Description
XP Power PBL500PS48B is a high-efficiency, Class II U-channel AC-DC switching power supply belonging to the medical-grade PBL500 series, where the PS48 designation specifies a single 48 VDC primary output and the B suffix denotes the open U-channel mechanical chassis equipped with an auxiliary 12 VDC fan header. Accepting a universal 80 to 264 VAC input (47 to 63 Hz) with active Power Factor Correction (PFC), this unit delivers 500 W (10.42 A) under 51 m³/h (30 CFM) forced-air cooling or 250 W (5.21 A) under natural convection. Engineered with a fixed-frequency PFC stage (65 kHz typical) and variable-frequency main and standby converters, it achieves 90% typical efficiency at 230 VAC full load while maintaining ±2% total line/load regulation, 480 mV peak-to-peak maximum ripple and noise, and a 5.0 VDC / 2.0 A always-on standby rail.
Housed in a compact 180.3 × 101.6 × 39.6 mm (7.10 × 4.00 × 1.56 in) aluminum U-channel chassis weighing 1.01 kg (2.23 lbs), the power supply utilizes a Dinkle screw terminal block for AC line/neutral input, M4×0.7 screw posts for the high-current 48 VDC bus, and keyed Molex and JST headers for remote sensing, Power OK (POK), TTL inhibit, and external cooling fan power. Comprehensive protection includes latching overvoltage protection (112% to 140%), trip-and-restart overcurrent protection (105% to 140%), and latching overtemperature shutdown. Certified to IEC/EN/ES 60601-1 and CSA C22.2 No. 60601-1 for Type BF patient-connected medical devices, it provides 4000 VAC (2×MOPP) input-to-output galvanic isolation, 80 µA maximum patient leakage current, Class B conducted/radiated EMI compliance, and full operation up to 5,000 m altitude across 0 °C to +70 °C.
Key Features
- Dual-Rating Forced-Air & Convection Output: Delivers 500 W (10.42 A at 48 VDC) with 30 CFM user-supplied airflow or 250 W (5.21 A) under natural convection, achieving 90% typical efficiency.
- BF-Rated Medical Isolation (2×MOPP): Provides 4000 VAC input-to-output and input-to-ground isolation (2×MOPP), 1500 VAC output-to-ground (1×MOPP), and low 50 µA typical patient leakage current.
- Integrated Auxiliary & Fan Supply Rails: Includes a dedicated 5.0 VDC / 2.0 A standby output (V2) for system logic and an independent 12.0 VDC / 0.3 A output (V3) to power external cooling fans.
- Comprehensive System Control & Monitoring: Features remote output voltage sensing compensating up to 0.5 V total lead drop, TTL-high inhibit control, and Power OK (POK) with 1 ms power-fail warning.
- Low-Profile U-Channel Thermal Chassis: Measures 180.3 × 101.6 × 39.6 mm (7.10 × 4.00 × 1.56 in) with dual-standard #6-32 and M3×0.5 threaded mounting inserts for flexible OEM equipment integration.
- Class B EMI & High-Altitude Compliance: Meets EN 55032/EN 55011 Class B conducted and radiated emissions without external filtering and carries safety certification up to 5,000 m (16,404 ft) operating altitude.
- Multi-Layer Fault Protection Suite: Safeguards downstream clinical electronics with latching overvoltage protection (112%–140%), auto-recovery overcurrent protection (105%–140%), latching thermal shutdown, and dual line/neutral internal fuses.
Technical Specifications
| Manufacturer | XP Power |
|---|---|
| Manufacturer Part Number (MPN) | PBL500PS48B |
| Product Series | PBL500 Series |
| Product Category | Converter Heatsinks |
| Component Type | U-Channel Chassis Mount Medical AC-DC Switching Power Supply |
| Circuit Topology | Active PFC + Variable-Frequency Main & Standby Switching Converters (Class II) |
| Nominal Output Voltage | 48 VDC (V1 Main Output) |
| Rated Output Current | 10.42 A (Forced-Air Cooled at 51 m³/h / 30 CFM); 5.21 A (Convection Cooled) |
| Rated Output Power | 500 W (Forced-Air Cooled); 250 W (Convection Cooled) |
| Number of Outputs | 3 Outputs (Main V1 + Standby V2 + Fan V3) |
| Ch1 Output Voltage & Current | 48 VDC @ 10.42 A max (Forced Air) / 5.21 A max (Convection) |
| Ch2 Output Voltage & Current | 5.0 VDC @ 2.0 A max (Always-On Standby Supply V2) |
| Ch3 Output Voltage & Current | 12.0 VDC @ 0.3 A (300 mA) max (Fan Supply V3) |
| Output Ripple & Noise (Max.) | 480 mV pk-pk (20 MHz bandwidth with 10 µF tantalum & 0.1 µF ceramic capacitors) |
| Output Voltage Tolerance | ±2% (combined line and load regulation; 0.1% minimum load required) |
| Transient Recovery Time & Deviation | 4% max deviation, recovery within 1% in less than 500 µs (25% step load change) |
| Setup Time | 100 ms to 1000 ms (Power OK turn-on delay) |
| Hold-Up Time (at 115 VAC) | 20 ms min (measured at 110 VAC and 500 W full load) |
| AC Input Voltage Range | 80 to 264 VAC (Single-Phase, Class II) |
| Input Frequency Range | 47 to 63 Hz |
| Efficiency (at 230 VAC) | 90% typical (at 100% full load) |
| AC Input Current (at 115 VAC) | 5.2 A rms max (at 60 Hz full load) |
| AC Input Current (at 230 VAC) | 2.6 A rms max (at 50 Hz full load) |
| Inrush Current (Cold Start at 115 VAC) | 30 A max (at 25 °C ambient) |
| Inrush Current (Cold Start at 230 VAC) | 60 A max (at 25 °C ambient) |
| Power Factor Correction (PFC) | Active PFC (EN 61000-3-2 Class A compliant) |
| No-Load Power Consumption | 1 W max (when remote inhibit is active) |
| Switching Frequency | PFC: 55–75 kHz (65 kHz typ fixed); Main: 90–300 kHz (variable); Standby: 80–120 kHz (variable) |
| Mounting Style | Chassis Mount (U-Channel with 8× #6-32 & 8× M3×0.5 threaded inserts, 4 mm max screw penetration) |
| Cooling Method | Forced-Air Cooled (51 m³/h / 30 CFM user-supplied for 500 W) or Natural Convection (250 W) |
| Overall Width (W) | 101.6 mm (4.00 in) [100.1 mm / 3.94 in U-channel nominal] |
| Overall Height (H) | 39.6 mm (1.56 in) [39.5 mm / 1.56 in U-channel nominal] |
| Overall Depth (D) | 180.3 mm (7.10 in) [180.0 mm / 7.09 in U-channel nominal] |
| Dimensional Tolerance | ±0.5 mm (±0.02 in) maximum |
| Net Weight | 1.01 kg (2.23 lbs) |
| Termination / Connector Style | P1: Dinkle DT-35C-B01W-03 M3 Screw Terminal; P2: M4×0.7 Screw Terminals; P3: Molex 87833-08; P4: JST S2B-ZR-3.4 |
| Input Terminal Pinout | Connector P1 (Dinkle DT-35C-B01W-03): Pin 1 = Line (L), Pin 2 = Neutral (N) |
| Output & Signal Terminal Pinout | P2 Main Output: Pin 1 = +V1 (48V), Pin 2 = -V1 (Common Return); P3 Signals/Standby (mates Molex 51110-0850): Pin 1 = Common Return, Pin 2 = +V1 Sense, Pin 3 = -V1 Sense, Pin 4 = POK, Pin 5 = Inhibit, Pin 6 = +V2 (5V Standby), Pin 7 = +V2 (5V Standby), Pin 8 = Common Return; P4 Fan (mates JST ZHR-2): Pin 1 = Common Return, Pin 2 = +V3 (12V Fan) |
| Control & Monitoring Signals | Remote Sense (compensates up to 0.5 V total drop), Inhibit (TTL high to inhibit), Power OK / POK (TTL high normal, 1 ms power-fail warning) |
| Overload Protection (OLP) Threshold | 105% to 140% of rated output current |
| Overload Protection (OLP) Mode | Trip and automatic restart (hiccup characteristic) |
| Overvoltage Protection (OVP) Threshold | 112% to 140% of nominal output voltage |
| Overvoltage Protection (OVP) Mode | Latching shutdown (requires AC input cycle to reset) |
| Overtemperature Protection (OTP) Mode | Latching thermal shutdown |
| Short Circuit Protection (SCP) | Protected via internal dual line/neutral fuses and overcurrent trip-and-restart circuitry |
| Withstand Isolation Voltage (I/P to O/P) | 4000 VAC (2×MOPP reinforced medical isolation) |
| Withstand Isolation Voltage (I/P to FG / Case) | 4000 VAC (2×MOPP reinforced medical isolation) |
| Withstand Isolation Voltage (O/P to FG / Case) | 1500 VAC (1×MOPP medical isolation) |
| Earth Leakage Current | Patient Leakage Current: 50 µA typical, 80 µA max (at 264 VAC, 63 Hz) |
| Medical Isolation Rating | 2×MOPP (Input-to-Output & Input-to-Ground), 1×MOPP (Output-to-Ground), Type BF Rated |
| Operating Ambient Temperature Range | 0 °C to +70 °C (+32 °F to +158 °F) |
| Full-Load Operating Temperature Limit | +40 °C (+104 °F) safety-approved full-load limit |
| High-Temperature Derating Rate | Derate linearly from 100% load at +40 °C (+104 °F) to 50% load at +60 °C (+140 °F) |
| Low-Line Input Derating | Derate output power linearly to 90% at 85 VAC and 80% at 80 VAC |
| Altitude Derating | Rated for operation up to 5,000 m (16,404 ft) |
| Storage Temperature Range | -10 °C to +85 °C (+14 °F to +185 °F) |
| Operating Relative Humidity | 5% to 95% RH (non-condensing) |
| Temperature Coefficient | ±0.04 %/°C |
| Safety Approvals (UL / North America) | UL / ES 60601-1, CSA C22.2 No. 60601-1 (Medical) |
| Safety Approvals (IEC / EN / TÜV) | IEC 60601-1, EN 60601-1, CB Scheme (Medical) |
| Safety Approvals (Regional / Global) | CE, UKCA |
| EMC Conducted & Radiated Emissions | EN 55032 Class B, EN 55011 Class B |
| EMC Harmonic & Flicker Compliance | EN 61000-3-2 Class A (Harmonic Currents), EN 61000-3-3 (Voltage Flicker) |
| EMC Immunity Standards | EN 61000-4-2 Level 4 (±8 kV contact, ±15 kV air, Criteria A), EN 61000-4-3 (10 V/m, Criteria A), EN 61000-4-4 (±2 kV EFT/burst, Criteria A), EN 61000-4-5 Level 3 (±2 kV L-N surge, Criteria A), EN 61000-4-6 (10 Vrms conducted, Criteria A), EN 61000-4-8 (30 A/m magnetic field, Criteria A), EN 61000-4-11 (voltage dips & interruptions) |
| MTBF (MIL-HDBK-217F) | 100,000 hours minimum (full load at +25 °C ambient, Ground Benign) |
| Environmental Compliance | RoHS, REACH |
Industrial Applications & Compatible Systems
The XP Power PBL500PS48B is purpose-built for Type BF patient-connected clinical instrumentation and high-reliability industrial 48 VDC bus architectures requiring Class II double insulation and low leakage current:
- Medical Diagnostic & Imaging Systems: Powers 48 VDC motor drives, gantry actuators, and ultrasound transducers in hospital diagnostic consoles where 2×MOPP isolation and 80 µA max patient leakage are mandatory.
- Surgical & Therapy Equipment: Supplies clean, Class B EMI-compliant 48 VDC bulk power and 5 VDC standby logic for electrosurgical generators, patient monitors, and respiratory ventilators.
- Clinical Laboratory & Analytical Instrumentation: Drives automated blood analyzers, centrifuges, and chromatography fluidics requiring high peak efficiency and operation up to 5,000 m altitude.
- Home Healthcare & Class II Medical Devices: Eliminates protective earth ground requirements in portable or home-use medical equipment via Class II double-insulated construction.
Installation, Thermal Management & MRO Guidelines
Follow these thermal, mechanical, and electrical integration practices when installing or replacing this U-channel power supply:
- Airflow & Thermal Derating Management: Supply at least 51 m³/h (30 CFM) of forced airflow across the U-channel chassis—optionally powered from the onboard 12 VDC / 0.3 A P4 header—to sustain the full 500 W rating up to +40 °C, or limit continuous load to 250 W under natural convection and derate linearly to 50% at +60 °C.
- Chassis Mounting Screw Depth Limit: Secure the U-channel using the bottom or side #6-32 or M3×0.5 threaded inserts, strictly limiting fastener penetration to 4.0 mm maximum from the outer chassis surface to maintain internal creepage and clearance distances.
- Remote Sense & Minimum Load Configuration: Connect P3 Pins 2 and 3 (+V1/-V1 Sense) directly to the 48 VDC point of load to compensate for up to 0.5 V of distribution drop, and ensure at least 0.1% minimum load is present to maintain ±2% regulation tolerance.
- Inrush Current & Upstream Breaker Sizing: Account for the 60 A cold-start inrush current at 230 VAC when selecting upstream medical-grade circuit breakers, noting that internal fuses are already fitted in both line and neutral poles.