HPT5K0TS200-L XP Power 5kW 200VDC Medical Power Supply

$2,744.95

XP Power HPT5K0TS200-L (HPT5K0TS200L) 5kW 200VDC 25A 3-phase enclosed medical power supply, 180-528VAC input, featuring 0-105% Vprog, PMBus, and 2xMOPP.

SKU: HPT5K0TS200-L Category: Brand:

Description

XP Power HPT5K0TS200-L (also known as HPT5K0TS200L) is a high-density 5 kW three-phase AC-DC enclosed chassis-mount programmable power supply from the HPT5K0-L series, where “5K0” denotes the 5000W power rating, “TS200” specifies the single 200 VDC nominal output, and “-L” indicates the wide-range 180–528 VAC input architecture. Engineered for both constant voltage (CV) and constant current (CC) operation, it delivers 0 to 210.0 VDC and 0.0 to 25.0 A alongside a continuous 5V / 2A standby output. The unit operates from a 3-wire plus earth three-phase input spanning 342–528 VAC for full 5 kW output and 180–264 VAC for 2.5 kW output, achieving 93% typical efficiency at 480 VAC full load and a 0.96 power factor via fixed-frequency 60 kHz active PFC. Output noise is tightly controlled within 1% peak-to-peak with ±0.5% line regulation and 1% load regulation.

Housed in a compact 2U low-profile aluminum chassis measuring 254.0 mm × 63.5 mm × 336.5 mm (10.00 in × 2.50 in × 13.25 in) and weighing 5.7 kg (12.5 lbs), the supply integrates intelligent variable-speed fan cooling and M4/M5 screw terminals with a 30-pin JST PHDR-30VS signal interface supporting PMBus, CANopen, RS485, MODBUS, SCPI, and 0–5V analog programming. Comprehensive protection circuits include 110–120% latching overvoltage protection (OVP), 108% ±3% constant-current overload protection (OLP), short-circuit protection, phase-loss shutdown, and auto-resetting thermal protection. Reinforced galvanic isolation rated at 6000 VDC input-to-output (2 x MOPP) and 4000 VDC input-to-ground (1 x MOPP) ensures full compliance with IEC/EN/ANSI/AAMI ES60601-1 Edition 3 medical, UL/IEC/EN 62368-1 ITE, and SEMI F47 semiconductor standards across an operating ambient range of -20 °C to +70 °C (-4 °F to +158 °F).

Key Features

  • Wide Programmable CV/CC Output: Delivers 200 VDC nominal output adjustable from 0 to 210.0 VDC (0–105%) and 0.0 to 25.0 A rated current programmable up to 110% (27.5 A) via analog or digital command.
  • Dual-Range Three-Phase AC Input: Accepts 3-phase 3-wire plus earth input from 342–528 VAC for full 5000W output or 180–264 VAC for 2500W output, incorporating internal F16A/500V fuses in each phase.
  • Multi-Protocol Digital & Analog Control: Standard built-in I2C PMBus, CANopen, RS485 (MODBUS/SCPI), and 0–5V analog Vprog/Iprog interfaces enable real-time telemetry, GUI configuration, and active current sharing up to 5 units.
  • High Power Density & Efficiency: Achieves up to 93% efficiency at 480 VAC full load with 0.96 active power factor correction, packing 15.38 W/in³ into a 2U 254.0 × 63.5 × 336.5 mm enclosure.
  • Medical 2xMOPP & ITE Safety Approvals: Certified to IEC/EN 60601-1 Ed. 3, ANSI/AAMI ES60601-1, and UL/IEC/EN 62368-1 with 6000 VDC input-to-output (2xMOPP) and 4000 VDC input-to-ground (1xMOPP) isolation.
  • Fast Slew & Dynamic Response: Supports rapid voltage programming slew rates under 40 ms (10–105% and 105–10%) and recovers within 1% in less than 500 µs during 50–75–50% load transients.
  • Comprehensive Fault & Phase Protection: Features latching OVP (110–120%), constant-current OLP (108% ±3%), auto-recovery phase-loss shutdown after 0.5 s, and a 10 s pre-shutdown fan fail / overtemperature alarm.
  • SEMI F47 & Class B Conducted EMI: Complies with SEMI F47-0706 voltage sag immunity at 480 VAC, EN 55011/EN 55032 Class B conducted emissions, and Level 4 ESD/Surge industrial immunity standards.

Technical Specifications

Manufacturer XP Power
Manufacturer Part Number (MPN) HPT5K0TS200-L
Alternate Part Number HPT5K0TS200L
Product Series HPT5K0-L Series
Product Category Medical Enclosed PSUs
Component Type Programmable Three-Phase AC-DC Enclosed Chassis Mount Power Supply
Circuit Topology Fixed-Frequency Active PFC (60 kHz) + Variable-Frequency Isolated Main Converter (40–250 kHz)
Nominal Output Voltage 200 VDC (Main V1) / 5 VDC (Standby)
Output Voltage Adjustment Range 0 to 210.0 VDC (0% to 105% via 0–5V Analog Vprog or Digital Bus; ±10% of set output via potentiometer up to 105% max)
Rated Output Current 25.0 A (at 200 VDC Main Output) / 2.0 A (at 5 VDC Standby)
Output Current Range 0.0 to 25.0 A (Programmable current limit 0% to 110% of nominal via 0–5V Iprog or Digital Bus; 0 A minimum load)
Rated Output Power 5000 W (5 kW at 342–528 VAC input) / 2500 W (2.5 kW at 180–264 VAC input)
Number of Outputs 2 Outputs (Ch1 Main: 200 VDC / 25.0 A; Ch2 Standby: 5 VDC / 2.0 A)
Output Ripple & Noise (Max.) 1% of nominal voltage (2.0 Vp-p max on 200 VDC main output) / 2.5% on 5V Standby (20 MHz bandwidth, 0–50 °C)
Output Voltage Tolerance ±0.5% of nominal voltage (Main V1) / ±4% (5V Standby)
Line Regulation ±0.5% of nominal voltage (Main V1) / ±0.5% (5V Standby)
Load Regulation 1% (0–100% or 100–0% load on Main V1) / 2% (5V Standby)
Setup Time 2000 ms typical / 2300 ms max (Start-up delay under all line and load conditions)
Rise Time 40 ms max (Start-up rise time); <40 ms voltage programming slew rate (10–105% & 105–10%)
Hold-Up Time (at 115 VAC) 40 ms min / 44 ms typical (at 180 VAC, 2.5 kW load, 25 °C)
Hold-Up Time (at 230 VAC) 20 ms min / 22 ms typical (at 380 VAC, 5 kW load, 25 °C)
AC Input Voltage Range 180 to 528 VAC Three-Phase (3-Wire + Earth: 342–528 VAC for 5 kW max; 180–264 VAC for 2.5 kW max)
Input Frequency Range 47 to 63 Hz
Efficiency 93% typical / 92% minimum (Measured at 480 VAC and 342–528 VAC input, full load)
AC Input Current (at 115 VAC) 11 A max per phase (at 180 VAC input, 2.5 kW output)
AC Input Current (at 230 VAC) 10 A max per phase (at 342 VAC input, 5 kW output)
Inrush Current (Cold Start at 230 VAC) 60 A max per phase (at 528 VAC, 5 kW)
Power Factor (PF at 230 VAC) 0.96 typical (Complies with EN 61000-3-2 Class A)
Power Factor Correction (PFC) Active PFC (Fixed switching frequency 55–65 kHz, 60 kHz typical)
Switching Frequency 55 to 65 kHz (60 kHz typical Fixed-Frequency PFC) / 40 to 250 kHz (Variable-Frequency Main Converter)
Mounting Style Chassis Mount (4 x M4 bottom mounting points and 4 x M4 side mounting points, 4.0 mm max screw penetration)
Cooling Method Forced Air Cooling via 3 x internal 60 x 38 mm fans with intelligent speed control
Package / Footprint Standard 2U Low-Profile Enclosed Chassis (15.38 W/in³ power density)
Overall Width (W) 254.0 mm (10.00 in)
Overall Height (H) 63.5 mm (2.50 in)
Overall Depth (D) 336.5 mm (13.25 in)
Mounting Hole Pitch Bottom: 241.3 x 152.4 mm (9.50 x 6.00 in); Side: 215.9 x 31.75 mm (8.50 x 1.25 in)
Net Weight 5.7 kg (12.5 lbs)
Termination / Connector Style AC Input: 3 x M4 Screw Terminals (13 mm pitch) + M4 GND Stud; DC Output: 4 x M5 Screw Terminals; Signals: 30-Pin JST S30B-PHDSS Header (mates with JST PHDR-30VS)
Input Terminal Pinout Front Panel AC Barrier Block: 3 x M4 Phase Terminals (L1, L2, L3) + Chassis M4 x 0.7 GND Stud (F16A/500V fuse fitted in each phase)
Output & Signal Terminal Pinout Rear Busbars: 4 x M5 Screws (+Vout, -Vout); J1 30-Pin Connector: Pins 1-2 DCOK/RTN, Pins 3-4 Remote Inhibit/RTN, Pins 5-7 I2C Address A0-A2, Pins 8/10 CANH/CANL, Pins 9/11/13/15 RS485 Y/Z/A/B, Pin 12 SGND, Pins 14/16 I2C SDA/SCL, Pin 17 Fan Fail/Temp Warn, Pin 18 SYNC, Pin 19 VPROG, Pins 20-21 NC on 200V model, Pin 22 IPROG, Pin 23 ISHARE, Pin 24 PMBUS_EN, Pins 25-26 ACOK/RTN, Pins 27-28 +5VSBY, Pins 29-30 5VSBY_RTN
Control & Monitoring Signals V Program (0–5V for 0–105% Vout, ±1% accuracy), I Program (0–5V for 0–110% Iout, ±2% accuracy), Active Current Share (up to 5 units in parallel, ±3% accuracy), AC OK (2 ms warning), DC OK, Fan Fail / Temp Warning (10 s pre-shutdown notice), Remote Inhibit, Turn-On Sync (Note: Remote Sense not fitted on HPT5K0TS200-L)
Communication Interface PMBus (I2C), CANopen, RS485 (Full/Half Duplex MODBUS & SCPI), compatible with XP PS Manager GUI
Status Indicator Dual Front-Panel LEDs: AC OK (Green) and DC OK (Green) with solid and diagnostic blink states (0.2s / 1.0s intervals)
Terminal Tightening Torque AC Input M4 Screws: 1.0–1.1 Nm (8.9–9.7 in-lbs); M4 GND Stud: 1.2–1.7 Nm (10.6–15.0 in-lbs); DC Output M5 Screws: 3.2 Nm max; M4 Mounting Screws: 1.6 Nm max
Overload Protection (OLP) Threshold 108% ±3% of maximum rated current (Main V1); <5 A max (5V Standby); 3.4 kW constant power limit at 180–264 VAC low line
Overload Protection (OLP) Mode Constant current limit characteristic on Main V1; Foldback current limiting on 5V Standby
Overvoltage Protection (OVP) Threshold 110% to 120% of nominal output voltage (220 to 240 VDC)
Overvoltage Protection (OVP) Mode Latching shutdown (Recycle AC input power to reset; no OVP on 5V Standby)
Overtemperature Protection (OTP) Mode Auto-resetting thermal shutdown (FAN_FAIL/Temp Warning logic high alarm triggered 10 seconds prior to output shutdown)
Short Circuit Protection (SCP) Continuous constant current characteristic on Main V1; Foldback <5 A max on 5V Standby; Loss of phase shuts down after 0.5 s with auto-recovery
Withstand Isolation Voltage (I/P to O/P) 6000 VDC (With M3x0.5 HI-POT screw removed; 2121 VDC max test voltage with HI-POT screw installed)
Withstand Isolation Voltage (I/P to FG / Case) 4000 VDC (With M3x0.5 HI-POT screw removed; 2121 VDC max test voltage with HI-POT screw installed)
Withstand Isolation Voltage (O/P to FG / Case) 500 VDC
Earth Leakage Current 1.0 mA typical (Normal condition at 528 VAC / 60 Hz) / 3.3 mA max (Single fault condition at 528 VAC / 60 Hz)
Medical Isolation Rating 2 x MOPP (Primary to Secondary) / 1 x MOPP (Primary to Earth) per IEC 60601-1 Ed. 3; Class I Equipment
Operating Ambient Temperature Range -20 °C to +70 °C (-4 °F to +158 °F)
Full-Load Operating Temperature Limit +50 °C (+122 °F) for 100% rated output power
High-Temperature Derating Rate Derate linearly at 2.5% per °C from 100% load at +50 °C to 50% rated power at +70 °C (+122 °F to +158 °F)
Low-Line Input Derating Derate output power linearly from 5.0 kW at 342 VAC down to 2.5 kW at 180 VAC input
Altitude Derating Operating: Up to 3000 m (9,842 ft) for Medical / 5000 m (16,404 ft) for ITE; Non-operating transport: Up to 10,000 m (32,808 ft)
Storage Temperature Range -40 °C to +85 °C (-40 °F to +185 °F)
Operating Relative Humidity 5% to 95% RH (Non-condensing)
Temperature Coefficient ±0.03% / °C of maximum load
Vibration Resistance 10 to 500 Hz at 2 g single-axis sweep and endurance at resonance in all 3 planes (EN 60068-2-6); Shock: ±3 x 30 g (11 ms half-sine, 18 shocks total per EN 60068-2-27 & EN 60068-2-47); Acoustic Noise: <70 dB(A) Lw
Safety Approvals (UL / North America) UL 62368-1, CSA C22.2 No. 62368-1, UL 60950-1 (ITE); ANSI/AAMI ES60601-1:2005 & CSA C22.2 No. 60601-1:08 (Medical)
Safety Approvals (IEC / EN / TÜV) CB Report IEC 62368-1 Ed. 2, TÜV EN 62368-1 (ITE); CB Report IEC 60601-1 Ed. 3 Including Risk Management, TÜV EN 60601-1/2006 (Medical)
Safety Approvals (Regional / Global) CE Mark, UKCA, SEMI F47-0706 (at 480 VAC nominal)
EMC Conducted & Radiated Emissions EN 55011 / EN 55032 Class B Conducted Emissions (up to ~80% output current, Class A above), EN 55011 / EN 55032 Class A Radiated Emissions
EMC Harmonic & Flicker Compliance EN 61000-3-2 Class A (Harmonic Currents)
EMC Immunity Standards EN 60601-1-2, EN 61000-4-2 (ESD Level 4: ±8 kV contact / ±15 kV air, Criteria A), EN 61000-4-3 (Radiated Level 3, Criteria A), EN 61000-4-4 (EFT/Burst Level 3, Criteria A), EN 61000-4-5 (Surge Installation Class 4, Criteria A), EN 61000-4-6 (Conducted RF Level 3, Criteria A), EN 61000-4-8 (Magnetic Field Level 4, Criteria A), EN 61000-4-11 (Voltage Dips & Interruptions at 200/380 VAC & 240/480 VAC, Criteria A/B), SEMI F47-0706 (Criteria A)
MTBF (Telcordia SR-332 / Bellcore) 600,000 Hours (600 kHrs per Telcordia SR-332)
Environmental Compliance RoHS Compliant, 3-Year Manufacturer Warranty

Industrial Applications & Compatible Systems

The XP Power HPT5K0TS200-L is engineered for mission-critical industrial, healthcare, and semiconductor platforms requiring high-voltage programmable DC power:

  • Medical Imaging & Diagnostic Systems: Provides 2xMOPP patient-isolated 200 VDC power for medical lasers, X-ray subsystems, MRI gradient amplifiers, and clinical sterilization equipment certified to IEC 60601-1 Ed. 3.
  • Semiconductor Wafer Fab & Plasma Processing: Delivers SEMI F47 sag-immune high-voltage DC rails for ion implanters, RF plasma generators, physical vapor deposition (PVD) chambers, and automated wafer inspection tools.
  • Automated Test Equipment (ATE) & Burn-In Racks: Acts as a fast-slewing (<40 ms) programmable 0–210 VDC / 0–25 A CV/CC source controlled via PMBus, CANopen, MODBUS, or SCPI for EV battery module testing and component burn-in.
  • High-Power Industrial Lasers & Electrolysis: Supports scalable parallel current sharing of up to 5 units (up to 25 kW combined system capacity) for industrial fiber lasers, UV curing arrays, and precision electrochemical rectifiers.

Installation, Thermal Management & MRO Guidelines

Follow these critical electrical and mechanical engineering protocols when integrating the HPT5K0TS200-L into OEM enclosures:

  • Digital vs. Analog Programming Pin Configuration: By default, leaving Pin 24 (PMBUS_EN) open selects digital bus control; to enable 0–5V analog Vprog (Pin 19) and Iprog (Pin 22), pull Pin 24 low to SGND (Pin 12) or 5VSBY_RTN (Pins 29/30), noting that open-circuit analog pins default to 0% output.
  • Three-Phase Line Derating & Torque Compliance: Supply 342–528 VAC three-phase input to unlock the full 5 kW rating (derating linearly to 2.5 kW at 180–264 VAC), and torque the M4 AC terminals to 1.0–1.1 Nm and M5 DC output busbars up to 3.2 Nm.
  • Hi-Pot Testing & Thermal Derating Management: Remove the side M3x0.5 HI-POT screw prior to performing 6000 VDC isolation tests (maximum test voltage is 2121 VDC with the screw installed), and derate output power linearly at 2.5%/°C above +50 °C up to +70 °C.