DTM300PW480D1 TDK-Lambda 300W 48VDC Medical Desktop Adapter

$300.60

TDK-Lambda DTM300PW480D1 300W 48VDC 6.25A Class I medical desktop power supply, 90-264VAC input, >88% DoE Level VI efficiency, IEC 60601-1 2xMOPP certified.

SKU: DTM300PW480D1 Category: Brand:

Description

TDK-Lambda DTM300PW480D1 is a 300W external Class I AC-DC desktop power supply from the DTM300-D series engineered for medical, industrial, and test-and-measurement equipment. In TDK-Lambda’s nomenclature, the 480 suffix specifies a fixed 48 VDC regulated output rated at 6.25 A continuous current, while the D1 suffix designates a Class I grounded input utilizing a standard 3-prong IEC 320-C14 inlet. Operating across a universal 90 to 264 VAC (47–63 Hz) input range with an 83 to 96 kHz full-load switching topology, the adapter achieves greater than 88% four-point average efficiency to satisfy US DoE Level VI energy mandates and maintains off-load power draw below 0.5 W. Output voltage accuracy is held within a ±5% total regulation band with zero minimum load requirement and 300 mVp-p maximum ripple and noise.

Comprehensive fault protection includes auto-recovery overcurrent limiting (105–150%), auto-recovery overvoltage clamping (50.0–57.2 VDC), and continuous hiccup-mode short-circuit protection. For patient-connected medical compliance, the Class I isolation barrier provides 4,000 VAC (2×MOPP) input-to-output dielectric withstand and 1,500 VAC (1×MOPP) input/output-to-ground isolation with touch current kept under 100 µA. Housed in a convection-cooled IP42 black polycarbonate enclosure measuring 222 × 112 × 45 mm (8.75 × 4.40 × 1.77 in) with an 800 mm UL 2464 #16 AWG output cable terminated in a 10-pin Mini-Fit connector, it operates from -30 °C to +60 °C (-22 °F to +140 °F) and carries full IEC/ES/UL/CSA/EN 60601-1, IEC 60601-1-2 Ed4, and IEC/UL/CSA/EN 62368-1 safety approvals.

Key Features

  • 300W Regulated 48 VDC Output: Delivers 48 VDC at 6.25 A continuous current with ±5% total regulation, 300 mVp-p ripple and noise, and zero minimum load requirement.
  • Dual Medical & ITE Safety Approvals: Certified to IEC/ES/UL/CSA/EN 60601-1 with 2×MOPP isolation and IEC/UL/CSA/EN 62368-1, making it suitable for Type B and BF medical devices.
  • DoE Level VI Energy Efficiency: Achieves greater than 88% average efficiency across 25% to 100% load points while limiting no-load standby consumption to under 500 mW.
  • 4,000 VAC 2×MOPP Class I Isolation: Features a grounded IEC 320-C14 inlet providing 4,000 VAC input-to-output withstand, 1,500 VAC to ground, and touch current below 100 µA.
  • Auto-Recovery Protection Suite: Integrates 105–150% overcurrent protection, 50.0–57.2 VDC overvoltage protection, and continuous hiccup-mode short-circuit protection with automatic fault recovery.
  • Wide Operating Temperature & Altitude: Convection-cooled IP42 housing operates from -30 °C to +60 °C (-22 °F to +140 °F) and up to 5,000 m (16,404 ft) elevation.
  • High-Current 10-Pin Mini-Fit Output Harness: Equipped with an 800 mm UL 2464 #16 AWG multi-conductor cable terminated in a 10-pin 4.20 mm pitch Mini-Fit connector.

Technical Specifications

Manufacturer TDK-Lambda
Manufacturer Part Number (MPN) DTM300PW480D1
Product Series DTM300-D Series
Product Category Medical Power Adapters
Component Type External Desktop AC-DC Power Supply (Class I)
Safety Protection Class Class I (Grounded Protective Earth via IEC 320-C14)
Nominal Output Voltage 48 VDC
Output Voltage Adjustment Fixed (None)
Rated Output Current 6.25 A
Minimum Load Requirement 0 A (No minimum load required)
Rated Output Power 300 W
Total Output Regulation ±5% (includes initial voltage accuracy, line, and load regulation)
Output Ripple & Noise (Max) 300 mVp-p (20 MHz bandwidth with 0.1 µF ceramic & 47 µF electrolytic capacitors)
Temperature Coefficient ±0.05 %/°C
Hold-Up Time (at 115 VAC) 10 ms (Typical at full load)
Parallel Operation Not supported
Input Voltage Range (AC) 90 to 264 VAC (Universal)
Input Frequency Range 47 to 63 Hz
AC Input Current (at 115 VAC) 3.0 A max
AC Input Current (at 230 VAC) 1.5 A max
Inrush Current (Cold Start at 230 VAC) < 70 A (at 25 °C ambient)
Average Active Efficiency > 88% (4-point average at 25%, 50%, 75%, 100% load; DoE Level VI compliant)
No-Load Power Consumption < 500 mW (< 0.5 W)
Switching Frequency 83 to 96 kHz (Full load)
Overcurrent Protection (OCP) Threshold 105% to 150% of rated output current
Overcurrent Protection (OCP) Mode Auto-Recovery
Overvoltage Protection (OVP) Threshold 50.0 to 57.2 VDC
Overvoltage Protection (OVP) Mode Auto-Recovery
Short-Circuit Protection (SCP) Continuous — Hiccup Mode (Auto-Recovery)
Withstand Isolation Voltage (I/P to O/P) 4,000 VAC for 1 minute (2×MOPP)
Withstand Isolation Voltage (I/P to FG / Ground) 1,500 VAC for 1 minute (1×MOPP)
Withstand Isolation Voltage (O/P to FG / Ground) 1,500 VAC for 1 minute (1×MOPP)
Isolation Resistance 20 MΩ minimum
Touch / Leakage Current < 100 µA
Operating Ambient Temperature Range -30 °C to +60 °C (-22 °F to +140 °F)
Full-Load Operating Temperature Window 0 °C to +40 °C (+32 °F to +104 °F) at 100% rated load
High-Temperature Derating Rate Linear derating from 100% load at +40 °C (+104 °F) to 50% load (150 W) at +60 °C (+140 °F) (2.5 %/°C)
Low-Temperature Derating Rate Linear derating from 100% load at 0 °C (+32 °F) to 70% load (210 W) at -30 °C (-22 °F) (1.0 %/°C)
Storage Temperature Range -30 °C to +85 °C (-22 °F to +185 °F)
Operating Relative Humidity 0% to 95% RH (non-condensing)
Maximum Operating Altitude 5,000 m (16,404 ft)
MTBF (Bellcore / Telcordia TR-332) 200,000 hours minimum (100% load at +25 °C ambient)
Medical Safety Certifications IEC/ES/UL/CSA/EN 60601-1 (2×MOPP, suitable for Type B & BF equipment)
ITE & Industrial Safety Certifications IEC/UL/CSA/EN 62368-1, IEC/UL/CSA/EN 60950-1
Regional Compliance Marks CE Mark (LVD, EMC, RoHS), US DoE Level VI
Conducted & Radiated EMI (CISPR / EN) EN 55032 Class B, EN 55011 Class B
Conducted & Radiated EMI (FCC) FCC Part 15 Class B, FCC Part 18 Class B
Medical EMC Immunity Standard IEC 60601-1-2 Edition 4:2014
Status Indicator Green LED (illuminates when DC output is active)
AC Input Connector IEC 320-C14 (3-prong grounded male inlet)
DC Output Cable Specification 800 mm (31.5 in) length, UL 2464 #16 AWG multi-conductor cable
DC Output Connector 10-Pin Mini-Fit Receptacle, 4.20 mm pitch (mates with Molex Mini-Fit 39-28-1103 or equivalent)
DC Output Pinout — Return / GND (Pins 1–5) Pin 1: Brown (GND), Pin 2: Black (GND), Pin 3: Blue (GND), Pin 4: Green (GND), Pin 5: D/W (GND)
DC Output Pinout — Positive +48V & N.C. (Pins 6–10) Pin 6: Red (+V), Pin 7: White (+V), Pin 8: Yellow (+V), Pin 9: Orange (+V), Pin 10: N.C. (No Connection)
Cooling Method Natural convection (fanless)
Ingress Protection Rating IP42
Overall Length (L) 222.0 mm (8.75 in) ±1 mm
Overall Width (W) 112.0 mm (4.40 in) ±1 mm
Overall Height (H) 45.0 mm (1.77 in) ±1 mm (44.0 mm body + 1.0 mm max rubber feet)
Overall Dimensions (L × W × H) 222 × 112 × 45 mm (8.75 × 4.40 × 1.77 in)
Net Weight 1,570 g (3.46 lbs / 55.38 oz)
Enclosure Material & Finish Black polycarbonate plastic housing with 4× bottom rubber feet
Manufacturer Warranty 3 Years

Industrial Applications & Compatible Systems

TDK-Lambda DTM300PW480D1 provides isolated, low-leakage 48 VDC bus power for high-wattage medical, laboratory, and industrial hardware:

  • Diagnostic Imaging & Patient-Monitoring Consoles: Powers mobile ultrasound carts, endoscopy processors, and Type B/BF bedside monitors requiring 4,000 VAC 2×MOPP isolation and IEC 60601-1-2 Ed4 EMC compliance.
  • Laboratory & Analytical Instrumentation: Supplies clean 48 VDC rails to automated blood analyzers, chromatography pumps, and clinical centrifuges with <100 µA touch current.
  • Industrial Test, Measurement & ATE Stations: Serves as an external fanless benchtop supply for automated optical inspection (AOI) rigs, semiconductor wafer handlers, and 48 VDC motorized actuators.
  • Telecom & PoE+ / PoE++ Peripheral Hubs: Delivers 300 W at 48 VDC for medical-grade network switches, wireless telemetry gateways, and distributed industrial controllers.

Installation, Thermal Management & MRO Guidelines

Installation & Engineering Guidelines: Because the DTM300PW480D1 relies entirely on fanless natural convection, mount or place it horizontally on its four rubber feet with at least 50 mm (2 in) of unobstructed air clearance around all sides and never enclose it in an unvented compartment. Maintain full 300 W (6.25 A) loading only between 0 °C and +40 °C (+32 °F to +104 °F); above +40 °C, derate linearly at 2.5 %/°C down to 150 W at +60 °C, and below 0 °C, derate linearly at 1.0 %/°C down to 210 W at -30 °C. Ensure all four +V conductors (Pins 6–9) and all five GND return conductors (Pins 1–5) are terminated in parallel through a mating Molex 39-28-1103 header so the 6.25 A current distributes evenly across the #16 AWG wires, and note that parallel unit operation is not permitted.