DTM250PW360D TDK-Lambda 250W 36VDC Medical Desktop PSU

$257.90

TDK-Lambda DTM250PW360D 250W 36VDC 6.94A medical & ITE desktop power supply, 90-264VAC input, >89% efficiency, featuring 2xMoPP isolation & DoE Level VI compliance.

SKU: DTM250PW360D Category: Brand:

Description

TDK-Lambda DTM250PW360D is a high-power 250W external AC-DC desktop adapter from the DTM250-D series, engineered for dual-certified medical and industrial equipment. Within the model nomenclature, DTM250PW designates the 250W medical/ITE desktop platform, while 360D specifies the fixed 36 VDC regulated output and series revision. Operating from a universal 90 to 264 VAC (47–63 Hz) input, the unit delivers up to 6.94 A continuous output current with ±5% total regulation, 300 mVp-p maximum ripple and noise, and a switching frequency of 75 to 90 kHz. Compliant with DoE Level VI and EU Tier 2 v5 energy efficiency mandates, it achieves an average active efficiency exceeding 89% while restricting off-load power consumption to under 150 mW.

Built for patient-connected and clinical environments, the adapter integrates 4,000 VAC (2xMoPP) input-to-output galvanic isolation, 1,500 VAC (1xMoPP) input/output-to-ground isolation, and ultra-low touch current below 100 µA. Comprehensive auto-recovery protection circuits guard against overcurrent (105–160%), overvoltage (37.8–43.2 VDC), and continuous short-circuit faults via hiccup mode. Housed in a fanless, convection-cooled IP30 black polycarbonate enclosure measuring 200 × 90 × 45 mm (7.874 × 3.543 × 1.770 in) with a 3-prong IEC 320-C14 inlet and an 800 mm #16 AWG output cable terminated in a 6-pin Mini-Fit connector, it operates across -20 °C to +60 °C up to 5,000 m altitude and carries full IEC/ES/UL/CSA/EN 60601-1 and 62368-1 certifications.

Key Features

  • High-Power 36 VDC Medical & ITE Output: Delivers 250 W continuous power at 36 VDC (6.94 A max) with ±5% total regulation, 300 mVp-p ripple, and zero minimum load requirement.
  • Dual 2xMoPP Medical & ITE Safety Certifications: Certified to IEC/ES/UL/CSA/EN 60601-1 (suitable for Type B and BF applied parts) and IEC/UL/CSA/EN 62368-1 for global medical and industrial deployment.
  • Reinforced 4 kVAC Isolation & Low Touch Current: Provides 4,000 VAC (2xMoPP) input-to-output withstand voltage, 1,500 VAC (1xMoPP) ground isolation, and less than 100 µA touch current.
  • DoE Level VI & EU Tier 2 Efficiency Compliance: Exceeds 89% average active efficiency across 25%–100% loads while consuming under 150 mW in no-load standby mode.
  • Comprehensive Auto-Recovery Fault Protection: Incorporates hiccup-mode continuous short-circuit protection, 105–160% overcurrent limiting, and 37.8–43.2 V overvoltage protection with automatic recovery.
  • High-Altitude & Wide-Temperature Operation: Operates reliably from -20 °C to +60 °C (full load up to +40 °C) and up to 5,000 m elevation using silent, fanless convection cooling.
  • High-Current 6-Pin Mini-Fit Output Interface: Equipped with a 3-prong IEC 320-C14 AC inlet and an 800 mm UL 2464 #16 AWG cable terminated in a 6-pin Molex Mini-Fit P4.20 plug.

Technical Specifications

Manufacturer TDK-Lambda
Manufacturer Part Number (MPN) DTM250PW360D
Product Series DTM250-D Series
Product Category Medical Power Adapters
Component Type External Desktop AC-DC Power Supply (Class I)
Nominal Output Voltage 36 VDC
Output Voltage Adjustment Fixed (None)
Rated Output Current 6.94 A
Minimum Load Requirement 0 A (No minimum load required)
Rated Output Power 250 W
Output Ripple & Noise (20 MHz BW) 300 mVp-p max (measured with 0.1 µF ceramic & 47 µF electrolytic capacitors in parallel)
Total Output Regulation ±5% (includes line regulation, load regulation, and initial voltage accuracy)
Temperature Coefficient ±0.05%/°C
Hold-Up Time (at 115 VAC) 10 ms typical
Parallel Operation Not supported
AC Input Voltage Range 90 to 264 VAC (Universal)
Input Frequency Range 47 to 63 Hz
AC Input Current (at 115 VAC) 2.5 A max
AC Input Current (at 230 VAC) 1.3 A max
Inrush Current (Cold Start at 230 VAC) < 80 A (at 25 °C / 77 °F ambient)
Average Active Efficiency > 89% (measured at 25%, 50%, 75%, and 100% load)
Energy Efficiency Compliance DoE Level VI, EU Tier 2 v5
No-Load Power Consumption < 150 mW (< 0.15 W)
Switching Frequency 75 to 90 kHz (at full load)
Overcurrent Protection (OCP) Threshold 105% to 160% of rated current
Overcurrent Protection (OCP) Mode Auto-Recovery
Overvoltage Protection (OVP) Threshold 37.8 to 43.2 VDC
Overvoltage Protection (OVP) Mode Auto-Recovery
Short-Circuit Protection (SCP) Continuous, Hiccup Mode (Auto-Recovery)
Withstand Isolation Voltage (I/P to O/P) 4,000 VAC for 1 minute (2xMoPP)
Withstand Isolation Voltage (I/P to FG) 1,500 VAC for 1 minute (1xMoPP)
Withstand Isolation Voltage (O/P to FG) 1,500 VAC for 1 minute (1xMoPP)
Isolation Resistance 20 MΩ min
Touch / Leakage Current < 100 µA
Medical Applied Part Suitability Type B and Type BF rated equipment (2xMoPP)
Operating Temperature Range -20 °C to +60 °C (-4 °F to +140 °F) with derating
Full-Load Operating Temperature Range 0 °C to +40 °C (+32 °F to +104 °F)
High-Temperature Derating Rate Linear derating from 100% load at +40 °C (+104 °F) to 50% load at +60 °C (+140 °F) (2.5%/°C)
Low-Temperature Derating Rate Linear derating from 100% load at 0 °C (+32 °F) to 80% load at -20 °C (-4 °F) (1.0%/°C)
Storage Temperature Range -20 °C to +85 °C (-4 °F to +185 °F)
Operating Relative Humidity 0% to 95% RH (non-condensing)
Maximum Operating Altitude 5,000 m (16,404 ft)
MTBF (Bellcore / Telcordia TR-332) 200,000 hours (at 100% load, 25 °C / 77 °F ambient)
Safety Certifications (Medical) IEC 60601-1, ES 60601-1, UL 60601-1, CSA C22.2 No. 60601-1, EN 60601-1
Safety Certifications (ITE / Industrial) IEC 62368-1, UL 62368-1, CSA C22.2 No. 62368-1, EN 62368-1, IEC/EN/UL 60950-1
Regional Conformity Marks CE Mark (LVD, EMC, and RoHS)
Conducted & Radiated EMI (Europe) EN 55032 Class B, EN 55011 Class B
Conducted & Radiated EMI (North America) FCC Part 15 Class B, FCC Part 18 Class B
Medical EMC Immunity Standard IEC 60601-1-2 4th Edition (2014)
Status Indicator Green LED (DC Output On)
AC Input Connector IEC 320-C14 (3-prong grounded inlet)
DC Output Cable Specification 800 mm (31.5 in) length, UL 2464 #16 AWG
DC Output Connector 6-Pin Mini-Fit P4.20 Plug (Mates with Molex Mini-Fit 39-28-1063 or equivalent)
DC Output Connector Pinout Pins 1, 2, 3: -Vout (Return) | Pins 4, 5, 6: +Vout (+36 VDC)
Cooling Method Natural Convection (Fanless)
Ingress Protection Rating IP30
Enclosure Material & Color Black Polycarbonate Plastic Case
Overall Length (L) 200 mm (7.874 in)
Overall Width (W) 90 mm (3.543 in)
Overall Height (H) 45 mm (1.770 in)
Dimensional Tolerance ±1 mm (±0.039 in)
Net Weight 1,300 g (2.87 lbs / 45.86 oz)
Manufacturer Warranty 3 Years

Industrial Applications & Compatible Systems

TDK-Lambda DTM250PW360D serves as an external high-wattage 36 VDC power source for clinical, laboratory, and portable industrial instrumentation requiring 2xMoPP isolation:

  • Patient-Connected Medical & Dental Systems: Portable diagnostic imaging units, surgical suction/irrigation pumps, dental delivery consoles, and Type B/BF patient monitoring carts requiring 4 kVAC isolation and <100 µA touch current.
  • Portable Test & Measurement Instrumentation: Benchtop spectrum analyzers, portable RF test sets, battery-backed field calibrators, and high-altitude mobile diagnostic equipment rated up to 5,000 m.
  • Industrial Automation & Laboratory Analyzers: Automated liquid handling workstations, compact 36 V stepper/servo actuators, and analytical chemistry spectrometers certified under IEC/UL 62368-1.
  • Broadcast & Communications Equipment: Field-deployable broadcast encoders, portable studio lighting controllers, and mobile telecom base-station test rigs needing fanless Class B EMI compliance.

Installation, Thermal Management & MRO Guidelines

Installation & Engineering Guidelines: Because the DTM250PW360D relies entirely on fanless natural convection cooling inside a sealed IP30 housing, position the adapter horizontally on a flat surface with at least 25 mm of unobstructed airflow clearance around all sides and avoid enclosing it inside unventilated pockets or beneath insulating drapes. Maintain ambient temperatures between 0 °C and +40 °C for full 250 W output; above +40 °C, derate the load linearly at 2.5%/°C down to 125 W at +60 °C, and below 0 °C derate at 1.0%/°C down to 200 W at -20 °C. Ensure the host chassis receptacle matches the 6-pin Molex Mini-Fit Jr. (39-28-1063) pinout (Pins 1–3 = -Vout, Pins 4–6 = +36 VDC) to distribute the 6.94 A load across all three parallel contact pairs, and note that parallel output connection of multiple units is not supported.