Description
XP Power PBL500PS12C is a high-density, Class II enclosed AC-DC switching power supply from the PBL500 series, engineered specifically for medical equipment requiring Body Floating (BF) patient-contact isolation. In XP Power’s nomenclature, PBL500 identifies the 500W medical series, PS designates single main output, 12 specifies the 12 VDC primary rail, and the C suffix denotes the factory-covered chassis configuration equipped with an integral temperature-controlled cooling fan (which utilizes the internal 12 V / 0.3 A fan supply). Accepting a universal 80 to 264 VAC input with active power factor correction, the unit delivers up to 450 W continuous output power (37.50 A at 12 VDC) alongside an isolated 5.0 VDC / 2.0 A always-on standby rail, achieving 90% typical efficiency and limiting output ripple and noise to 120 mV pk-pk.
Housed in a compact perforated metal enclosure measuring 180.3 x 101.6 x 67.0 mm (7.10 x 4.00 x 2.64 in) and weighing 1.14 kg (2.52 lbs), the unit features dual-standard #6-32 and M3x0.5 mounting inserts and screw-terminal power connections (Dinkle M3 input, M4x0.7 output) plus an 8-pin Molex signal header for Remote Sense, TTL Inhibit, and AC OK/POK monitoring. Comprehensive protection includes latching overvoltage (112–140%), auto-recovery trip-and-restart overcurrent (105–140%), and latching thermal shutdown. Reinforced galvanic isolation provides 4000 VAC (2xMOPP) input-to-output and input-to-ground withstand ratings with ultra-low 80 µA maximum patient leakage current, certified to IEC/EN/ES/CSA 60601-1 and EN 55011/55032 Class B emissions up to 5000 m altitude.
Key Features
- High-Density 450W Medical Power: Delivers 12 VDC at 37.50 A continuous forced-air output (20.83 A / 250 W convection) with 90% typical efficiency at 230 VAC full load.
- 2xMOPP Reinforced Isolation & BF Rating: Provides 4000 VAC input-to-output and input-to-ground isolation with 80 µA maximum patient leakage current at 264 VAC for Type BF medical devices.
- Integral Smart Fan Cooling (‘C’ Version): Factory-fitted top cover houses a variable-speed fan triggered at 30 °C transformer temperature and reaching full speed at 60 °C for low acoustic noise.
- Auxiliary Standby & System Control Signals: Integrates an always-on 5.0 VDC / 2.0 A standby output, 0.5 V remote sense compensation, TTL high inhibit (1 W no-load), and AC OK/POK status monitoring.
- Universal Wide-Range Input & High Altitude: Operates across 80 to 264 VAC (47–63 Hz) with dual line/neutral fuses and full safety approval for operation up to 5000 m (16,404 ft) elevation.
- Class B EMC & Comprehensive Protection: Meets EN 55011/55032 Class B conducted and radiated emissions without external filtering, backed by latching OVP/OTP and trip-and-restart OCP.
Technical Specifications
| Manufacturer | XP Power |
|---|---|
| Manufacturer Part Number (MPN) | PBL500PS12C |
| Product Series | PBL500 Series (Covered ‘C’ Version) |
| Product Category | Converter Heatsinks |
| Component Type | AC-DC Medical Switching Power Supply (Class II) |
| Circuit Topology | Active PFC + Variable-Frequency Main Converter + Auxiliary Standby Converter |
| Nominal Output Voltage | 12 VDC (Main V1) |
| Rated Output Current | 37.50 A (Integral Fan Cooled) / 20.83 A (Convection Cooled) |
| Rated Output Power | 450 W (Integral Fan Cooled) / 250 W (Convection Cooled) |
| Number of Outputs | 2 User Outputs (Main V1 + Auxiliary Standby V2; Internal V3 dedicated to integral fan) |
| Ch1 Output Voltage & Current | 12 VDC @ 37.50 A (450 W max) |
| Ch2 Output Voltage & Current | 5.0 VDC @ 2.0 A (10 W Isolated Standby Supply V2) |
| Output Ripple & Noise (Max.) | 120 mV pk-pk (1% pk-pk max, 20 MHz BW with 10 µF tantalum + 0.1 µF ceramic capacitors) |
| Output Voltage Tolerance | ±2% (Combined line & load regulation; 0.1% minimum load required) |
| Transient Recovery Time & Deviation | <500 µs recovery within 1% for a 25% step load change (4% max deviation) |
| Hold-Up Time (at 115 VAC) | 20 ms min (at 110 VAC & 500 W load) |
| AC Input Voltage Range | 80 to 264 VAC (Single-Phase) |
| Input Frequency Range | 47 to 63 Hz |
| Efficiency (at 230 VAC) | 90% typical (at 100% load) |
| AC Input Current (at 115 VAC) | 5.2 A rms max (at 60 Hz full load) |
| AC Input Current (at 230 VAC) | 2.6 A rms max (at 50 Hz full load) |
| Inrush Current (Cold Start at 115 VAC) | 30 A max (at 25 °C) |
| Inrush Current (Cold Start at 230 VAC) | 60 A max (at 25 °C) |
| Power Factor Correction (PFC) | Active PFC (EN 61000-3-2 Class A compliant) |
| No-Load Power Consumption | 1 W max (when TTL Inhibit is active) |
| Switching Frequency | PFC: 55–75 kHz (65 kHz typ) / Main Converter: 90–300 kHz (variable) / Standby: 80–120 kHz |
| Mounting Style | Chassis Mount (8x #6-32 & 8x M3x0.5 threaded inserts on base and sides; 4 mm max screw penetration) |
| Cooling Method | Forced Air via Integral Temperature-Controlled Fan (starts ~30 °C, full speed ~60 °C at transformer T1) |
| Overall Width (W) | 101.6 mm / 4.00 in (Chassis body 100.0 mm / 3.94 in) |
| Overall Height (H) | 67.0 mm / 2.64 in (Covered enclosure height) |
| Overall Depth (D) | 180.3 mm / 7.10 in (Chassis length 180.0 mm / 7.09 in) |
| Dimensional Tolerance | ±0.5 mm (±0.02 in) |
| Net Weight | 1.14 kg (2.52 lbs) |
| Termination / Connector Style | Input P1: Dinkle DT-35C-B01W-03 (M3 screws); Output P2: M4x0.7 screw terminals; Signals P3: Molex 87833-08 |
| Input Terminal Pinout | Connector P1 — Pin 1: Line (L), Pin 2: Neutral (N) (Class II; dual internal fuses in L and N) |
| Output & Signal Terminal Pinout | Connector P2 — Pin 1: +V1 (+12 VDC), Pin 2: -V1 (Common Return); Connector P3 (mates Molex 51110-0850) — Pin 1: Common Return, Pin 2: +V1 Sense, Pin 3: -V1 Sense, Pin 4: AC OK / POK, Pin 5: Inhibit, Pin 6: +V2 (5 VDC Standby), Pin 7: +V2 (5 VDC Standby), Pin 8: Common Return |
| Control & Monitoring Signals | Remote Sense (compensates up to 0.5 V total drop), TTL High Inhibit, AC OK / Power OK (TTL high normal, 100–1000 ms turn-on delay, 1 ms power-loss warning) |
| Overload Protection (OLP) Threshold | 105% to 140% of rated load |
| Overload Protection (OLP) Mode | Trip & Restart (Auto-Recovery) |
| Overvoltage Protection (OVP) Threshold | 112% to 140% of nominal V1 voltage |
| Overvoltage Protection (OVP) Mode | Latching (Requires AC input recycle to reset) |
| Overtemperature Protection (OTP) Mode | Latching thermal shutdown |
| Withstand Isolation Voltage (I/P to O/P) | 4000 VAC (2xMOPP) |
| Withstand Isolation Voltage (I/P to FG / Case) | 4000 VAC (2xMOPP) |
| Withstand Isolation Voltage (O/P to FG / Case) | 1500 VAC (1xMOPP) |
| Earth Leakage Current | Patient Leakage Current: 50 µA typical / 80 µA max (at 264 VAC, 63 Hz) |
| Medical Isolation Rating | 2xMOPP (Input-to-Output & Input-to-Ground), 1xMOPP (Output-to-Ground); Type BF Rated |
| Operating Ambient Temperature Range | 0 °C to +70 °C (32 °F to 158 °F) |
| Full-Load Operating Temperature Limit | +40 °C (+104 °F) at 450 W full load |
| High-Temperature Derating Rate | Linear derating from 100% load (450 W) at +40 °C to 50% load (225 W) at +60 °C, and 0 W at +70 °C |
| Low-Line Input Derating | Linear derating to 90% load at 85 VAC and 80% load at 80 VAC |
| Altitude Derating | Certified for continuous operation up to 5000 m (16,404 ft) |
| Storage Temperature Range | -10 °C to +85 °C (14 °F to 185 °F) |
| Operating Relative Humidity | 5% to 95% RH (non-condensing) |
| Temperature Coefficient | ±0.04% / °C |
| Safety Approvals (UL / North America) | UL / ES 60601-1, CSA C22.2 No. 60601-1 (Medical) |
| Safety Approvals (IEC / EN / TÜV) | IEC 60601-1, EN 60601-1 (CB Scheme Medical) |
| Safety Approvals (Regional / Global) | CE, UKCA |
| EMC Conducted & Radiated Emissions | EN 55032 Class B, EN 55011 Class B |
| EMC Harmonic & Flicker Compliance | EN 61000-3-2 Class A, EN 61000-3-3 |
| EMC Immunity Standards | EN 61000-4-2 Level 4 (±8 kV contact / ±15 kV air, Criteria A); EN 61000-4-3 (10 V/m, Criteria A); EN 61000-4-4 (±2 kV EFT/Burst, Criteria A); EN 61000-4-5 Level 3 (±2 kV L-N Surge, Criteria A); EN 61000-4-6 (10 Vrms Conducted, Criteria A); EN 61000-4-8 (30 A/m Magnetic Field, Criteria A); EN 61000-4-11 Voltage Dips & Interruptions |
| MTBF (MIL-HDBK-217F) | 100,000 hours min (Full load at 25 °C, Ground Benign) |
Industrial Applications & Compatible Systems
XP Power PBL500PS12C is engineered for Class II medical and precision laboratory equipment requiring reinforced 2xMOPP patient isolation, low leakage current, and self-contained forced-air cooling:
- Patient-Connected Diagnostic & Monitoring Systems: Powers ultrasound consoles, patient vital-signs monitors, and ECG/EEG carts requiring Type BF isolation and 80 µA max patient leakage.
- Surgical & Therapeutic Equipment: Supplies clean 12 VDC main power and 5 VDC standby logic for surgical pumps, endoscopic light sources, and respiratory ventilators.
- Clinical & Analytical Laboratory Instrumentation: Drives high-throughput hematology analyzers, automated immunoassay platforms, and spectrophotometers operating in high-altitude clinics up to 5000 m.
- Home Healthcare & Dental Systems: Ideal for Class II ungrounded medical enclosures, dental imaging units, and portable dialysis consoles requiring EN 55011 Class B quiet electromagnetic performance.
Installation, Thermal Management & MRO Guidelines
Follow these mechanical, thermal, and electrical integration guidelines during OEM assembly and MRO replacement:
- Chassis Screw Penetration & Airflow Clearance: Limit #6-32 or M3x0.5 mounting screw insertion depth to 4.0 mm maximum from the outer chassis surface to avoid violating internal 2xMOPP creepage distances, and maintain unobstructed intake/exhaust paths for the top-mounted fan.
- Thermal & Low-Line Load Derating: Keep ambient intake temperature at or below +40 °C (+104 °F) for full 450 W output; derate linearly to 225 W at +60 °C and apply additional derating to 90% at 85 VAC or 80% at 80 VAC.
- Minimum Load & Remote Sense Wiring: Maintain at least a 0.1% minimum load on the 12 VDC V1 output to guarantee ±2% regulation, and route P3 twisted-pair sense leads (+V1 Sense / -V1 Sense) directly to the load terminals to compensate up to 0.5 V of distribution drop.