 °C to +40 °C (+32 °F to +104 °F) |
| High-Temperature Derating Rate | Linear derating from 100% load at +40 °C (+104 °F) to 50% load at +60 °C (+140 °F) (2.5%/°C) |
| Low-Temperature Derating Rate | Linear derating from 100% load at 0 °C (+32 °F) to 80% load at -20 °C (-4 °F) (1.0%/°C) |
| Storage Temperature Range | -20 °C to +85 °C (-4 °F to +185 °F) |
| Operating Relative Humidity | 0% to 95% RH (non-condensing) |
| Maximum Operating Altitude | 5,000 m (16,404 ft) |
| MTBF (Bellcore / Telcordia TR-332) | 200,000 hours (at 100% load, 25 °C / 77 °F ambient) |
| Safety Certifications (Medical) | IEC 60601-1, ES 60601-1, UL 60601-1, CSA C22.2 No. 60601-1, EN 60601-1 |
| Safety Certifications (ITE / Industrial) | IEC 62368-1, UL 62368-1, CSA C22.2 No. 62368-1, EN 62368-1, IEC/EN/UL 60950-1 |
| Regional Conformity Marks | CE Mark (LVD, EMC, and RoHS) |
| Conducted & Radiated EMI (Europe) | EN 55032 Class B, EN 55011 Class B |
| Conducted & Radiated EMI (North America) | FCC Part 15 Class B, FCC Part 18 Class B |
| Medical EMC Immunity Standard | IEC 60601-1-2 4th Edition (2014) |
| Status Indicator | Green LED (DC Output On) |
| AC Input Connector | IEC 320-C14 (3-prong grounded inlet) |
| DC Output Cable Specification | 800 mm (31.5 in) length, UL 2464 #16 AWG |
| DC Output Connector | 6-Pin Mini-Fit P4.20 Plug (Mates with Molex Mini-Fit 39-28-1063 or equivalent) |
| DC Output Connector Pinout | Pins 1, 2, 3: -Vout (Return) | Pins 4, 5, 6: +Vout (+36 VDC) |
| Cooling Method | Natural Convection (Fanless) |
| Ingress Protection Rating | IP30 |
| Enclosure Material & Color | Black Polycarbonate Plastic Case |
| Overall Length (L) | 200 mm (7.874 in) |
| Overall Width (W) | 90 mm (3.543 in) |
| Overall Height (H) | 45 mm (1.770 in) |
| Dimensional Tolerance | ±1 mm (±0.039 in) |
| Net Weight | 1,300 g (2.87 lbs / 45.86 oz) |
| Manufacturer Warranty | 3 Years |
Industrial Applications & Compatible Systems
TDK-Lambda DTM250PW360D serves as an external high-wattage 36 VDC power source for clinical, laboratory, and portable industrial instrumentation requiring 2xMoPP isolation:
- Patient-Connected Medical & Dental Systems: Portable diagnostic imaging units, surgical suction/irrigation pumps, dental delivery consoles, and Type B/BF patient monitoring carts requiring 4 kVAC isolation and <100 µA touch current.
- Portable Test & Measurement Instrumentation: Benchtop spectrum analyzers, portable RF test sets, battery-backed field calibrators, and high-altitude mobile diagnostic equipment rated up to 5,000 m.
- Industrial Automation & Laboratory Analyzers: Automated liquid handling workstations, compact 36 V stepper/servo actuators, and analytical chemistry spectrometers certified under IEC/UL 62368-1.
- Broadcast & Communications Equipment: Field-deployable broadcast encoders, portable studio lighting controllers, and mobile telecom base-station test rigs needing fanless Class B EMI compliance.
Installation, Thermal Management & MRO Guidelines
Installation & Engineering Guidelines: Because the DTM250PW360D relies entirely on fanless natural convection cooling inside a sealed IP30 housing, position the adapter horizontally on a flat surface with at least 25 mm of unobstructed airflow clearance around all sides and avoid enclosing it inside unventilated pockets or beneath insulating drapes. Maintain ambient temperatures between 0 °C and +40 °C for full 250 W output; above +40 °C, derate the load linearly at 2.5%/°C down to 125 W at +60 °C, and below 0 °C derate at 1.0%/°C down to 200 W at -20 °C. Ensure the host chassis receptacle matches the 6-pin Molex Mini-Fit Jr. (39-28-1063) pinout (Pins 1–3 = -Vout, Pins 4–6 = +36 VDC) to distribute the 6.94 A load across all three parallel contact pairs, and note that parallel output connection of multiple units is not supported.