PBL500PS12C XP Power 450W 12VDC Medical Enclosed PSU

$288.11

XP Power PBL500PS12C 450W 12VDC 37.5A enclosed medical AC-DC PSU with integral fan, 80-264VAC input, 5V/2A standby, 2xMOPP 4000VAC isolation, and IEC 60601-1 BF.

SKU: PBL500PS12C Category: Brand:

Description

XP Power PBL500PS12C is a high-density, Class II enclosed AC-DC switching power supply from the PBL500 series, engineered specifically for medical equipment requiring Body Floating (BF) patient-contact isolation. In XP Power’s nomenclature, PBL500 identifies the 500W medical series, PS designates single main output, 12 specifies the 12 VDC primary rail, and the C suffix denotes the factory-covered chassis configuration equipped with an integral temperature-controlled cooling fan (which utilizes the internal 12 V / 0.3 A fan supply). Accepting a universal 80 to 264 VAC input with active power factor correction, the unit delivers up to 450 W continuous output power (37.50 A at 12 VDC) alongside an isolated 5.0 VDC / 2.0 A always-on standby rail, achieving 90% typical efficiency and limiting output ripple and noise to 120 mV pk-pk.

Housed in a compact perforated metal enclosure measuring 180.3 x 101.6 x 67.0 mm (7.10 x 4.00 x 2.64 in) and weighing 1.14 kg (2.52 lbs), the unit features dual-standard #6-32 and M3x0.5 mounting inserts and screw-terminal power connections (Dinkle M3 input, M4x0.7 output) plus an 8-pin Molex signal header for Remote Sense, TTL Inhibit, and AC OK/POK monitoring. Comprehensive protection includes latching overvoltage (112–140%), auto-recovery trip-and-restart overcurrent (105–140%), and latching thermal shutdown. Reinforced galvanic isolation provides 4000 VAC (2xMOPP) input-to-output and input-to-ground withstand ratings with ultra-low 80 µA maximum patient leakage current, certified to IEC/EN/ES/CSA 60601-1 and EN 55011/55032 Class B emissions up to 5000 m altitude.

Key Features

  • High-Density 450W Medical Power: Delivers 12 VDC at 37.50 A continuous forced-air output (20.83 A / 250 W convection) with 90% typical efficiency at 230 VAC full load.
  • 2xMOPP Reinforced Isolation & BF Rating: Provides 4000 VAC input-to-output and input-to-ground isolation with 80 µA maximum patient leakage current at 264 VAC for Type BF medical devices.
  • Integral Smart Fan Cooling (‘C’ Version): Factory-fitted top cover houses a variable-speed fan triggered at 30 °C transformer temperature and reaching full speed at 60 °C for low acoustic noise.
  • Auxiliary Standby & System Control Signals: Integrates an always-on 5.0 VDC / 2.0 A standby output, 0.5 V remote sense compensation, TTL high inhibit (1 W no-load), and AC OK/POK status monitoring.
  • Universal Wide-Range Input & High Altitude: Operates across 80 to 264 VAC (47–63 Hz) with dual line/neutral fuses and full safety approval for operation up to 5000 m (16,404 ft) elevation.
  • Class B EMC & Comprehensive Protection: Meets EN 55011/55032 Class B conducted and radiated emissions without external filtering, backed by latching OVP/OTP and trip-and-restart OCP.

Technical Specifications

Manufacturer XP Power
Manufacturer Part Number (MPN) PBL500PS12C
Product Series PBL500 Series (Covered ‘C’ Version)
Product Category Converter Heatsinks
Component Type AC-DC Medical Switching Power Supply (Class II)
Circuit Topology Active PFC + Variable-Frequency Main Converter + Auxiliary Standby Converter
Nominal Output Voltage 12 VDC (Main V1)
Rated Output Current 37.50 A (Integral Fan Cooled) / 20.83 A (Convection Cooled)
Rated Output Power 450 W (Integral Fan Cooled) / 250 W (Convection Cooled)
Number of Outputs 2 User Outputs (Main V1 + Auxiliary Standby V2; Internal V3 dedicated to integral fan)
Ch1 Output Voltage & Current 12 VDC @ 37.50 A (450 W max)
Ch2 Output Voltage & Current 5.0 VDC @ 2.0 A (10 W Isolated Standby Supply V2)
Output Ripple & Noise (Max.) 120 mV pk-pk (1% pk-pk max, 20 MHz BW with 10 µF tantalum + 0.1 µF ceramic capacitors)
Output Voltage Tolerance ±2% (Combined line & load regulation; 0.1% minimum load required)
Transient Recovery Time & Deviation <500 µs recovery within 1% for a 25% step load change (4% max deviation)
Hold-Up Time (at 115 VAC) 20 ms min (at 110 VAC & 500 W load)
AC Input Voltage Range 80 to 264 VAC (Single-Phase)
Input Frequency Range 47 to 63 Hz
Efficiency (at 230 VAC) 90% typical (at 100% load)
AC Input Current (at 115 VAC) 5.2 A rms max (at 60 Hz full load)
AC Input Current (at 230 VAC) 2.6 A rms max (at 50 Hz full load)
Inrush Current (Cold Start at 115 VAC) 30 A max (at 25 °C)
Inrush Current (Cold Start at 230 VAC) 60 A max (at 25 °C)
Power Factor Correction (PFC) Active PFC (EN 61000-3-2 Class A compliant)
No-Load Power Consumption 1 W max (when TTL Inhibit is active)
Switching Frequency PFC: 55–75 kHz (65 kHz typ) / Main Converter: 90–300 kHz (variable) / Standby: 80–120 kHz
Mounting Style Chassis Mount (8x #6-32 & 8x M3x0.5 threaded inserts on base and sides; 4 mm max screw penetration)
Cooling Method Forced Air via Integral Temperature-Controlled Fan (starts ~30 °C, full speed ~60 °C at transformer T1)
Overall Width (W) 101.6 mm / 4.00 in (Chassis body 100.0 mm / 3.94 in)
Overall Height (H) 67.0 mm / 2.64 in (Covered enclosure height)
Overall Depth (D) 180.3 mm / 7.10 in (Chassis length 180.0 mm / 7.09 in)
Dimensional Tolerance ±0.5 mm (±0.02 in)
Net Weight 1.14 kg (2.52 lbs)
Termination / Connector Style Input P1: Dinkle DT-35C-B01W-03 (M3 screws); Output P2: M4x0.7 screw terminals; Signals P3: Molex 87833-08
Input Terminal Pinout Connector P1 — Pin 1: Line (L), Pin 2: Neutral (N) (Class II; dual internal fuses in L and N)
Output & Signal Terminal Pinout Connector P2 — Pin 1: +V1 (+12 VDC), Pin 2: -V1 (Common Return); Connector P3 (mates Molex 51110-0850) — Pin 1: Common Return, Pin 2: +V1 Sense, Pin 3: -V1 Sense, Pin 4: AC OK / POK, Pin 5: Inhibit, Pin 6: +V2 (5 VDC Standby), Pin 7: +V2 (5 VDC Standby), Pin 8: Common Return
Control & Monitoring Signals Remote Sense (compensates up to 0.5 V total drop), TTL High Inhibit, AC OK / Power OK (TTL high normal, 100–1000 ms turn-on delay, 1 ms power-loss warning)
Overload Protection (OLP) Threshold 105% to 140% of rated load
Overload Protection (OLP) Mode Trip & Restart (Auto-Recovery)
Overvoltage Protection (OVP) Threshold 112% to 140% of nominal V1 voltage
Overvoltage Protection (OVP) Mode Latching (Requires AC input recycle to reset)
Overtemperature Protection (OTP) Mode Latching thermal shutdown
Withstand Isolation Voltage (I/P to O/P) 4000 VAC (2xMOPP)
Withstand Isolation Voltage (I/P to FG / Case) 4000 VAC (2xMOPP)
Withstand Isolation Voltage (O/P to FG / Case) 1500 VAC (1xMOPP)
Earth Leakage Current Patient Leakage Current: 50 µA typical / 80 µA max (at 264 VAC, 63 Hz)
Medical Isolation Rating 2xMOPP (Input-to-Output & Input-to-Ground), 1xMOPP (Output-to-Ground); Type BF Rated
Operating Ambient Temperature Range 0 °C to +70 °C (32 °F to 158 °F)
Full-Load Operating Temperature Limit +40 °C (+104 °F) at 450 W full load
High-Temperature Derating Rate Linear derating from 100% load (450 W) at +40 °C to 50% load (225 W) at +60 °C, and 0 W at +70 °C
Low-Line Input Derating Linear derating to 90% load at 85 VAC and 80% load at 80 VAC
Altitude Derating Certified for continuous operation up to 5000 m (16,404 ft)
Storage Temperature Range -10 °C to +85 °C (14 °F to 185 °F)
Operating Relative Humidity 5% to 95% RH (non-condensing)
Temperature Coefficient ±0.04% / °C
Safety Approvals (UL / North America) UL / ES 60601-1, CSA C22.2 No. 60601-1 (Medical)
Safety Approvals (IEC / EN / TÜV) IEC 60601-1, EN 60601-1 (CB Scheme Medical)
Safety Approvals (Regional / Global) CE, UKCA
EMC Conducted & Radiated Emissions EN 55032 Class B, EN 55011 Class B
EMC Harmonic & Flicker Compliance EN 61000-3-2 Class A, EN 61000-3-3
EMC Immunity Standards EN 61000-4-2 Level 4 (±8 kV contact / ±15 kV air, Criteria A); EN 61000-4-3 (10 V/m, Criteria A); EN 61000-4-4 (±2 kV EFT/Burst, Criteria A); EN 61000-4-5 Level 3 (±2 kV L-N Surge, Criteria A); EN 61000-4-6 (10 Vrms Conducted, Criteria A); EN 61000-4-8 (30 A/m Magnetic Field, Criteria A); EN 61000-4-11 Voltage Dips & Interruptions
MTBF (MIL-HDBK-217F) 100,000 hours min (Full load at 25 °C, Ground Benign)

Industrial Applications & Compatible Systems

XP Power PBL500PS12C is engineered for Class II medical and precision laboratory equipment requiring reinforced 2xMOPP patient isolation, low leakage current, and self-contained forced-air cooling:

  • Patient-Connected Diagnostic & Monitoring Systems: Powers ultrasound consoles, patient vital-signs monitors, and ECG/EEG carts requiring Type BF isolation and 80 µA max patient leakage.
  • Surgical & Therapeutic Equipment: Supplies clean 12 VDC main power and 5 VDC standby logic for surgical pumps, endoscopic light sources, and respiratory ventilators.
  • Clinical & Analytical Laboratory Instrumentation: Drives high-throughput hematology analyzers, automated immunoassay platforms, and spectrophotometers operating in high-altitude clinics up to 5000 m.
  • Home Healthcare & Dental Systems: Ideal for Class II ungrounded medical enclosures, dental imaging units, and portable dialysis consoles requiring EN 55011 Class B quiet electromagnetic performance.

Installation, Thermal Management & MRO Guidelines

Follow these mechanical, thermal, and electrical integration guidelines during OEM assembly and MRO replacement:

  • Chassis Screw Penetration & Airflow Clearance: Limit #6-32 or M3x0.5 mounting screw insertion depth to 4.0 mm maximum from the outer chassis surface to avoid violating internal 2xMOPP creepage distances, and maintain unobstructed intake/exhaust paths for the top-mounted fan.
  • Thermal & Low-Line Load Derating: Keep ambient intake temperature at or below +40 °C (+104 °F) for full 450 W output; derate linearly to 225 W at +60 °C and apply additional derating to 90% at 85 VAC or 80% at 80 VAC.
  • Minimum Load & Remote Sense Wiring: Maintain at least a 0.1% minimum load on the 12 VDC V1 output to guarantee ±2% regulation, and route P3 twisted-pair sense leads (+V1 Sense / -V1 Sense) directly to the load terminals to compensate up to 0.5 V of distribution drop.